Current through Vol. 42, No. 1, September 16, 2024
(a)
Limitation of useful beam.
(1)
Primary barrier.
(A) The
fluoroscopic imaging assembly shall be provided with a primary protective
barrier which intercepts the entire cross section of the useful beam at any
source-to-image receptor distance (SID).
(B) The x-ray tube used for fluoroscopy shall
not produce x-rays unless the barrier is in position to intercept the useful
beam and the imaging device is in place and operable.
(2)
X-Ray field.
(A) Means shall be provided for stepless
(continuous) adjustment of the field size.
(B) The minimum field size at the greatest
SID shall be equal to or less than 5 centimeters by 5 centimeters.
(C) Neither the length nor the width of the
x-ray field in the plane of the image receptor shall exceed that of the visible
area of the image receptor by more than 3 percent of the SID. The sum of the
excess length and the excess width shall be no greater than 4 percent of the
SID.
(D) Spot-film devices shall
meet the following additional requirements:
(i) Means shall be provided between the
source and the patient for adjustment of the x-ray field size in the plane of
the film to the size of that portion of the film which has been selected on the
spot-film selector.
(ii) Such
adjustment shall be automatically accomplished except when the x-ray field size
in the plane of the film is smaller than that of the selected portion of the
film.
(iii) The total misalignment
of the edges of the x- ray field with the respective edges of the selected
portion of the image receptor along the length or width dimensions of the x-ray
field in the plane of the image receptor shall not exceed 3 percent of the SID
when adjusted for full coverage of the selected portion of the image receptor.
(iv) The sum, without regard to
sign of the misalignment along any 2 orthogonal dimensions, shall not exceed 4
percent of the SID.
(v) The center
of the x-ray field in the plane of the film shall be aligned with the center of
the selected portion of the film to within 2 percent of the SID.
(E) Compliance with
310:281-11-13(a)(2) is determined with the beam axis perpendicular to the plane
of the image receptor.
(b)
Activation of the fluoroscopic
tube. X-ray production in the fluoroscopic mode shall be controlled by a
device which requires continuous pressure by the fluoroscopist for the entire
time of the exposure (dead-man switch). When recording serial fluoroscopic
images, the fluoroscopist shall be able to terminate the x-ray exposures at any
time, but means may be provided to permit completion of any single exposure of
the series in process.
(c)
Exposure rate limits:
(1)
Entrance Exposure Rate Allowable Limits.
(A) Fluoroscopic equipment provided with
automatic exposure rate control mode (no manual mode) shall not be operable at
any combination of tube potential and current that will result in an exposure
rate in excess of 2.6 mC/kg (10 roentgens) per minute at the point where the
center of the useful beam enters the patient, except:
(i) During recording of fluoroscopic images;
or
(ii) When an optional high level
control is provided. When so provided, the equipment shall not be operable at
any combination of tube potential and current which will result in an exposure
rate in excess of 2.6 mC/kg (10 roentgens) per minute at the point where the
center of the useful beam enters the patient unless the high level control is
activated. Special means of activation of high level controls shall be
required. The high level control shall only be operable when continuous manual
activation is provided by the operator. A continuous signal audible to the
fluoroscopist shall indicate that the high level control is being employed.
(B) Fluoroscopic
equipment (manual mode only) not provided with automatic exposure rate control
shall not be operable at any combination of tube potential and current that
will result in an exposure rate in excess of 2.6 mC/kg (10 roentgens) per
minute at the point where the center of the useful beam enters the patient,
except:
(i) During recording of fluoroscopic
images; or
(ii) When an optional
high level control is activated. The high level control shall only be operable
when continuous manual activation is provided by the operator. A continuous
signal audible to the fluoroscopist shall indicate that the high level control
is being employed.
(C)
Fluoroscopic equipment, provided with both automatic exposure rate control and
manual modes, shall not be operable at any combination of tube potential and
current that shall result in an exposure rate in excess of 2.6 mC/kg (10
roentgens) per minute in either mode at the point where the center of the
useful beam enters the patient, except:
(i)
During recording of fluoroscopic images; or
(ii) When the mode or modes have an optional
high level control, in which case the mode or modes shall not be operable at
any combination of tube potential and current that results in an exposure rate
in excess of 2.6 mC/kg (10 roentgens) per minute at the point where the center
of the useful beam enters the patient, unless the high level control is
activated. Special means of activation of high level controls shall be
required. The high level control shall only be operable when continuous manual
activation is provided by the operator. A continuous signal audible to the
fluoroscopist will indicate that the high level control is being employed.
