Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 240 - RADON CERTIFICATION
Subchapter G - QA REQUIREMENTS
Section 240.606 - QA requirements for laboratories
Universal Citation: 25 PA Code ยง 240.606
Current through Register Vol. 54, No. 44, November 2, 2024
(a) CRMs for laboratories.
(1)
Calibration. Each
Department-listed CRM must have a current calibration. To have a current
calibration, the CRM shall be calibrated in a Department-approved calibration
facility within 1 year from the date of the previous calibration and when
alterations or repairs are made to the CRM. A current calibration certificate
shall be retained for each monitor. Analysis may not be performed on a monitor
that was not calibrated during any portion of the testing period.
(2)
Data collection log. CRM
data shall be tracked on a form that contains all of the following:
(i) The CRM serial number.
(ii) The exposure dates and times.
(iii) The test result.
(iv) The address of the building
tested.
(v) The test location in
the building.
(vi) The name of the
tester who placed the CRM.
(vii)
The name of the tester who retrieved the CRM.
(viii) The calibration, repair and Department
listing dates.
(b) CWLM for laboratories.
(1)
Calibration. Each
Department-listed CWLM must have a current calibration. To have a current
calibration, the CWLM shall be calibrated in a Department-approved calibration
facility within 1 year from the date of the previous calibration and when
alterations or repairs are made to the CWLM. A current calibration certificate
shall be retained for each monitor. Analysis may not be performed on a monitor
that was not calibrated during any portion of the testing period.
(2)
Data collection log.
CWLM data shall be tracked on a form that contains all of the following:
(i) The CWLM serial number.
(ii) The exposure dates and times.
(iii) The test result.
(iv) The address of the building
tested.
(v) The test location in
the building.
(vi) The name of the
tester who placed the CWLM.
(vii)
The name of the tester who retrieved the CWLM.
(viii) The calibration, repair and Department
listing dates.
(c) Electret ion chamber for laboratory analysis.
(1)
Calibration. Each Department-listed electret reader shall have
a current calibration. To have a current calibration, the electret reader shall
be calibrated in a Department-approved calibration facility within 1 year from
the date of the previous calibration and when alterations or repairs are made
to the electret reader. Each electret reader shall be calibrated simultaneously
with its corresponding reference electret's recertification.
(2)
Voltmeter routine instrument
checks.
(i) Proper operation of the
surface voltmeter shall be monitored following the manufacturer's procedures
for zeroing the voltmeter and analyzing the reference electrets.
(ii) A voltage reading of a reference
electret difference of more than 2 volts from its specified value shall be
considered a wrong reading and corrective action shall be taken.
(iii) If the voltmeter displays more than
(±) 3 volts, corrective action shall be taken.
(iv) Voltmeter checks shall be conducted at
least once each week while the voltmeter is in use and shall be documented.
Documentation of routine instrument checks must include all of the following:
(A) The reader serial number.
(B) The date of analysis.
(C) Zero value.
(D) The reference electret values.
(E) Corrective actions performed.
(3)
Known
exposure measurements (spikes).
(i)
Spikes shall be conducted at a rate of 3 for each 100 test devices deployed,
with a minimum of 3 spikes for each certification year when tests were
conducted in the certification year, and with a maximum of 6 spikes each
month.
(ii) Spikes shall be
analyzed in the same manner as all other testing.
(iii) Spikes shall be monitored using a means
control chart. The means control chart must be established as follows:
(A) Using an RPE value of plus and minus 10%,
which corresponds to the 1 sigma level.
(B) A warning level of the RPE of plus and
minus 20%, which corresponds to the 2 sigma warning level.
(C) Control limits of the RPE of plus and
minus 30%, which correspond to the 3 sigma control level.
(iv) Each RPE value shall be plotted on the
means control chart within 1 week of return of the device from the radon
chamber. If the RPE value is outside the 3 sigma control level, all
measurements shall cease until the problem is evaluated and corrected. All
evaluations shall be documented.
(v) In addition to the means control chart,
all spikes shall be documented on a form that contains all of the following:
(A) The radon chamber name.
(B) The electret serial numbers.
(C) The RV from the radon chamber.
(D) The measured spike value or
values.
(E) The individual RPE
results.
(F) The certification year
beginning date and end date.
(G)
The exposure dates.
