Section 228.69 - Termination of irradiation by the dose monitoring system or systems

Universal Citation: 25 PA Code ยง 228.69

Current through Register Vol. 54, No. 44, November 2, 2024

(a) A dose monitoring system shall be capable of independently terminating irradiation.

(b) A primary system shall terminate irradiation when the preselected number of dose monitor units has been detected by the system.

(c) A secondary dose monitoring system shall terminate irradiation when either 110% of the preselected number of dose monitor units or 10 dose monitor units (whichever is greater) has been detected by the secondary dose monitoring system.

(d) For new equipment, an indicator on the control panel shall show which dose monitoring system has terminated irradiation.

Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.