Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 221 - X-RAYS IN THE HEALING ARTS
DIAGNOSTIC INSTALLATIONS GENERAL REQUIREMENTS
Section 221.30 - Exposure reproducibility for noncertified systems

Universal Citation: 25 PA Code ยง 221.30
Current through Register Vol. 54, No. 44, November 2, 2024

The coefficient of variation of exposure reproducibility may not exceed 0.10 when technique factors are held constant. This requirement shall be deemed to have been met when four exposures are made. This requirement applies when either manual techniques or automatic exposure control is used.

The provisions of this §221.30 issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. §§ 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); and the Radon Certification Act (63 P. S. §§ 2001-2014).

This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).

Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.