Section 228.66 - Beam monitors
Current through Register Vol. 54, No. 44, November 2, 2024
(a) Therapy systems shall be provided with radiation detectors in the radiation head.
(b) New equipment shall be provided with at least two radiation detectors incorporated into two separate dose monitoring systems.
(c) Existing equipment shall be provided with at least one radiation detector incorporated into a primary dose monitoring system.
(d) The detector in a dose monitoring system shall be:
(e) For new equipment, the design of dose monitoring systems shall assure that:
(f) A dose monitoring system shall have a legible display at the control panel. For new equipment, a display shall:
This section cited in 25 Pa. Code § 228.74 (relating to absorbed dose rate).