(4)
Duplicate measurements.
(i) Duplicates shall be made in at least 10%
of the total number of test devices deployed each month, or 50 each month,
whichever is smaller.
(ii) The RPD
shall be calculated for all duplicate results with an average of greater than
or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor
duplicate precision:
(A) One for duplicates
when the average is greater than or equal to 4.0 pCi/L.
(B) One for duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.
(iii) Each RPD value shall be
plotted on the control chart within 1 week of performing the duplicate
measurement.
(iv) The RPD shall be
tracked using control charts from "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2
and B-3.
(v) For duplicates when
the average is greater than or equal to 4.0 pCi/L, all of the following apply:
(A) The control level shall be set at an RPD
of 14%.
(B) The warning level shall
be set at an RPD of 28%.
(C) The
control limit shall be set at an RPD of 36%.
(vi) For duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the
following apply:
(A) The control level shall
be set at an RPD of 25%.
(B) The
warning level shall be set at an RPD of 50%.
(C) The control limit shall be set at an RPD
of 67%.
(vii) If the
plotted RPD result falls outside of the control limit, the measurements shall
cease until the problem is identified and corrected.
(viii) If the plotted RPD result falls
outside of the warning level, "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5,
shall be used to determine the action to be taken.
(ix) Documentation of duplicates must include
all of the following:
(A) The device serial
numbers.
(B) The exposure
dates.
(C) Each duplicate
measurement result.
(D) The RPD
results.
(d) AC and LS.
(1)
Calibration. All AC or
LS laboratory systems shall be calibrated at least once every 12 months, when
alterations or repairs are made to the system, or when a new batch of charcoal
is received. This requires a determination of calibration factors for AC and LS
devices by the exposure of these devices to a known concentration of radon in a
Department-approved radon chamber. Calibration factors shall be determined for
a range of exposure times and humidity levels.
(2)
Laboratory control
devices. The laboratory background level for each batch of AC and LS
devices shall be established by each laboratory. Laboratories shall measure the
background of at least 5% of unexposed AC and LS devices that have been
processed according to their standard operating procedures (laboratory
blanks).
(3)
Routine
counting system checks. Daily counting of a reference source shall be
performed and documented. The characteristics of the check source (geometry,
type of radiation emitted, and the like) must be similar to the samples to be
analyzed. The count rate of the check sources must be high enough to yield
reliable counting statistics in a short period of time, such as 1,000 to 10,000
counts per minute, to provide a maximum random uncertainty of 5%.
(4)
Known exposure measurements
(spikes).
(i) Spikes shall be
conducted at a rate of 3 for each 100 test devices deployed, with a minimum of
3 spikes for each certification year when tests were conducted in the
certification year, and with a maximum of 6 spikes each month.
(ii) Spikes shall be analyzed in the same
manner as all other testing.
(iii)
Spikes shall be monitored using a means control chart. The means control chart
must be established as follows:
(A) Using an
RPE value of plus and minus 10%, which corresponds to the 1 sigma
level.
(B) A warning level of the
RPE of plus and minus 20%, which corresponds to the 2 sigma warning
level.
(C) Control limits of the
RPE of plus and minus 30%, which correspond to the 3 sigma control
level.
(iv) Each RPE
value shall be plotted on the means control chart within 1 week of receiving
the result from the laboratory. If the RPE value is outside the 3 sigma control
level, all measurements shall cease until the problem is evaluated and
corrected. All evaluations shall be documented.
(v) In addition to the means control chart,
all spikes shall be documented on a form that contains all of the following:
(A) The radon chamber name.
(B) The device serial numbers.
(C) The RV from the radon chamber.
(D) The measured spike value or
values.
(E) The individual RPE
results.
(F) The certification year
beginning date and end date.
(G)
The exposure dates.
(5)
Duplicate measurements.
(i) Duplicates shall be made in at least 10%
of the total number of test devices deployed each month, or 50 each month,
whichever is smaller.
(ii) The RPD
shall be calculated for all duplicate results with an average of greater than
or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor
duplicate precision:
(A) One for duplicates
when the average is greater than or equal to 4.0 pCi/L.
(B) One for duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.
(iii) Each RPD value shall be
plotted on the control chart within 1 week of performing the duplicate
measurement.
(iv) The RPD shall be
tracked using control charts from "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2
and B-3.
(v) For duplicates when
the average is greater than or equal to 4.0 pCi/L, all of the following apply:
(A) The control level shall be set at an RPD
of 14%.
(B) The warning level shall
be set at an RPD of 28%.
(C) The
control limit shall be set at an RPD of 36%.
(vi) For duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the
following apply:
(A) The control level shall
be set at an RPD of 25%.
(B) The
warning level shall be set at an RPD of 50%.
(C) The control limit shall be set at an RPD
of 67%.
(vii) If the
plotted RPD result falls outside of the control limit, the measurements shall
cease until the problem is identified and corrected.
(viii) If the plotted RPD result falls
outside of the warning level, "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, shall be used
to determine the action to be taken.
(ix) Documentation of duplicates must include
all of the following:
(A) The device serial
numbers.
(B) The exposure
dates.
(C) Each duplicate
measurement result.
(D) The RPD
results.
(e) ATs.
(1)
Calibration. All AT
laboratory systems shall be calibrated at least once every 12 months, when
alterations or repairs are made to the system, or when each new batch or sheet
of detector material is received. This requires a determination of calibration
factors for AT devices by the exposure of these devices to different
concentrations of radon in a Department-approved radon chamber.
(2)
Laboratory control
detectors. Laboratory control detectors for each batch of ATs shall be
established and documented. Each laboratory shall measure the background of a
statistically significant number of unexposed ATs. The laboratory control
background value shall be subtracted from the field readings to produce a final
result.
(3)
Known exposure
measurements (spikes).
(i) Spikes
shall be conducted at a rate of 3 for each 100 test devices deployed, with a
minimum of 3 spikes for each certification year when tests were conducted in
the certification year, and with a maximum of 6 spikes each month.
(ii) Spikes shall be analyzed in the same
manner as all other testing. The RV of a spike may not be revealed to the
laboratory prior to analysis.
(iii)
Spikes shall be monitored using a means control chart. The means control chart
must be established as follows:
(A) Using an
RPE value of plus and minus 10%, which corresponds to the 1 sigma
level.
(B) A warning level of the
RPE of plus and minus 20%, which corresponds to the 2 sigma warning
level.
(C) Control limits of the
RPE of plus and minus 30%, which correspond to the 3 sigma control
level.
(iv) Each RPE
value shall be plotted on the means control chart within 1 week of receiving
the result from the laboratory. If the RPE value is outside the 3 sigma control
level, all measurements shall cease until the problem is evaluated and
corrected. All evaluations shall be documented.
(v) In addition to the means control chart,
all spikes shall be documented on a form that contains all of the following:
(A) The radon chamber name.
(B) The device serial numbers.
(C) The RV from radon chamber.
(D) The measured spike value or
values.
(E) The individual RPE
results.
(F) The certification year
beginning date and end date.
(G)
The exposure dates.
(4)
Duplicate measurements.
(i) Duplicates shall be made in at least 10%
of the total number of test devices deployed each month, or 50 each month,
whichever is smaller.
(ii) The RPD
shall be calculated for all duplicate results with an average of greater than
or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor
duplicate precision:
(A) One for duplicates
when the average is greater than or equal to 4.0 pCi/L.
(B) One for duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.
(iii) Each RPD value shall be
plotted on the control chart within 1 week of performing the duplicate
measurement.
(iv) The RPD shall be
tracked using control charts from "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2
and B-3.
(v) For duplicates when
the average is greater than or equal to 4.0 pCi/L, all of the following apply:
(A) The control level shall be set at an RPD
of 14%.
(B) The warning level shall
be set at an RPD of 28%.
(C) The
control limit shall be set at an RPD of 36%.
(vi) For duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the
following apply:
(A) The control level shall
be set at an RPD of 25%.
(B) The
warning level shall be set at an RPD of 50%.
(C) The control limit shall be set at an RPD
of 67%.
(vii) If the
plotted RPD result falls outside of the control limit, the measurements shall
cease until the problem is identified and corrected.
(viii) If the plotted RPD result falls
outside of the warning level, "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5,
shall be used to determine the action to be taken.
(ix) Documentation of duplicates shall
include all of the following:
(A) The device
serial numbers.
(B) The exposure
dates.
(C) Each duplicate
measurement result.
(D) The RPD
results.
Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.