Current through Register Vol. 54, No. 44, November 2, 2024
(a) In addition to the requirements of §
252.401 (relating to basic
requirements), environmental laboratories performing testing or analysis in the
area of microbiology shall comply with this section.
(b) When the method selected by an
environmental laboratory in accordance with §
252.307 (relating to methodology)
contains more stringent requirements than the requirements of this section, the
environmental laboratory shall follow the more stringent requirements contained
in the method.
(c) The following
pieces of equipment shall be maintained according to this subsection:
(1)
Autoclave.
(i) An environmental laboratory shall use
autoclaves that meet specified temperature tolerances of the method. Pressure
cookers may not be used.
(ii) A
continuous temperature-recording device or a maximum-temperature-registering
thermometer shall be used during each autoclave cycle.
(iii) An environmental laboratory shall
verify the sterilization capability of each autoclave by utilizing appropriate
biological indicators (for example, spore strips or ampoules) once a month.
Records of biological indicator tests shall be maintained in a laboratory
notebook and include the autoclave identification, date, incubation time and
temperature, results and initials of the responsible individual.
(iv) An environmental laboratory shall verify
the mechanical timing device, if used, for each autoclave every 3 months.
Records of mechanical timer verification shall be maintained in a laboratory
notebook and include the autoclave identification, date, mechanical timing
device time, actual time and initials of the responsible individual. Correction
factors shall be documented and used.
(v) Autoclaves shall be properly cleaned and
maintained. Copies of service contracts or internal maintenance protocols and
maintenance records shall be kept.
(vi) Required times for autoclaving items at
121°C are set forth in this subparagraph. The following items must be at
temperature for the required amount of time. Except for membrane filters and
pads and carbohydrate-containing media, indicated times are minimum times and
may necessitate adjustment depending upon volumes, containers and loads. For
autoclave runs that include membrane filters and pads and media, the total
cycle time may not exceed 45 minutes. Autoclaved membrane filters and pads and
media shall be removed immediately after completion of the autoclave cycle.
(A) Membrane filters and pads | 10
minutes |
(B) Carbohydrate-containing media
| 12-15 minutes |
(C) Contaminated test materials | 30
minutes |
(D) Membrane filtration units | 15
minutes |
(E) Sample containers | 15 minutes |
(F) Individual glassware | 15 minutes
|
(G) Dilution water | 15 minutes |
(H) Rinse water | 15-30 minutes |
(vii) Records of each autoclave run shall be
maintained in a laboratory notebook and include the date, contents,
sterilization time and temperature, total cycle time (recorded as time in and
time out) and initials of the responsible individual.
(viii) If an autoclave cycle fails to meet
any requirement, corrective action shall be documented. Media may not be
reautoclaved.
(2)
Hot air oven.
(i) An
environmental laboratory shall maintain a thermometer, graduated in 10°C
increments or less with the bulb placed in sand, in each hot air
oven.
(ii) An environmental
laboratory shall verify the sterilization capability of each hot air oven by
utilizing appropriate biological indicators (for example, spore strips) once a
month. Records of biological indicator tests shall be maintained in a
laboratory notebook and include the hot air oven identification, date,
incubation time and temperature, results and initials of the responsible
individual.
(iii) An environmental
laboratory shall sterilize items in a hot air oven maintaining a temperature of
170°-180°C for a minimum of 2 hours. Only dry items may be sterilized
in a hot air oven.
(iv) Records of
each hot air oven operation shall be maintained and include the date, contents,
sterilization time and temperature, and initials of the responsible
individual.
(3)
Inoculating equipment.
(i)
An environmental laboratory shall use appropriate sterile inoculating
equipment.
(ii) Metal loops and
needles must be made of nickel alloy or platinum.
(iii) Wooden applicator sticks must be
sterilized using dry heat.
(iv) For
oxidase tests, nickel alloy loops may not be used.
(4)
Membrane filtration
equipment.
(i) Membrane filtration
funnels must be stainless steel, glass, porcelain or autoclaveable or
presterilized plastic. Membrane filtration funnels may not be scratched or
corroded and may not leak.
(ii)
Membrane filtration units shall be sterilized before the beginning of a
filtration series. A filtration series ends when 30 minutes or longer elapses
after a sample is filtered.
(iii)
Forceps must be blunt and smooth-tipped without corrugations on the inner sides
of tips.
(iv) Membrane filters must
meet the following requirements:
(A) Membrane
filters must be made of cellulose ester, white, grid marked, 47 mm diameter and
0.45-µm pore size unless otherwise specified by the method.
(B) Membrane filters must be either purchased
presterilized or autoclaved for 10 minutes at 121°C before use. Membrane
filters may not be brittle or distorted.
