Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 221 - X-RAYS IN THE HEALING ARTS
DIAGNOSTIC INSTALLATIONS GENERAL REQUIREMENTS
Section 221.28 - Technique indicators

Universal Citation: 25 PA Code ยง 221.28

Current through Register Vol. 54, No. 44, November 2, 2024

(a) The technique factors for radiographic systems shall be indicated before exposure except for units utilizing automatic exposure controls, in which case the maximum mAs shall be indicated.

(b) The requirement of subsection (a) may be met by permanent markings on equipment having fixed technique factors. Indication of technique factors shall be visible from the operator's position except in the case of spot films made by a fluoroscopist.

The provisions of this §221.28 issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).

This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).

Disclaimer: These regulations may not be the most recent version. Pennsylvania may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.