Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 240 - RADON CERTIFICATION
Subchapter G - QA REQUIREMENTS
Section 240.604 - QA requirements for testing using primary devices
Universal Citation: 25 PA Code ยง 240.604
Current through Register Vol. 54, No. 44, November 2, 2024
(a) CRMs for primary testers.
(1)
Calibration. Each Department-listed CRM must have a current
calibration. To have a current calibration, the CRM shall be calibrated in a
Department-approved calibration facility within 1 year from the date of the
previous calibration and when alterations or repairs are made to the
CRM.
(2)
Background
measurements. Background measurements shall be performed and
documented after every 1,000 hours of operation of scintillation cell-type CRM.
These background measurements shall be checked by purging the unit with clean,
aged air or nitrogen in accordance with the manufacturer's instructions. For
all CRMs, the background shall be monitored in accordance with the
manufacturer's instructions.
(3)
Check source counting. For a CRM with a check source, check
source counting shall be documented and completed with that check source prior
to each test.
(4)
Routine
instrument checks. Before and after each measurement, the CRM shall be
checked according to the manufacturer's instructions. For each check, all of
the following shall be verified:
(i) The
correct input parameters and the unit's clock or timer are set
properly.
(ii) The pump's flow
rates are within the range of the manufacturer's specifications.
(5)
Data collection
log.
(i) CRM data shall be tracked
on a form that contains all of the following:
(A) The CRM serial number.
(B) The exposure dates and times.
(C) The test result.
(D) The address of the building
tested.
(E) The test location in
the building.
(F) The name of the
tester who placed the CRM.
(G) The
name of the tester who retrieved the CRM.
(H) The calibration, repair and Department
listing dates.
(ii) For a
CRM without a check source, the data collection log must also contain all of
the following intercomparison measurement information:
(A) The intercomparison devices' serial
numbers.
(B) The RPD
value.
(C) The intercomparison
measurements results.
(6)
Intercomparison
measurements. An intercomparison measurement shall be performed for
each CRM without a check source.
(i)
Intercomparison measurements shall be made at least every tenth test with
another Department-listed passive device that is analyzed by a
Department-certified laboratory or with another CRM with a hard copy printout.
The intercomparison measurements shall be distributed systematically throughout
the entire population of test locations. Original printouts or
Department-certified laboratory results, or both, shall be kept for each
intercomparison. Each intercomparison measurement must be performed with the
devices side by side for the measurement for at least 48 hours.
(ii) For intercomparison measurements, the
RPD shall be used to track performance. The RPD value shall be tracked using
control charts from "Protocols for Radon and Radon Decay Product Measurements
in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2 and
B-3.
(iii) If the RPD value exceeds
the control limit, the CRM may not be used for radon measurements until the
problem is identified and corrected. If the RPD value exceeds the warning
level, the criteria in "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5,
shall be followed.
(iv) In addition
to the control charts, intercomparison measurements shall be documented on the
CRM data collection log.
(b) CWLMs for primary testers.
(1)
Calibration. Each Department-listed CWLM must have a current
calibration. To have a current calibration, the CWLM shall be calibrated in a
Department-approved calibration facility within 1 year from the date of the
previous calibration and when alterations or repairs are made to the
CWLM.
(2)
Background
measurements. CWLM background measurements shall be performed and
documented at least every 168 hours of operation and when the unit is
calibrated.
(3)
Routine
instrument checks. Routine instrument checks for each CWLM shall be
documented and performed before and after each test by using an Am-241 or
similar energy check source. Pumps and flow meters shall be checked in
accordance with the manufacturer's instructions and documented. The pump and
flow meter check shall be performed with a dry-gas meter or other flow
measurement device of traceable accuracy.
(4)
Data collection log.
(i) CWLM data shall be tracked on a form that
contains all of the following:
(A) The CWLM
serial number.
(B) The exposure
dates and times.
(C) The test
result.
(D) The address of the
building tested.
(E) The test
location in the building.
(F) The
name of the tester who placed the CWLM.
(G) The name of the tester who retrieved the
CWLM.
(H) The calibration, repair
and Department listing dates.
(ii) For CWLMs without a check source, the
data collection log must also contain all of the following intercomparison
measurement information:
(A) The
intercomparison devices' serial numbers.
(B) The RPE value or RPD value.
(C) The intercomparison measurement
results.
(5)
Intercomparison measurements. An intercomparison measurement
shall be performed for each CWLM monitor without a check source.
(i) A CWLM without check source capability
must have an informal intercomparison measurement made with another CWLM with a
hard copy printout at least every tenth test. This printout shall be retained
for each intercomparison. The intercomparison measurements shall be distributed
systematically throughout the entire population of test locations. Each
intercomparison measurement must be performed with the devices side by side for
the measurement for at least 48 hours.
(ii) Each intercomparison shall be documented
on the data collection log.
(iii)
For intercomparison measurements the RPD shall be used to track performance.
The RPD value shall be tracked using control charts from "Protocols for Radon
and Radon Decay Product Measurements in Homes," EPA 402-R-92-003, May 1993,
Appendix B, Exhibits B-2 and B-3.
