Section 221.38a - Entrance exposure rate
Current through Register Vol. 54, No. 44, November 2, 2024
(a) Fluoroscopic systems without high level control. The exposure rate may not exceed 10 roentgens (2.58 mC/kg) per minute except during recording of fluoroscopic images.
(b) Fluoroscopic systems with high level control.
(c) Frequency of output measurements. Output measurements to show compliance with this section shall be made at least annually and after maintenance that could affect the output of the machine.
(d) Compliance requirements. Compliance with subsections (a)-(c) shall be determined as follows: