Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 240 - RADON CERTIFICATION
Subchapter G - QA REQUIREMENTS
Section 240.605 - QA requirements for testing using secondary devices
Universal Citation: 25 PA Code ยง 240.605
Current through Register Vol. 54, No. 44, November 2, 2024
(a) CRMs for secondary testers.
(1)
Calibration. Each Department-listed CRM must have a current
calibration. To have a current calibration, the CRM shall be calibrated in a
Department-approved calibration facility within 1 year from the date of the
previous calibration and when alterations or repairs are made to the CRM. A
current calibration certificate shall be retained for each monitor.
(2)
Check source counting.
For a CRM with a check source, check source counting shall be documented and
completed with that check source prior to each test.
(3)
Routine instrument
checks. Before and after each measurement, the CRM shall be checked
according to the manufacturer's instructions. For each check, all of the
following shall be verified:
(i) The correct
input parameters and the unit's clock or timer are set properly.
(ii) The pump's flow rates are within the
range of the manufacturer's specifications.
(4)
Data collection log.
(i) CRM data shall be tracked on a form that
contains all of the following:
(A) The CRM
serial number.
(B) The exposure
dates and times.
(C) The test
result.
(D) The address of the
building tested.
(E) The test
location in the building.
(F) The
name of the tester who placed the CRM.
(G) The name of the tester who retrieved the
CRM.
(H) The calibration, repair
and Department listing dates.
(ii) For a CRM without a check source, the
data collection log must also contain all of the following intercomparison
measurement information:
(A) The
intercomparison device serial number.
(B) The RPE value or RPD value.
(C) The intercomparison measurement
result.
(5)
Intercomparison measurements. An intercomparison measurement
shall be performed for each CRM without a check source.
(i) Intercomparison measurements shall be
made at least every tenth test with another Department-listed passive device
that is analyzed by a Department-certified laboratory or with another CRM with
a hard copy printout. The intercomparison measurements shall be distributed
systematically throughout the entire population of test locations. Original
printouts or Department-certified laboratory results, or both, shall be kept
for each intercomparison. Each intercomparison measurement must be performed
with the devices side by side for the measurement for at least 48
hours.
(ii) When performing
intercomparison measurements, the RPD shall be used to track performance. The
RPD value shall be tracked using control charts from "Protocols for Radon and
Radon Decay Product Measurements in Homes," EPA 402-R-92-003, May 1993,
Appendix B, Exhibits B-2 and B-3.
(iii) If the RPD value exceeds the control
limit, the CRM may not be used for radon measurements until the problem is
identified and corrected. If the RPD value exceeds the warning level, the
criteria in the "Protocols for Radon and Radon Decay Product Measurements in
Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5, shall be
followed.
(iv) In addition to the
control charts, intercomparison measurements shall be documented on the CRM
data collection log.
(b) CWLM for secondary testers.
(1)
Calibration. Each Department-listed CWLM must have a current
calibration. To have a current calibration, the CWLM shall be calibrated in a
Department-approved calibration facility within 1 year from the date of the
previous calibration and when alterations or repairs are made to the CWLM. A
current calibration certificate shall be retained for each monitor.
(2)
Data collection log.
(i) CWLM data shall be tracked on a form that
contains all of the following:
(A) The CWLM
serial number.
(B) The exposure
dates and times.
(C) The test
result.
(D) The address of the
building tested.
(E) The test
location in the building.
(F) The
name of the tester who placed the CWLM.
(G) The name of the tester who retrieved the
CWLM.
(H) The calibration, repair
and Department listing dates.
(ii) For CWLMs without a check source, the
data collection log must also contain all of the following intercomparison
measurement information:
(A) The
intercomparison device serial number.
(B) The RPD value.
(C) The intercomparison measurement
result.
(3)
Intercomparison measurements. An intercomparison measurement
shall be performed for all CWLM monitors without a check source.
(i) A CWLM without check source capability
shall have an informal intercomparison measurement made with another CWLM with
a hard copy printout at least every tenth test. This printout shall be retained
for each intercomparison. The intercomparison measurements shall be distributed
systematically throughout the entire population of test locations. Each
intercomparison measurement must be performed with the devices side by side for
the measurement for at least 48 hours.
