Current through Register Vol. 48, No. 9, September 27, 2024
(1) Sample
Location, Collection Methods, and Number of Samples
(a) A water system that fails to meet the
lead or copper action level on the basis of tap samples collected in accordance
with R.61-58.11.H above, shall collect
lead and copper source water samples in accordance with the requirements
regarding sample location, number of samples, and collection methods.
(i) Groundwater systems shall take a minimum
of one (1) sample at every entry point to the distribution system after any
application of treatment or in the distribution system at a point which is
representative of each source after treatment (hereafter called a sampling
point). The system shall take one (1) sample at the same sampling point unless
conditions make another sampling point more representative of each source or
treatment plant.
(ii) Surface water
systems shall take a minimum of one sample at every entry point to the
distribution system after any application of treatment or in the distribution
system at a point which is representative of each source after treatment
(hereafter called a sampling point). The system shall take each sample at the
same sampling point unless conditions make another sampling point more
representative of each source or treatment plant.
NOTE: For the purposes of this paragraph, surface water systems
include systems with a combination of surface and ground sources.
(iii) if a system draws water from
more than one source and the sources are combined before distribution, the
system must sample at an entry point to the distribution system during periods
of normal operating conditions (i.e., when water is representative of all
sources being used).
(iv) The
Department may reduce the total number of samples which must be analyzed by
allowing the use of compositing. Compositing of samples must be done by
certified laboratory personnel. Composite samples from a maximum of five (5)
samples are allowed, provided that if the lead concentration in the composite
sample is greater than or equal to 0.001 mg/L or the copper concentration is
greater than or equal to 0.160 mg/L, then either:
(A) A follow-up sample shall be taken and
analyzed within fourteen (14) days at each sampling point included in the
composite; or
(B) If duplicates of
or sufficient quantities from the original samples from each sampling point
used in the composite are available, the system may use these instead of
resampling.
(b) Where the results of sampling indicate an
exceedance of maximum permissible source water levels established under
R.61-58.11.E(2)(d)
above, the Department may require that one additional sample be collected as
soon as possible after the initial sample was taken (but not to exceed two
weeks) at the same sampling point. If a Department- required confirmation
sample is taken for lead or copper, then the results of the initial and
confirmation sample shall be averaged in determining compliance with the
Department- specified maximum permissible levels. Any sample value below the
detection limit shall be considered to be zero. Any value above the detection
limit but below the PQL shall either be considered as the measured value or be
considered one-half the PQL.
(2) Monitoring Frequency After System Exceeds
Tap Water Action Level
Any system which exceeds the lead or copper action level at the
tap for the first time or for the first time after an addition of a new source
or installation of source water treatment required under R.61-58.11.E(2)(b)
shall collect one (1) source water sample from each entry point to the
distribution system no later than six (6) months after the end of the tap
sampling period during which the lead or copper action level was exceeded. For
tap sampling periods that are annual or less frequent, the end of the tap
sampling period is September 30 of the calendar year in which the sampling
occurs, or if the Department has established an alternate monitoring period,
the last day of that period. If the Department determines that source water
treatment is not required under R.61-58.11.E(2)(b),
the Department may waive source water monitoring, for any subsequent lead or
copper action level exceedance at the tap, in accordance with the requirements
in paragraphs (2)(a)(i) through (iii) of this section.
(a) The Department may waive source water
monitoring for lead or copper action level exceedance at the tap under the
following conditions:
(i) The water system has
already conducted source water monitoring following a previous action level
exceedance;
(ii) The Department has
determined that source water treatment is not required; and
(iii) The system has not added any new water
sources.
(b)
[Reserved]
(3)
Monitoring Frequency After Installation of Source Water Treatment and Addition
of New Source
(a) Any system which installs
source water treatment pursuant to R.61-58.11.E(1)(c)
shall collect one (1) source water sample from each entry point to the
distribution system during two (2) consecutive six (6)-month monitoring periods
by the deadline specified in R.61-58.11.E(1)(d).
(b) Any system which adds a new source shall
collect one (1) source water sample from each entry point to the distribution
system until the system demonstrates that finished drinking water entering the
distribution system has been maintained below the maximum permissible lead and
copper concentrations specified by the Department in R.61-58.11.E(2)(d)
or the Department determines that source water treatment is not
needed.
(4) Monitoring
Frequency after Department Specifies Maximum Permissible Source Water Levels
(a) A system shall monitor at the frequency
specified in paragraphs (4)(a) and (b) of this section, in cases where the
Department specifies maximum permissible source water levels under
R.61-58.11.E(2)(d).
(i) A water system using only groundwater
shall collect samples once during the three (3) year compliance period (as that
term is defined in R.61-58.B, Definitions) in effect
when the applicable Department determination under paragraph (4)(a) of this
section is made. Such systems shall collect samples once during each subsequent
compliance period. Triennial samples shall be collected every third calendar
year.
(ii) A water system using
surface water (or a combination of surface and groundwater) shall collect
samples once during each calendar year, the first annual monitoring period to
begin during the year in which the applicable Department determination is made
under paragraph (4)(a) of this section.
(b) A system is not required to conduct
source water sampling for lead and/or copper if the system meets the action
level for the specific contaminant in tap water samples during the entire
source water sampling period applicable to the system under paragraph (4)(a)(i)
or (ii) of this section.
(5) Reduced Monitoring Frequency
(a) A water system using only ground water
may reduce the monitoring frequency for lead and copper in source water to once
during each nine (9)-year compliance cycle (as that term is defined in
R.61-58.B, Definitions) provided
that the samples are collected no later than every ninth calendar year and if
the systems meets the following criteria:
(i)
The system demonstrates that finished drinking water entering the distribution
system has been maintained below the maximum permissible lead and copper
concentrations specified by the Department in R.61-58.11.E(2)(d),
during at least three (3) consecutive monitoring periods under paragraph (4)(a)
of this section; or
(ii) The
Department has determined that source water treatment is not needed and the
system demonstrates that, during at least three consecutive compliance periods
in which sampling was conducted under paragraph (4)(a) of this section, the
concentration of lead in source water was less than or equal to 0.005 mg/L and
the concentration of copper in source water was less than or equal to 0.65
mg/L.
(b) A water system
using surface water (or a combination of surface and ground waters) may reduce
the monitoring frequency in paragraph (4)(a) of this section to once during
each nine (9)-year compliance cycle (as that term is defined in R.61-58.B, Definitions) provided
that the samples are collected no later than every ninth calendar year and if
the system meets the following criteria:
(i)
The system demonstrates that finished drinking water entering the distribution
system has been maintained below the maximum permissible lead and copper
concentrations specified by the Department in Section E(2)(d) above, for at
least three (3) consecutive years; or
(ii) The Department has determined that
source water treatment is not needed and the system demonstrates that, during
at least three (3) consecutive years, the concentration of lead in source water
was less than or equal to 0.005 mg/L and the concentration of copper in source
water was less than or equal to 0.65 mg/L.
(c) A water system that uses a new source of
water is not eligible for reduced monitoring for lead and/or copper until
concentrations in samples collected from the new source during three
consecutive monitoring periods are below the maximum permissible lead and
copper concentrations specified by the Department in R.61-58.11.E(1)(e).