Pennsylvania Code
Title 67 - TRANSPORTATION
Part I - Department of Transportation
Subpart A - Vehicle Code Provisions
Article IV - Licensing
Chapter 77 - EQUIPMENT AND TRAINING REQUIRED FOR ADMINISTERING CHEMICAL TESTS; TEST PROCEDURES AND ACCURACY CERTIFICATION FOR BREATH TEST DEVICES
Subchapter B - PERMANENT REGULATIONS
Section 77.26 - Periodic calibration of Type A breath test equipment
Current through Register Vol. 54, No. 44, November 2, 2024
(a) Frequency. Type A alcohol breath test equipment shall be calibrated annually within 1 year of using the breath test equipment to perform an actual alcohol breath test.
(b) Procedures for calibration testing.
Constant value | Reading | Difference (absolute value) | |
First | .10 | .103 | .003 |
Series | .10 | .104 | .004 |
.10 | .097 | .003 | |
.10 | .104 | .004 | |
.10 | .099 | .001 | |
.015 = .003 | |||
5 | |||
Second | .05 | .049 | .001 |
Series | .05 | .054 | .004 |
.05 | .048 | .002 | |
.05 | .051 | .001 | |
.05 | .052 | .002 | |
.010 = .002 | |||
5 | |||
Third | .15 | .149 | .001 |
Series | .15 | .151 | .001 |
.15 | .148 | .002 | |
.15 | .151 | .001 | |
.15 | .160 | .010 | |
.015 = .003 | |||
5 |
First series-Average deviation = .003 and .003 does not exceed .005 Second series-Average deviation = .002 and .002 does not exceed .005 Third series-Average deviation = .003 and .003 does not exceed .005 but the fifth test of the series yields a result which is more than .009 above the constant value so the device fails the calibration testing.
(c) Procedures for adjustment. Breath test equipment which fails the testing under § 77.25(b) (relating to accuracy inspection tests for Type A equipment) or subsection (b) shall be placed out of service and shall be serviced, repaired and adjusted, as necessary, by the manufacturer or its authorized representative or a person who has received comparable training or instruction prior to being placed back into service. In addition, the breath test device shall be tested under subsection (b) prior to being placed back into service.
(d) Certificate of calibration. Upon satisfactory completion of the calibration procedure, the test record shall be recorded on a certificate of calibration of a type provided or approved by the Department. The certificate of calibration shall be signed and dated by the individual who performed the calibration procedure and shall be retained for a period of 3 years from the date of the calibration procedure.
This section cited in 67 Pa. Code § 77.25 (relating to accuracy inspection tests for Type A equipment).