COMPUTED TOMOGRAPHY X-RAY SYSTEMS
Section 221.202 - Equipment requirements
Current through Register Vol. 54, No. 44, November 2, 2024
(a) Accreditation. All diagnostic CT X-ray systems must be accredited by an accrediting organization recognized by the Department within 1 year from first patient use.
(b) Technical and safety information. The technical and safety information relating to the conditions of operation, dose information and imaging performance provided by the CT manufacturer shall be maintained by the facility and readily accessible to the operators.
(c) Termination of exposure. The operator shall be able to terminate the X-ray exposure at any time during a scan, or series of scans under X-ray system control, of greater than 0.5 second duration. Termination of the X-ray exposure shall necessitate resetting of the conditions of operation prior to initiation of another scan.
(d) Tomographic plane indication and alignment.
(e) Status indicators and control switches.
(f) Indication of CT conditions of operation. The CT X-ray system shall be designed so that the CT conditions of operation to be used during a scan or a scan sequence are indicated prior to the initiation of a scan or a scan sequence. On equipment having all or some of these conditions of operation at fixed values, this requirement may be met by permanent markings. Indication of CT conditions of operation shall be visible from any position from which scan initiation is possible.
(g) Leakage radiation. The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source may not exceed 100 milliroentgens (25.8 µC/kg) in 1 hour when the X-ray tube is operated at its leakage technique factors. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
(h) Additional requirements applicable to CT X-ray systems containing a gantry manufactured after September 3, 1985.
The provisions of this §221.202 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P.S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20).
This section cited in 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems); 25 Pa. Code § 221.64 (relating to CBCT); 25 Pa. Code § 221.65 (relating to X-ray attenuation systems); 25 Pa. Code § 221.201 (relating to definitions); and 25 Pa. Code § 223.31 (relating to registrant responsibilities).