(iv) If the RPD value exceeds the control
limit, the CWLM may not be used for radon measurements until the problem is
identified and corrected. If the RPD value exceeds the warning level, the
criteria in "Protocols for Radon and Radon Decay Product Measurements in
Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5, shall be
followed.
(c) Electret ion chambers for primary testers.
(1)
Calibration. Each
Department-listed electret reader must have a current calibration. To have a
current calibration, the electret reader shall be calibrated in a
Department-approved calibration facility within 1 year from the date of the
previous calibration and when alterations or repairs are made to the electret
reader. Each electret reader shall be calibrated simultaneously with its
corresponding reference electret's recertification.
(2)
Data collection log.
Electret custody shall be tracked on a form that contains all of the following:
(i) The electret serial number.
(ii) The initial voltage reading.
(iii) The final voltage reading.
(iv) The exposure dates and times.
(v) The test result.
(vi) The serial number of duplicate
electret.
(vii) The RPD
value.
(viii) The address of the
building tested.
(ix) The test
location in the building.
(x) The
name of the tester who placed the electret.
(xi) The name of the tester who retrieved the
electret.
(3)
Known exposure measurements (spikes).
(i) Spikes shall be conducted at a rate of 3
for each 100 test devices deployed, with a minimum of 3 spikes for each
certification year when tests were conducted in the certification year, and
with a maximum of 6 spikes each month.
(ii) Spikes shall be analyzed in the same
manner as all other testing.
(iii)
Spikes shall be monitored using a means control chart. The means control chart
must be established as follows:
(A) Using an
RPE value of plus and minus 10%, which corresponds to the 1 sigma
level.
(B) A warning level of the
RPE of plus and minus 20%, which corresponds to the 2 sigma warning
level.
(C) Control limits of the
RPE of plus and minus 30%, which corresponds to the 3 sigma control
level.
(iv) Each RPE
value shall be plotted on the means control chart within 1 week of return of
the device from the chamber. If the RPE value is outside the 3 sigma control
level, all measurements shall cease until the problem is evaluated and
corrected. All evaluations shall be documented.
(v) In addition to the means control chart,
all spikes shall be documented on a form that contains all of the following:
(A) The radon chamber name.
(B) The electret serial numbers.
(C) The RV from radon chamber.
(D) The measured spike value or
values.
(E) The individual RPE
results.
(F) The certification year
beginning date and end date.
(G)
The exposure dates.
(4)
Duplicate measurements.
(i) Duplicates shall be made in at least 10%
of the total number of test devices deployed each month, or 50 each month,
whichever is smaller.
(ii) The RPD
shall be calculated for all duplicate results with an average of greater than
or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor
duplicate precision:
(A) One for duplicates
when the average is greater than or equal to 4.0 pCi/L.
(B) One for duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.
(iii) Each RPD value shall be
plotted on the control chart within 1 week of performing the duplicate
measurement.
(iv) The RPD shall be
tracked using control charts from "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2
and B-3.
(v) For duplicates when
the average is greater than or equal to 4.0 pCi/L, all of the following apply:
(A) The control level shall be set at an RPD
of 14%.
(B) The warning level shall
be set at an RPD of 28%.
(C) The
control limit shall be set at an RPD of 36%.
(vi) For duplicates when the average is
greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the
following apply:
(A) The control level shall
be set at an RPD of 25%.
(B) The
warning level shall be set at an RPD of 50%.
(C) The control limit shall be set at an RPD
of 67%.
(vii) If the
plotted RPD result falls outside of the control limit, the measurements shall
cease until the problem is identified and corrected.
(viii) If the plotted RPD result falls
outside of the warning level, "Protocols for Radon and Radon Decay Product
Measurements in Homes," EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5,
shall be used to determine the action to be taken.
(ix) Documentation of duplicates must include
all of the following:
(A) The device serial
numbers.
(B) The exposure
dates.
(C) Each duplicate
measurement result.
(D) The RPD
results.
(5)
Voltmeter routine instrument checks.
(i) Proper operation of the surface voltmeter
shall be monitored following the manufacturer's procedures for analyzing the
reference electrets and zeroing the voltmeter.
(ii) A voltage reading of a reference
electret difference of more than 2 volts from the reference electret specified
value shall be considered a wrong reading. The second reference electret in the
set shall be read to determine whether the wrong reading is in the first
reference electret or in the reader. Corrective action shall be taken in
consultation with the manufacturer.
(iii) When zeroing the reader, if the
voltmeter displays more than (±) 3 volts, corrective action shall be
taken in consultation with the manufacturer.
(iv) Voltmeter checks shall be conducted at
least once each week while the voltmeter is in use and shall be documented.
Documentation of routine instrument checks must include all of the following:
(A) The reader serial number.
(B) The date of analysis.
(C) Zero value.
(D) The reference electret values.
(E) Corrective actions performed.
This section cited in 25 Pa. Code § 240.102 (relating to prerequisites for radon testing certification); and 25 Pa. Code § 240.122 (relating to prerequisites for radon laboratory certification).
Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.