Oregon Administrative Rules
Chapter 811 - BOARD OF CHIROPRACTIC EXAMINERS
Division 30 - X-RAY
Section 811-030-0030 - X-Ray Departments, Equipment and Procedures
Universal Citation: OR Admin Rules 811-030-0030
Current through Register Vol. 63, No. 12, December 1, 2024
(1) All X-ray departments, equipment, and procedures, including fluoroscopy, shall be in compliance with the current rules and regulations of the Oregon Radiation Protection Services, including, but not limited to: the physical design of the department, occupational exposure, collimation, shielding, exposure charts.
(2) In addition:
(a) The patient shall be an adequate
candidate for the radiographic or fluoroscopic procedure employed;
(b) The radiographic field shall be
restricted to the area of clinical interest;
(c) Specialized views shall be used any time
the area of clinical interest is not clearly visualized on a standard
film;
(d) Every exposure, including
post-treatment exposures, and scanograms, shall have clinical justification
with adequate documentation consistent with the patient's case
history;
(e) The operator shall
maintain a record on each exposure of each patient containing the patient's
name, the date, the operator's name or initials, the type of exposure, and the
radiation factors of time, mA, kVp and target film distance, including those
exposures resulting in the necessity of repeat exposure for better diagnostic
information, such as patient motion or poor technical factors. For computerized
and automated systems, the recording of technique factors is not necessary as
long as the equipment is calibrated and maintained. OAR
333-106-0045 requires the
facility to determine the typical patient exposure for their most common
radiographic examinations, i.e. technique chart.
(f) Each film shall be properly identified by
date of exposure, location of X-ray department, patient's name or number,
patient's age, right or left marker, postural position marker, and indication
of the position of the patient;
(g)
The patient with tremors must be immobilized;
(h) The radiographs of a patient with an
antalgic posture may be taken in an upright position only if the patient is
adequately supported and immobilized to insure diagnostic quality. Otherwise,
the recumbent position shall be used;
(i) Upright or postural views shall not be
used for any patient whose size exceeds the capacity of the X-ray equipment.
Penetration must be adequate on all films;
(j) Sectional views shall be taken in
preference to a single 14 x 36 inch film if the patient's size or height
prevents diagnostic quality on a single 14 x 36 inch film;
(k) If two exposures are made on a single
film, the area of exposure shall be critically collimated to avoid double
exposure of the overlapping area;
(l) All views shall employ graduated
filtration or adequate devices to attenuate the primary beam for the purpose of
reducing unnecessary radiation and to improve film quality. Split screens,
gradient or graded screens, paper light barriers inside the cassette, or any
other attenuating device in the beam between the patient and the film shall not
be permitted, other than the grid controlling scattered radiation;
(m) A record of radiographic findings on
every set of radiographs reviewed shall be included in the patient's permanent
file; and
(n) Radiographs shall be
kept and available for review for a minimum of seven years or until a minor
becomes 18 years of age, whichever is longer.
Statutory/Other Authority: ORS 441 & 684
Statutes/Other Implemented: ORS 684.025, 684.150, 684.155 & 441.059
Disclaimer: These regulations may not be the most recent version. Oregon may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.