(C) Membrane filters must be approved (based
upon manufacturer data from tests for toxicity, recovery, retention and absence
of growth-promoting substances) for the specified analysis for which they are
to be used.
(v) An
environmental laboratory using an ultraviolet sanitation lamp to sanitize
filtration funnels between successive filtrations shall test the ultraviolet
sanitation lamp every 3 months for effectiveness with an appropriate UV light
meter or by plate count agar spread plates. Records of ultraviolet lamp tests
shall be maintained and bulbs shall be replaced if output is less than 70% of
original for light tests or if count reduction is less than 99% for a plate
containing 200 to 300 organisms.
(5)
Culture dishes.
(i) Culture dishes must be presterilized
plastic or sterilizable glass and of appropriate size for the method.
(ii) Stainless steel canisters, aluminum
canisters or a wrap of heavy aluminum foil or char-resistant paper, shall be
used for autoclave sterilization of glass culture dishes.
(iii) Loose-lid culture dishes shall be
incubated in a tight fitting container containing a moistened paper
towel.
(iv) Opened packs of
disposable culture dishes shall be resealed between use periods.
(6)
Culture tubes and
closures. Culture tubes and containers must be of sufficient size to
contain medium and sample without being more than three quarters full. Tube
closures must be stainless steel, aluminum, plastic or a screw cap with a
nontoxic liner.
(7)
Pipettes.
(i) Pipettes must
have legible markings and may not be chipped or etched and must be accurate to
within 2.5% tolerance.
(ii)
Stainless steel canisters, aluminum canisters or a wrap of heavy aluminum foil
or char-resistant paper shall be used for autoclave sterilization of
pipettes.
(iii) Opened packs of
disposable sterile pipettes shall be resealed between use periods.
(8)
Sample
containers.
(i) Sample containers
must be sterile plastic bags or wide-mouth plastic or noncorrosive glass
bottles with nonleaking ground glass stoppers or caps with nontoxic liners that
can withstand repeated sterilization. Sample containers must be capable of
holding sufficient volume of sample for all required tests while maintaining
adequate air space for mixing.
(ii)
Glass stoppers must be covered with aluminum foil or char-resistant paper for
sterilization.
(iii) Glass and
plastic bottles that have not been presterilized shall be sterilized by
autoclaving. Glass bottles may be sterilized by dry heat. Empty containers
shall be moistened with several drops of water prior to autoclaving.
(9)
Plastic and glassware
washing procedure.
(i) Prior to the
initial use of a lot of detergent or washing procedure, an environmental
laboratory shall perform an inhibitory residue test utilizing the method
described in the currently approved editions of Standard Methods for
the Examination of Water and Wastewater (available from the American
Public Health Association, 800 I Street, NW, Washington, D.C. 20001). Records
of inhibitory residue tests shall be maintained and include the detergent
identification, date, calculations, results and initials of responsible
individual.
(ii) Washed plastic and
glassware shall be tested at least once each month for possible acid or
alkaline residue by testing at least one piece of plastic and glassware with a
suitable pH indicator such as 0.04% bromothymol blue. Records of pH tests shall
be maintained and include the date, results and identification of the
responsible individual.
(10)
Ultraviolet lamp. An
environmental laboratory shall use a 365-nm, 6-watt ultraviolet lamp in a
darkened room to view sample fluorescence.
(11)
Quanti-TrayTM Sealer.
(i) An environmental laboratory shall perform
a sealer check on each Quanti-Tray Sealer once a month by adding a dye to a
water sample and performing the sealing procedure.
(ii) Records of the sealer check shall be
maintained and include the sealer identification, date, results and initials of
responsible individual. If dye is observed outside the wells, the Quanti-Tray
Sealer may not be used.
(d) The requirements for reagent water are as
follows:
(1) An environmental laboratory
shall use reagent water in the preparation of media, solutions and
buffers.
(2) An environmental
laboratory shall demonstrate that reagent water meets the following criteria on
a monthly basis or whenever maintenance is performed on the water treatment
system or at startup after a period of nonuse longer than 1 month:
(i) Total chlorine residual must be less than
0.1 mg/L.
(ii) Conductivity must be
less than 2.0 µmhos/cm or resistance greater than 0.5 megohms at
25°C.
(iii) Heterotrophic plate
count must be less than 500 CFU/mL.
(3) An environmental laboratory shall
demonstrate that reagent water meets the following criteria every 12 months:
(i) The individual concentration of lead,
cadmium, chromium, copper, nickel and zinc must be less than 0.05
mg/L.
(ii) The total concentration
of lead, cadmium, chromium, copper, nickel and zinc must be less than 0.1
mg/L.
