Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 228 - RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS
RADIATION SAFETY REQUIREMENTS FOR ACCELERATORS USED IN THE HEALING ARTS
Section 228.71 - Timer

Universal Citation: 25 PA Code ยง 228.71

Current through Register Vol. 54, No. 44, November 2, 2024

(a) The control panel shall have a timer that is graduated in minutes and fractions of minutes or seconds. The timer shall have a preset time selector and an elapsed time indicator.

(b) The timer shall be cumulative and activated only during irradiation and shall retain its reading after irradiation is interrupted or terminated.

(c) The timer shall terminate irradiation when a preselected time has elapsed if the dose monitoring systems fail to do so.

Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.