Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 221 - X-RAYS IN THE HEALING ARTS
DIAGNOSTIC INSTALLATIONS GENERAL REQUIREMENTS
Section 221.50 - Facilities using CR or DR

Universal Citation: 25 PA Code ยง 221.50

Current through Register Vol. 54, No. 44, November 2, 2024

(a) When exposure indicators are available, the facility shall establish, document and post an acceptable range for the exposure values for examinations routinely performed at the facility. The indicated exposure values for each image shall be compared to the established range. Consistent deviations from established ranges shall be investigated, corrective actions taken as necessary and results documented.

(b) Facilities shall establish and follow an image QC program in accordance with the recommendations of a QMP, the system manufacturer or a nationally-recognized organization.

(c) Facilities other than dental, podiatric and veterinary shall complete phantom image evaluation using a phantom approved by a QMP, system manufacturer or the Department. The evaluation shall be completed on a quarterly basis and include, at a minimum, all of the following:

(1) Artifacts.

(2) Spatial resolution.

(3) Contrast/noise.

(4) Workstation monitors.

(5) Exposure indicator constancy.

(d) In addition to subsections (a)-(c), CR facilities shall erase all CR cassettes, at a minimum, on a weekly basis.

(e) Dental and podiatric facilities shall maintain and operate photostimulable storage phosphor and DDR systems in accordance with manufacturer specifications.

(f) The facility shall maintain records for 5 years for inspection by the Department.

The provisions of this §221.50 issued under sections 301 and 302 of the Radiation Protection Act (35 P.S. §§ 7110.301 and 7110.302); and section 1920-A of the Administrative Code (71 P.S. § 510.20).

Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.