Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 221 - X-RAYS IN THE HEALING ARTS
THERAPEUTIC X-RAY SYSTEMS WITH ENERGIES LESS THAN 1 MEV
Section 221.71 - Equipment requirements
Current through Register Vol. 54, No. 44, November 2, 2024
(a) When the tube is operated at its leakage technique factors, the leakage radiation may not exceed:
(b) Fixed diaphragms or cones used for limiting the useful beam must provide at least the same protection as required by the tube housing assembly.
(c) Beam limiting devices may, for the portion of the useful beam blocked by these devices, transmit not more than 5% of the original X-ray beam intensity at the maximum voltage and maximum treatment filter. This requirement does not apply to auxiliary blocks or materials placed in the useful beam to shape the useful beam to the individual patient.
(d) The filter system shall be designed so that:
(e) The tube housing assembly shall be immobilized during stationary treatments.
(f) The tube housing assembly shall be so marked that it is possible to determine the location of the focal spot to within 5 millimeters, and the marking shall be readily accessible for use during calibration procedures.
(g) Contact therapy tube housing assemblies shall have a removable shield of at least .5 millimeter lead equivalency at 100 kVp that can be positioned over the entire useful beam exit port during periods when the beam is not in use.
(h) Systems of greater than 150 kVp manufactured after December 19, 1987, must have a beam monitor system which meets the following requirements:
(i) The following apply to timers on the equipment:
(j) The control panel, in addition to the displays required in this section, must have:
(k) When a control panel may energize more than one X-ray tube, the following requirements shall be met:
(l) There must be a means of determining the SSD to within 5 millimeters.
(m) Unless it is possible to bring the X-ray output to the prescribed exposure parameters within 5 seconds, the entire useful beam shall be automatically attenuated by a shutter having a lead equivalency not less than that of the tube housing assembly.
(n) Electronic brachytherapy devices are exempt from the requirements in subsections (k)-(m).
The provisions of this §221.71 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P.S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20); and the Radon Certification Act (63 P.S. §§ 2001-2014).
This section cited in 25 Pa. Code § 216.1 (relating to purpose and scope); and 25 Pa. Code § 216.3 (relating to exemptions).