(iii) Except as provided in
subsection (d)(6), the bacteriological water quality test ratio must be between
0.8 and 3.0. The bacteriological water quality test shall be performed
according to the currently approved editions of Standard Methods for
the Examination of Water and Wastewater (available from the American
Public Health Association, 800 I Street, NW, Washington, D.C. 20001).
(4) The metals analyses may only
be performed by an environmental laboratory accredited under this chapter for
those fields of accreditation.
(5)
Results of the monthly and annual reagent water analysis shall be maintained
and include the date, type of test, results and initials of responsible
individual. Reagent water that does not meet the required criteria may not be
used.
(6) The bacteriological water
quality test need not be performed if the environmental laboratory can supply
documentation to show that their laboratory pure water or reagent water meets
the criteria, as specified in section 1080 of the currently approved editions
of Standard Methods for the Examination of Water and
Wastewater (available from the American Public Health Association, 800
I Street, NW, Washington, D.C. 20001), for Type I (high-quality) or Type II
(medium-quality) reagent water.
(7)
The heterotrophic plate count and bacteriological water quality test ratio
analyses described in paragraphs (2) and (3) shall be performed by an
environmental laboratory accredited under this chapter for the appropriate
field of accreditation.
(e) The requirements for dilution/rinse water
are as follows:
(1) Stock buffer solution or
peptone water shall be prepared as specified in the currently approved editions
of Standard Methods for the Examination of Water and
Wastewater (available from the American Public Health Association, 800
I Street, NW, Washington, D.C. 20001).
(2) Stock buffers shall be autoclaved or
filter-sterilized. Stock buffers shall be refrigerated and must be free from
turbidity.
(3) Dilution/rinse water
solutions shall be prepared as specified in the currently approved editions of
Standard Methods for the Examination of Water and Wastewater
(available from the American Public Health Association, 800 I Street, NW,
Washington, D.C. 20001).
(f) The requirements for media are as
follows:
(1) An environmental laboratory
shall use dehydrated or commercially manufactured prepared media. Dehydrated
media shall be stored in a cool, dry location. Caked or discolored dehydrated
media shall be discarded.
(2) An
environmental laboratory that prepares media from dehydrated stock shall follow
method specifications.
(3) Media
may not be reautoclaved.
(4) After
preparation, media shall be stored and maintained as follows:
(i) Stored away from sources of direct
light.
(ii) Prepared plates shall
be stored in sealed plastic bags or containers.
(iii) Each bag, container or rack of broth or
agar media shall be labeled with the date prepared or expiration
date.
(iv) Fermentation media
stored in a refrigerator shall be brought to room temperature before use. Media
that shows growth or false positive results may not be used.
(v) Prepared liquid media shall be discarded
if evaporation exceeds 10% of the original volume.
(vi) Poured agar plates and broth in tubes,
bottles or flasks with loose-fitting closures shall be discarded if not used
within 2 weeks of sterilization unless otherwise specified by the
method.
(vii) Broth in tightly
closed screw-cap tubes, bottles or flasks shall be discarded if not used within
3 months of sterilization unless otherwise specified by the method.
(g) An environmental
laboratory shall demonstrate that the filtration equipment and filters, sample
containers, media and reagents have not been contaminated through improper
handling or preparation, inadequate sterilization or environmental exposure as
follows:
(1) A sterility blank shall be
analyzed for each lot of preprepared, ready-to-use medium and for each batch of
medium prepared in the laboratory prior to first use of the medium. Records
shall be maintained and include media identification, date and time of the
start and end of incubation, results and initials of the responsible
individuals. If sterility blank indicates contamination, the media may not be
used.
(i) For chromogenic/fluorogenic media,
add single-strength media to sterile reagent water and incubate at the
appropriate temperature and time.
(ii) For all other media, incubate
uninoculated, single-strength at the appropriate temperature and
time.
(2) For each
reusable membrane filtration unit used during a filtration series, the
laboratory shall prepare at least one sterility blank at the beginning and at
the end of the series. A series is considered ended when more than 30 minutes
elapses between filtrations. The laboratory shall insert a sterility blank
after every ten sample aliquots filtered through each membrane filtration unit
or sanitize filtration units by UV light after each sample filtration in
addition to the regular rinsing procedure. Records of sterility blank results
shall be maintained in the same manner as the associated sample and include the
date and time of the start and end of the incubation, results and initials of
the responsible individuals. If sterility blanks indicate contamination, the
laboratory must treat each affected sample according to program
requirements.
(3) For presterilized
single use filtration funnel units, a sterility check shall be performed on one
funnel unit per lot.
(4) Sterility
checks on sample containers shall be performed on at least one container for
each lot of purchased, presterilized containers with an appropriate
nonselective growth media. For containers prepared and sterilized in the
laboratory, a sterility check shall be performed on one container per
sterilized batch with an appropriate nonselective growth media. Results shall
be maintained and include sample container identification, date and time of the
start and end of incubation, results and initials of responsible individuals.
