RADIATION SAFETY REQUIREMENTS FOR ACCELERATORS USED IN THE HEALING ARTS
Section 228.72 - Selection of radiation type

Universal Citation: 25 PA Code ยง 228.72

Current through Register Vol. 54, No. 44, November 2, 2024

Equipment capable of X-ray therapy or electron therapy, or both, must meet all of the following additional requirements:

(1) Irradiation may not be possible until a selection of radiation type and appropriate energy has been made and displayed at the control panel.

(2) An interlock system shall be provided to insure that the equipment can emit only the radiation type which has been selected.

(3) An interlock system shall be provided to prevent irradiation if selected operations carried out in the treatment room do not agree with the selected operations carried out at the control panel.

(4) An interlock system shall be provided to prevent:

(i) Irradiation with X-rays except to obtain a port film when electron applicators are fitted.

(ii) Irradiation with electrons when accessories specific for X-ray therapy are fitted.

(5) For new equipment, a system shall be provided to terminate irradiation if the energy of the electrons striking either the X-ray target or electron window deviates by more than +20% or 3 MeV, whichever is smaller, from the selected nominal energy.

Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.