Oklahoma Administrative Code
Title 40 - Board of Tests for Alcohol and Drug Influence
Chapter 50 - Ignition Interlock
Subchapter 1 - Ignition Interlock Devices, Service Centers, Technicians
Section 40:50-1-2.1 - Approval and denial

Universal Citation: OK Admin Code 40:50-1-2.1

Current through Vol. 42, No. 1, September 16, 2024

(a) The Board shall notify the applicant of certification of a device or of denial to certify the device within 10 days of such determination.

(b) The Board may deny certification of a device upon finding any of the following:

(1) A Defect in the design, materials or workmanship causing the device to fail to function as intended.

(2) False or inaccurate information provided by the manufacturer, manufacturer representative or independent laboratory.

(3) Modification of the components or design of the device or modification of National Highway Traffic Safety Administration specifications that causes the device to no longer satisfy the current National Highway Traffic Safety Administration specifications.

(4) The device fails the compliance testing conducted by the Board.

(5) The device fails to meet the requirements for certification or is no longer in compliance with all the requirements in this title.

(6) Any final action against a device or manufacturer by other state administrators or agencies in the United States.

Added at 27 Ok Reg 2667, eff 8-26-10

Disclaimer: These regulations may not be the most recent version. Oklahoma may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.