Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 221 - X-RAYS IN THE HEALING ARTS
DIAGNOSTIC INSTALLATIONS GENERAL REQUIREMENTS
Section 221.36a - Limitation of useful beam of fluoroscopic equipment
Current through Register Vol. 54, No. 44, November 2, 2024
(a) The fluoroscopic imaging assembly shall be provided with a primary protective barrier which intercepts the entire cross section of the useful beam at any source-to-image receptor distance.
(b) The X-ray tube used for fluoroscopy may not produce X-rays unless a barrier is in position to intercept the useful beam and the imaging device is in place and operable.
(c) A means shall be provided for stepless (continuous) adjustment of the field size.
(d) The minimum field size at the greatest source to image receptor distance shall be containable in a square of 5 centimeters by 5 centimeters unless otherwise provided in 21 CFR 1020.32(b) (relating to fluoroscopic equipment).
(e) Equipment may not be operated at a source to skin distance less than 30 centimeters or as required under 21 CFR 1020.32(g).
(f) The width of the X-ray field in the plane of the image receptor may not exceed that of the visible area of the image receptor by more than 3% of the source to image receptor distance. The sum of the excess length and the excess width may not be greater than 4% of the source to image receptor distance.
(g) For rectangular X-ray fields used with a circular image receptor, the error in alignment shall be determined along the length and width dimensions of the X-ray field which passes through the center of the visible area of the image receptor.
(h) Compliance with subsections (a)-(g) shall be determined with the beam axis perpendicular to the plane of the image receptor.
(i) Spot-film devices shall meet the following additional requirements:
The provisions of this §221.36a issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.43a (relating to mobile fluoroscopes).