South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-63 - Radioactive Materials (Title A)
Part IV - USE OF RADIONUCLIDES IN THE HEALTH PROFESSIONS
Subpart F - MANUAL BRACHYTHERAPY
Section 61-63.IV.F.RHA 4.51 - Calibration Measurements of Brachytherapy Sources

Current through Register Vol. 48, No. 9, September 27, 2024

4.51.1 Before the first medical use of a brachytherapy source on or after the effective date of these regulations, a licensee shall have--

4.51.1.1 Determined the source output or activity using a dosimetry system that meets the requirements of RHA 4.63.1;

4.51.1.2 Determined source positioning accuracy within applicators; and

4.51.1.3 Used published protocols currently accepted by nationally recognized bodies to meet the requirements of RHA 4.51.1.1 and 4.51.1.2.

4.51.2 A licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine that are made in accordance with RHA 4.51.1.

4.51.3 A licensee shall mathematically correct the outputs or activities determined in RHA 4.51.1 for physical decay at intervals consistent with 1 percent physical decay.

4.51.4 A licensee shall retain a record of each calibration in accordance with RHA 4.105.

Disclaimer: These regulations may not be the most recent version. South Carolina may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.