(ii) Each intercomparison shall be documented
on the data collection log.
(iii)
When performing intercomparison measurements, the RPD shall be used to track
performance. The RPD value shall be tracked using control charts from
"Protocols for Radon and Radon Decay product measurements in Homes," EPA
402-R-92-003, May 1993, Appendix B, Exhibits B-2 and B-3.
(iv) If the RPD value exceeds the control
limit, the CWLM may not be used for radon measurements until the problem is
identified and corrected. If the RPD value exceeds the warning level, the
criteria in the "Protocols for Radon and Radon Decay Product Measurements in
Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5, shall be
followed.
(c) Electret ion chambers for secondary testers.
(1)
Data collection log.
Electret data shall be tracked on a form that contains all of the following:
(i) The electret serial number.
(ii) The initial voltage reading.
(iii) The final voltage reading.
(iv) The exposure dates and times.
(v) The test results.
(vi) The serial number of duplicate
electret.
(vii) The RPD
value.
(viii) The address of the
building tested.
(ix) The test
location in the building.
(x) The
name of the tester who placed the electret.
(xi) The name of the tester who retrieved the
electret.
(2)
Known exposure measurements (spikes).
(i) Spikes shall be conducted at a rate of 3
for each 100 test devices deployed, with a minimum of 3 spikes for each
certification year when tests were conducted in the certification year, and
with a maximum of 6 spikes each month.
(ii) Spikes shall be submitted to a
Department-certified laboratory labeled as QA. The RV of the spiked device may
not be revealed to the laboratory prior to analysis.
(iii) Spikes shall be monitored using a means
control chart. The means control chart must be established as follows:
(A) Using an RPE value of plus and minus 10%,
which corresponds to the 1 sigma level.
(B) A warning level of the RPE of plus and
minus 20%, which corresponds to the 2 sigma warning level.
(C) Control limits of the RPE of plus and
minus 30%, which correspond to the 3 sigma control level.
(iv) Each RPE value shall be plotted on the
means control chart within 1 week of return of the device from the chamber. If
the RPE value is outside the 3 sigma control level, all measurements shall
cease until the problem is evaluated and corrected. All evaluations shall be
documented.
(v) In addition to the
means control chart, all spikes shall be documented on a form that contains all
of the following:
(A) The radon chamber
name.
(B) The electret serial
numbers.
(C) The RV from radon
chamber.
(D) The measured spike
value or values.
(E) The individual
RPE results.
(F) The certification
year beginning date and end date.
(G) The exposure dates.
(3)
Duplicate
measurements.
(i) Duplicates shall
be made in at least 10% of the total number of test devices deployed each
month, or 50 each month, whichever is smaller.
(ii) The RPD shall be calculated for all
duplicate results with an average of greater than or equal to 2.0 pCi/L. Two
control charts shall be constructed to monitor duplicate precision:
(A) One for duplicates when the average is
greater than or equal to 4.0 pCi/L.
(B) One for duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.
(iii) Each RPD value shall be
plotted on the control chart within 1 week of performing the duplicate
measurement.
(iv) The RPD shall be
tracked using control charts from "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2
and B-3.
(v) For duplicates when
the average is greater than or equal to 4.0 pCi/L, all of the following apply:
(A) The control level shall be set at an RPD
of 14%.
(B) The warning level shall
be set at an RPD of 28%.
(C) The
control limit shall be set at an RPD of 36%.
(vi) For duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the
following apply:
(A) The control level shall
be set at an RPD of 25%.
(B) The
warning level shall be set at an RPD of 50%.
(C) The control limit shall be set at an RPD
of 67%.
(vii) If the
plotted RPD result falls outside of the control limit, the measurements shall
cease until the problem is identified and corrected.
(viii) If the plotted RPD result falls
outside of the warning level, "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5,
shall be used to determine the action to be taken.
(ix) Documentation of duplicates must include
all of the following:
(A) The device serial
numbers.
(B) The exposure
dates.
(C) Each duplicate
measurement result.
(D) The RPD
results.
(d) LS, AC and ATs for secondary testers.
(1)
Data collection
log. Detector data shall be tracked on a form that contains all of the
following:
(i) The device serial
number.
