Pennsylvania Code
Title 25 - ENVIRONMENTAL PROTECTION
Part I - Department of Environmental Protection
Subpart D - Environmental Health and Safety
Article V - Radiological Health
Chapter 221 - X-RAYS IN THE HEALING ARTS
ADMINISTRATIVE CONTROLS
Section 221.11 - Registrant responsibilities
Current through Register Vol. 54, No. 52, December 28, 2024
(a) The registrant is responsible for directing the operation of X-ray systems under his administrative control and shall assure that the requirements of this article are met in the operation of the X-ray systems.
(b) An individual who operates an X-ray system shall be instructed adequately in the safe operating procedures and be competent in the safe use of the equipment. The instructions shall include items included in Appendix A (relating to determination of competence) and there shall be continuing education in radiation safety, biological effects of radiation, quality assurance and quality control.
(c) Protocol information, which specifies the techniques for examinations performed with the system, shall be provided in the vicinity of each diagnostic X-ray system's control panel. The protocol shall include information pertinent to the particular examination, such as:
(d) Written safety procedures and rules shall be available at a facility including restrictions of the operating technique required for the safe operation of the particular X-ray system. The operator shall be able to demonstrate familiarity with the rules.
(e) Except for patients who cannot be moved out of the room, only the staff and ancillary personnel or other persons required for the medical procedure or training shall be in the room during the radiographic exposure. The following apply for individuals other than the patient being examined:
(f) During diagnostic procedures in which the gonads are in the useful beam, gonad shielding of at least 0.5 millimeter lead equivalent shall be used for patients except for cases in which this would interfere with the diagnostic procedure.
(g) An individual may not be exposed to the useful beam except for healing arts purposes or under § 221.15 (relating to use of X-rays in research on humans). An exposure shall be authorized by a licensed practitioner of the healing arts. This provision specifically prohibits deliberate exposure for the following purposes:
(h) If a patient or image receptor requires auxiliary support during a radiation exposure the following apply:
(i) Procedures and auxiliary equipment designed to minimize patient and personnel exposure commensurate with the needed diagnostic information shall be utilized.
(j) The screen and film system used shall be spectrally compatible. Defective screens may not be used for diagnostic radiological imaging.
(k) With the exception of intraoral dental radiography, film may not be used without intensifying screens for routine diagnostic radiological imaging.
(l) The registrant shall have a quality assurance program. This quality assurance program shall be documented and be in accordance with guidelines established by the Department or by another appropriate organization recognized by the Department. At a minimum, the quality assurance program shall address repeat rate, DRLs, image recording, processing and viewing, image quality and artifacts, and maintenance and modifications to the quality assurance program. For CT, each study shall be checked. If an artifact is present, the registrant shall take corrective action as appropriate. Records shall be maintained by the registrant for inspection by the Department for 5 years. The Department's guidelines and a list of recognized organizations will be maintained and made available on the Department's website and on request.
(m) Neither the X-ray tube housing nor the collimating device may be handheld during the exposure unless specifically designed to be handheld.
(n) Functional damage to a patient organ or a physiological system that results from a prescribed causative procedure shall be reported to the Department as outlined in § 219.229 (relating to diagnostic or interventional procedure medical reports).
(o) The registrant shall maintain records documenting the QMP's qualifications and compliance with continuing education requirements.
The provisions of this § 221.11 amended under sections 301 and 302 of the Radiation Protection Act (35 P.S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20).
This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.42a (relating to control of scattered radiation).