THERAPEUTIC X-RAY SYSTEMS WITH ENERGIES LESS THAN 1 MEV
Section 221.72 - Facility design requirements for systems capable of operating above 50 kVp
Current through Register Vol. 54, No. 44, November 2, 2024
(a) Provision shall be made to permit continuous observation of and communication with the patient during irradiation.
(b) Windows, mirror systems or closed-circuit television viewing screens used for observing the patient shall be so located that the operator can maintain direct surveillance over both the control panel and the patient.
(c) Treatment rooms which contain an X-ray system capable of operating above 150 kVp shall meet the following additional requirements:
The provisions of this §221.72 issued under section 302 of the Radiation Protection Act (35 P.S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20).