(ii) The serial number of
duplicate devices.
(iii) The serial
number of spiked devices.
(iv) The
exposure dates and times.
(v) The
test results.
(vi) The RPE value or
RPD value.
(vii) The address of the
building tested.
(viii) The test
location in the building.
(ix) The
name of the tester who placed the device.
(x) The name of the tester who retrieved the
device.
(xi) The name of the
laboratory to which device was sent.
(2)
Known exposure measurements
(spikes).
(i) Spikes shall be
conducted at a rate of 3 for each 100 test devices deployed, with a minimum of
3 spikes for each certification year when tests were conducted in the
certification year, and with a maximum of 6 spikes each month.
(ii) Spikes shall be submitted to a
Department-certified laboratory labeled as QA. The RV of the spiked device may
not be revealed to the laboratory prior to analysis.
(iii) Spikes shall be monitored using a means
control chart. The means control chart must be established as follows:
(A) Using an RPE value of plus and minus 10%,
which corresponds to the 1 sigma level.
(B) A warning level of the RPE of plus and
minus 20%, which corresponds to the 2 sigma warning level.
(C) Control limits of the RPE of plus and
minus 30%, which correspond to the 3 sigma control level.
(iv) Each RPE value shall be plotted on the
means control chart within 1 week of receiving the result from the laboratory.
If the RPE value is outside the 3 sigma control level, all measurements shall
cease until the problem is evaluated and corrected. All evaluations shall be
documented.
(v) In addition to the
means control chart, all spikes shall be documented on a form that contains all
of the following:
(A) The radon chamber
name.
(B) The device serial
numbers.
(C) The RV from radon
chamber.
(D) The measured spike
value or values.
(E) The individual
RPE results.
(F) The certification
year beginning date and end date.
(G) The exposure dates.
(3)
Duplicate
measurements.
(i) Duplicates shall
be made in at least 10% of the total number of test devices deployed each
month, or 50 each month, whichever is smaller.
(ii) The RPD shall be calculated for all
duplicate results with an average of greater than or equal to 2.0 pCi/L. Two
control charts shall be constructed to monitor duplicate precision:
(A) One for duplicates when the average is
greater than or equal to 4.0 pCi/L.
(B) One for duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.
(iii) Each RPD value shall be
plotted on the control chart within 1 week of performing the duplicate
measurement.
(iv) The RPD shall be
tracked using control charts from "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2
and B-3.
(v) For duplicates when
the average is greater than or equal to 4.0 pCi/L, all of the following apply:
(A) The control level shall be set at an RPD
of 14%.
(B) The warning level shall
be set at an RPD of 28%.
(C) The
control limit shall be set at an RPD of 36%.
(vi) For duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the
following apply:
(A) The control level shall
be set at an RPD of 25%.
(B) The
warning level shall be set at an RPD of 50%.
(C) The control limit shall be set at an RPD
of 67%.
(vii) If the
plotted RPD result falls outside of the control limit, the measurements shall
cease until the problem is identified and corrected.
(viii) If the plotted RPD result falls
outside of the warning level, "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, shall be used
to determine the action to be taken.
(ix) Documentation of duplicates must include
all of the following:
(A) The device serial
numbers.
(B) The exposure
dates.
(C) Each duplicate
measurement result.
(D) The RPD
results.
(4)
Field blanks.
(i) Field
blank results shall be monitored and recorded. Field blanks shall be performed
at a rate of 5% of the devices that are deployed each month, or 25 each month,
whichever is smaller, or a minimum of 1 per certification year, unless tests
are not performed. These devices shall be set aside, kept in a low-radon
environment and labeled as QA when submitted to the laboratory.
(ii) If a field blank has a concentration
greater than the lowest level of detection (LLD) as established by the
laboratory, all of the following shall occur:
(A) The occurrence shall be documented and
reported to the laboratory.
(B) The
cause shall be investigated in conjunction with the laboratory and
documented.
(iii)
Documentation of field blanks must include all of the following:
(A) The device serial numbers.
(B) The date submitted to
laboratory.
(C) The measurement
results.
(D) The laboratory's
reported LLD.
Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.