Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 225 - RADIATION SAFETY REQUIREMENTS FOR INDUSTRIAL RADIOGRAPHIC OPERATIONS
Subchapter B - RADIATION-PRODUCING DEVICES
RADIATION SURVEY INSTRUMENT AND PERSONNEL MONITORING REQUIREMENTS
Section 225.92 - Radiation survey meter calibration requirements
Universal Citation: 25 PA Code ยง 225.92
Current through Register Vol. 54, No. 44, November 2, 2024
(a) In addition to the requirements of § 225.91 (relating to survey meter requirements), instruments required by this chapter shall have a range so that 2 mR (0.516 µC/kg) per hour through 1 R (258 µC/kg) per hour can be measured.
(b) Each radiation instrument shall be calibrated:
(1) At
energies appropriate for use.
(2)
At intervals not to exceed 6 months.
(3) After each instrument servicing, other
than battery replacement.
(4) To
within an accuracy of +/- 20%.
(5)
At two points located approximately one-third and two-thirds of full scale on
each scale of linear scale instruments; at mid-range of each decade and at two
points of at least 1 decade for logarithmic scale instruments; and for digital
instruments, at three points between 2 mR (0.516 µC/kg) and 1,000 mR (258
µC/kg) per hour.
(6) By a
person authorized by the Department, the NRC or an agreement state.
(c) Calibration records shall be maintained for inspection by the Department for 5 years after the date of calibration.
Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.