Current through Register Vol. 48, No. 9, September 27, 2024
(1) Triggered
source water monitoring.
(a) General
Requirement. A ground water system must conduct triggered source water
monitoring if the conditions identified in paragraphs (1)(a)(i) and either
(1)(a)(ii) or 1(a)(iii) of this section exist.
(i) The system does not provide at least
4-log treatment of viruses (using inactivation, removal, or a
Department-approved combination of 4-log virus inactivation and removal) before
or at the first customer for each ground water source; and either
(ii) The system is notified that a sample
collected under R.61-58.5.G(1)
is total coliform-positive and the sample is not invalidated under
R.61-58.5.G(3)
until March 31, 2016, or
(iii) The
system is notified that a sample collected under R.61-58.17.E through
R.61-58.17.H is total coliform-positive and the sample is not invalidated under
R.61-58.17.D(3) beginning April 1, 2016.
(b) Sampling Requirements. A ground water
system must collect, within 24 hours of notification of the total
coliform-positive sample, at least one ground water source sample from each
ground water source in use at the time the total coliform-positive sample was
collected under R.61-58.5.G(1)
until March 31, 2016, or collected under R.61-58.17.E through
R.61-58.17.H beginning April 1, 2016, except as provided in
R.61-58.16.E(1)(b)(ii).
(i) The Department
may extend the 24-hour time limit on a case-by-case basis if the system cannot
collect the ground water source sample within 24 hours due to circumstances
beyond its control. In the case of an extension, the Department must specify
how much time the system has to collect the sample.
(ii) If approved by the Department, systems
with more than one ground water source may meet the requirements of
R.61-58.16.E(1)(b) by sampling a representative ground water source or sources.
If directed by the Department, systems must submit a triggered source water
monitoring plan for Department approval that identifies one or more ground
water sources that are representative of each monitoring site in the system's
sample siting plan under R.61-58.5.G(1)
until March 31, 2016, or under R.61-58.17.D beginning April 1, 2016, and that
the system intends to use for representative sampling under this
paragraph.
(iii) Until, March 31,
2016, a ground water system serving 1,000 or fewer people may use a repeat
sample collected from a ground water source to meet both the requirements of
R.61-58.5.G(2)
and to satisfy the monitoring requirements of R.61-58.16.E(1)(b) for that
ground water source only if the Department approves the use of E.coli as a
fecal indicator for source water monitoring under R.61-58.16.E(1). If the
repeat sample collected from the ground water source is E.coli-positive, the
system must comply with R.61-58.16.E(1)(c).
(iv) Beginning April 1, 2016, a ground water
system serving 1,000 or fewer people may use a repeat sample collected from a
ground water source to meet both the requirements of R.61-58.17 and to satisfy
the monitoring requirements of R.61-58.16.E(1)(b) for that ground water source
only if the Department approves the use of E. coli as a fecal indicator for
source water monitoring under R.61-58.16.E(1) and approves the use of a single
sample for meeting both the triggered source water monitoring requirements in
R.61-58.16.E(1) and the repeat monitoring requirements in R.61-58.17.I. If the
repeat sample collected from the ground water source is E. coli-positive, the
system must comply with R.61-58.16.E(1)(c).
(c) Additional Requirements. If the
Department does not require corrective action under R.61-58.16.F(1)(b)
for a fecal indicator positive source water sample collected under
R.61-58.16.E(1)(b) that is not invalidated under R.61-58.16.E(4), the system
must collect five additional source water samples from the same source within
24 hours of being notified of the fecal indicator positive sample.
(d) Consecutive and wholesale systems.
(i) In addition to the other requirements of
R.61-58.16.E(1), a consecutive ground water system that has a total
coliform-positive sample collected under R.61-58.5.G(1)
until March 31, 2016, or under R.61-58.17.E through R.61-58.17.H beginning
April 1, 2016 must notify the wholesale system(s) within 24 hours of being
notified of the total coliform-positive sample.
(ii) In addition to the other requirements of
R.61-58.16.E(1), a wholesale ground water system must comply with
R.61-58.16.E(1)(d)(ii)(A) and R.61-58.16.E(1)(d)(ii)(B).
(A) A wholesale ground water system that
receives notice from a consecutive system it serves that a sample collected
under R.61-58.5.G(1)
until March 31, 2016, or collected under R.61-58.17.E through R.61-58.17.H
beginning April 1, 2016, is total coliform-positive must, within 24 hours of
being notified, collect a sample from its ground water source(s) under
R.61-58.16.E(1)(b) and analyze it for a fecal indicator under
R.61-58.16.E(3).
(B) If the sample
collected under R.61-58.16.E(1)(d)(ii)(A) is fecal indicator positive, the
wholesale ground water system must notify all consecutive systems served by
that ground water source of the fecal indicator positive sample within 24 hours
of being notified of the ground water source sample monitoring result and must
meet the requirements of R.61-58.16.E(1)(c).
(e) Exceptions to the triggered source water
monitoring requirements. A ground water system is not required to comply with
the source water monitoring requirements of R.61-58.16.E(1) if either one of
the following conditions exists:
(i) The
Department determines, and documents in writing, that the total
coliform-positive sample collected under R.61-58.5(G)(1)
until March 31, 2016, or under R.61-58.17.E through R.61-58.17.H beginning
April 1, 2016, is caused by a distribution system deficiency; or
(ii) The total coliform-positive sample
collected under R.61-58.5(G)(1) until March 31, 2016, or under R.61-58.17.E
through R.61-58.17.H beginning April 1, 2016, is collected at a location that
meets Department criteria for distribution system conditions that will cause
total coliform-positive samples.
(2) Assessment source water monitoring. If
directed by the Department, ground water systems must conduct assessment source
water monitoring that meets Department-determined requirements for such
monitoring. A ground water system conducting assessment source water monitoring
may use a triggered source water sample collected under R.61-58.16.E(1)(b) to
meet the requirements of R.61-58.16.E(2). Department-determined assessment
source water monitoring may include, but not be limited to the following:
(a) Collection of a total of 12 ground water
source samples that represent each month the system provides ground water to
the public.
(b) Collection of
samples from each well unless the system obtains written Department approval to
conduct monitoring at one or more wells within the ground water system that are
representative of multiple wells used by that system and that draw water from
the same hydrogeologic setting.
(c)
Collection of a standard sample volume of at least 100 mL for fecal indicator
analysis regardless of the fecal indicator or analytical method used.
(d) Analysis of all ground water source
samples using one of the analytical methods listed in R.61-58.16.E(3) for the
presence of E. coli, enterococci, or coliphage.
(e) Collection of ground water source samples
at a location prior to any treatment of the ground water source unless the
Department approves a sampling location after treatment.
(f) Collection of ground water source samples
at the well itself unless the system's configuration does not allow for
sampling at the well itself and the Department approves an alternate sampling
location that is representative of the water quality of that well.
(3) Analytical methods.
(a) A ground water system subject to the
source water monitoring requirements of R.61-58.16.E(1) must collect a standard
sample volume of at least 100 ml for fecal indicator analysis regardless of the
fecal indicator or analytical method used.
(b) A ground water system must analyze all
ground water source samples collected under R.61-58.16.E(1) for E.coli,
enterococci, or coliphage using EPA-approved methods listed in
40
CFR 141.402(c)(2) (Federal
Register 11-8-2006 edition).
(4) Invalidation of a fecal indicator
positive ground water source sample.
(a) A
ground water system may obtain Department invalidation of a fecal indicator
positive ground water source sample collected under R.61-58.16.E(1) only under
the conditions specified as follows:
(i) The
system provides the Department with written notice from the laboratory that
improper sample analysis occurred.
(ii) The Department determines and documents
in writing that there is substantial evidence that a fecal indicator positive
ground water source sample is not related to source water quality.
(b) If the Department invalidates
a fecal indicator positive ground water source sample, the ground water system
must collect another source water sample under R.61-58.16.E(1) within 24 hours
of being notified by the Department of its invalidation decision and have it
analyzed for the same fecal indicator using the analytical methods listed in
40
CFR 141.402(c)(2) (Federal
Register 11-8-2006 edition). The Department may extend the 24-hour time limit
on a case-by-case basis if the system cannot collect the source water sample
within 24 hours due to circumstances beyond its control. In the case of an
extension, the Department will specify how much time the system has to collect
the sample.
(5) Sampling
location.
(a) Any ground water source sample
required under R.61-58.16.E(1) must be collected at a location prior to any
treatment of the groundwater source unless the Department approves a sampling
location after treatment.
(b) If
the system's configuration does not allow for sampling at the well itself, the
system may collect a sample at a Department-approved location to meet the
requirements under R.61-58.16.E(1) if the sample is representative of the water
quality of that well.
(6) New sources. If directed by the
Department, a ground water system that places a new ground water source into
service after November 30, 2009, must conduct assessment source water
monitoring under R.61-58.16.E(2). If directed by the Department, the system
must begin monitoring before the ground water source is used to provide water
to the public.
(7) Public
notification. A ground water system with a ground water source sample collected
under R.61-58.16.E(1) or (2) that is fecal indicator positive and that is not
invalidated under R.61-58.16.E(4), including consecutive systems served by the
ground water source, must conduct public notification under R.61-58.6.E(2).
(8) Monitoring violations. Failure to meet
the requirements of R.61-58.16.E(1) through (6) is a monitoring violation and
requires the ground water system to provide public notification under
R.61-58.6.E(4).