Rhode Island Code of Regulations
Title 216 - Department of health
Chapter 50 - Environmental health
Subchapter 15 - HEALTHY ENVIRONMENT
Part 2 - Radon Control
Section 216-RICR-50-15-2.13 - Requirements for Radon Analytical Services
Universal Citation: 216 RI Code of Rules 50 15 2.13
Current through September 18, 2024
2.13.1 Certification Requirements
A. In
addition to the general requirements for all applicants, as listed in §
2.7 of this Part, the
applicant for a radon analytical service must:
1. Provide documentation of successful
completion of the EPA Radon Measurement Proficiency Program (RMP) for each
primary testing device for which certification is desired.
2. Identify all personnel performing analysis
or readings.
3. Provide the
Department with a copy of the quality assurance and quality control procedures
for each radon testing method established by the applicant.
4. Provide evidence of licensure as an
analytical laboratory by the RIDOH in accordance with R.I. Gen. Laws Chapter
23-16.2 entitled "Laboratories" for the following methods:
a. Alpha Track Detectors
b. Charcoal Adsorption Devices
c. Charcoal Liquid Scintillation
Devices
d. Continuous Radon
Monitors
e. Electret Ion
Chambers
2.13.2 Performance Requirements
A. Radon analytical services must maintain a
listing in the EPA Radon Measurement Proficiency Program (RMP) for each primary
testing device for which certification is requested.
B. Analytical services certified in
accordance with this Part must comply with any other applicable state and local
licensing and/or certification requirements.
Disclaimer: These regulations may not be the most recent version. Rhode Island may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.