If sample container sterility check indicates contamination, the affected
sample container may not be used.
(5) A sterility blank shall be performed on
each batch of dilution/rinse water prepared in the laboratory and on each batch
of preprepared, ready-to-use dilution water with an appropriate nonselective
growth media. The concentration of media shall be single strength after
addition of dilution water. Results shall be maintained and include
dilution/rinse water identification, date and time of the start and end of
incubation, results and initials of the responsible individuals. If
dilution/rinse water sterility check indicates contamination, the affected
dilution water may not be used.
(6)
At least one filter from each new lot of membrane filters shall be checked for
sterility with an appropriate nonselective growth media. Results shall be
maintained and include membrane filter identification, date and time of the
start and end of incubation, results and initials of the responsible
individuals. If the membrane filter sterility check indicates contamination,
the affected membrane filters may not be used.
(7) Sterility checks on
Quanti-TrayTM sample trays shall be performed on at
least one sample tray for each lot of purchased presterilized sample trays with
an appropriate nonselective growth media. Results shall be maintained and
include sample tray identification, date and time of the start and end of
incubation, results and initials of the responsible individuals. If the sample
tray sterility check indicates contamination, the affected lot of sample trays
may not be used.
(h) The
requirements for positive and negative culture control checks are as follows:
(1) Each preprepared, ready-to-use lot of
medium and each batch of medium prepared in the laboratory shall be tested by
the laboratory with at least one pure culture of a known positive reaction
prior to first use of the medium. Records shall be maintained and include the
date and time of the start and end of incubation, media lot or batch number,
type of media, positive culture control organism identification, results and
initials of the responsible individuals. If positive culture control checks do
not meet expected results, the affected media may not be used.
(2) Each preprepared, ready-to-use lot of
selective medium and each batch of selective medium prepared in the laboratory
shall be tested by the laboratory with at least one pure culture of a known
negative reaction prior to first use of the medium. Records shall be maintained
and include the date and time of the start and end of incubation, media lot or
batch number, type of media, negative culture control organism identification,
results and initials of the responsible individuals. If negative culture
control checks do not meet expected results, the affected media may not be
used.
(3) An environmental
laboratory shall use stock positive and negative culture controls that are
known and traceable to a recognized National collection. Documentation of
traceability shall be maintained.
(4) Stock positive and negative culture
controls shall be discarded after the manufacturer's expiration date.
(5) Culture controls may be single use or
cultures maintained by the laboratory using a documented procedure that
maintains the purity and viability of the organisms.
(6) For cultures maintained by the
laboratory, the following criteria must be met:
(i) Reference control cultures may be revived
and subcultured once to provide reference stocks.
(ii) Reference stocks shall be preserved
using a method which maintains the characteristics of the organism strains. If
reference stocks are thawed, they may not be refrozen and reused.
(iii) Working stocks shall be prepared from
reference stocks for routine laboratory work.
(iv) If the laboratory sequentially cultures
working stocks, the laboratory shall prepare a second working stock. Sequential
culturing may not be performed from a working stock that has been used for
routine laboratory work.
(v)
Working stocks may not be used for more than 30 days.
(vi) Working stocks may not be sequentially
cultured more than five times and may not be subcultured to replace reference
stocks.
(vii) Secondary working
stocks shall be used to prepare sequential working stocks.
(7) Positive and negative controls must be
processed under the same conditions and using the same equipment as routine
environmental samples, including all steps of the preparation and analytical
procedure.
(i) For test
methods that specify colony counts, duplicate counts shall be performed monthly
on one positive sample for each month that the test is performed. If the
laboratory has two or more analysts, each analyst shall count typical colonies
on the same plate. Counts may not differ by more than 10%. In an environmental
laboratory with only one analyst, the analyst shall count the same plate twice.
Counts may not differ by more than 5%.
(j) Quality control checks, including
sterility checks and positive and negative controls, shall be conducted after
the laboratory receives the material or supply and before or during first use.
These checks shall be performed by an environmental laboratory accredited under
this chapter and utilizing the same supplies, reagents and media to be used
during laboratory analysis of environmental samples. Certificates of analysis
from a manufacturer may not be used to demonstrate compliance with the
requirements of this subsection.
(k) Records of all equipment, reference
materials, reagents, media and supplies shall be maintained in accordance with
§
252.306 (relating to equipment,
supplies and reference materials).
The provisions of this §252.404 amended under
27 Pa.C.S. §§
4103(a),
4104 and
4105; and section 1920-A
of The Administrative Code of 1929 (71 P.S. §
510-20).