Oklahoma Administrative Code
Title 310 - Oklahoma State Department of Health
Chapter 281 - Medical Marijuana Regulations [Revoked]
Subchapter 11 - Use of X-Rays in the Healing Arts and Veterinary Medicine
Section 310:281-11-16 - Dental radiographic systems General requirements
Current through Vol. 42, No. 1, September 16, 2024
(1) Timers. Means shall be provided to terminate the exposure at a preset time interval, preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. In addition, it shall not be possible to make an exposure when the timer is set to a "zero" or "off" position if either position is provided.
(2) Reproducibility. With a timer setting of 0.3 second or less, the average exposure period (T) shall be greater than or equal to 5 times the maximum exposure period (T max) minus the minimum exposure period (T min) when 4 timer tests are performed: T > 5(T max - T min)
(3) X-Ray control.
(4) Exposure reproducibility. The coefficient of variation shall not exceed 0.10 when all technique factors are held constant. This requirement is met when 4 exposures are made at identical technique factors, the value of the average exposure (E) is greater than or equal to 5 times the maximum exposure (E max) minus the minimum exposure (E min): E > 5(E max - E min)
Added at 8 Ok Reg 3503, eff 8-15-91 (emergency); Added at 9 Ok Reg 1643, eff 5-29-92; Amended at 21 Ok Reg 1037, eff 5-13-04