(D) Any fluoroscopic
equipment manufactured after May 19, 1995 which can exceed 1.3 mC\kg (5
roentgens) per minute shall be equipped with an automatic exposure rate
control. All entrance exposure rates are limited to 2.6 mc/kg (10 roentgens)
per minute with an upper limit of 5.2 mC/kg (20 roentgens) per minute when the
high level control is activated.
(E) Compliance with the requirements of OAC
310:281- 11-13 is determined as follows:
(i)
If the source is below the x-ray table, then the exposure rate is measured 1
centimeter above the tabletop or cradle:
(ii) If the source is above the x-ray table,
the exposure rate shall be measured at 30 centimeters above the tabletop with
the end of the beam- limiting device or spacer positioned as closely as
possible to the point of measurement;
(iii) For a c-arm type of fluoroscope, then
the exposure rate is measured 30 centimeters from the input surface of the
fluoroscopic imaging assembly, with the source positioned at any available SID,
provided that the end of the beam-limiting device or spacer is no closer than
30 centimeters from the input surface of the fluoroscopic imaging assembly;
(iv) For a lateral type
fluoroscope, the exposure rate is measured at a point 15 centimeters for the
centerline of the x-ray table and in the direction of the x-ray source with the
end of the beam-limiting device or spacer positioned as closely to the point of
measurement. If the tabletop is movable, it shall be positioned as closely as
possible to the lateral x-ray source, with the end of the beam-limiting device
or spacer no closer than 15 centimeters to the centerline of the x-ray table.
(2)
Periodic Measurements. Periodic measurements of the entrance
exposure rate shall be performed by a qualified expert for both typical and
maximum values as follows (Materials should be placed in the useful beam to
protect the imaging system when conducting these periodic measurements.):
(A) Such measurements shall be made annually
or after any maintenance of the system which might affect the exposure rate;
(B) Measurement results must be
kept where any fluoroscopist may have ready access to such results while using
the fluoroscope, and include the following information:
(i) stated in coulombs per kilogram
(roentgens) per minute;
(ii)
include the technique factors used in determining such results;
(iii) The name of the individual performing
the measurements; and
(iv) the date
the measurements were performed.
(C) Conditions of periodic measurement of
typical entrance exposure rate are as follows:
(i) The measurements shall be made under the
conditions that satisfy the conditions of OAC 310:281-11-13(c)(1)(E);
(ii) The kVp, mA, and/or other
selectable parameters shall be adjusted to those settings typical of clinical
use on a 23 cm thick abdominal patient;
(iii) The x-ray system that incorporates
automatic exposure rate control shall have sufficient attenuative material
placed in the useful beam to produce a milliamperage and/or kilovoltage to
satisfy the conditions of OAC 310:281-11- 13(c)(2)(C)(ii);
(D) Conditions of periodic measurement of
maximum entrance exposure rate are as follows:
(i) The measurement shall be made under the
conditions that satisfy 310:281-11-13(c)(1)(E);
(ii) The kVp, mA and/or other selectable
parameters shall be adjusted to those settings which give the maximum entrance
exposure rate;
(iii) The x-ray
system(s) that incorporates automatic exposure control shall have sufficient
attenuative material placed in the useful beam to produce the maximum entrance
exposure rate of the system.
(d)
Barrier transmitted radiation rate
limits.
(1)
Primary barrier
transmission. The exposure rate due to transmission through the primary
protective barrier with the attenuation block in the useful beam combined with
radiation from the image intensifier shall not exceed 2 milliroentgens per hour
for each roentgen per minute exposure rate. Measurements shall be made at 10
centimeters from any accessible surface of the fluoroscopic imaging assembly
beyond the plane of the image receptor.
(2)
Measuring compliance of barrier
transmission.
(A) The exposure rate
due to transmission through the primary protective barrier combined with
radiation from the image intensifier is determined by measurements averaged
over an area of 100 square centimeters with no linear dimension greater than 20
centimeters.
(B) If the source is
below the tabletop, then the measurement is made with the input surface of the
fluoroscopic imaging assembly positioned 30 centimeters above the tabletop.
(C) If the source is above the
tabletop and the SID is variable, then the measurement is made with the end of
the beam-limiting device or spacer as close to the tabletop as it can be
placed, provided that it shall not be closer than 30 centimeters.
(D) Movable grids and compression devices
shall be removed from the useful beam during the measurement.
(E) The attenuation block shall be positioned
in the useful beam 10 centimeters from the point of measurement of entrance
exposure rate and between this point and the input surface of the fluoroscopic
imaging assembly.
(F) The
collimator shall be fully open when the measurement is made.
(e)
Indication of
potential and current. During fluoroscopy and cinefluorography, the kV
and the mA shall be continuously indicated at the control panel and/or the
fluoroscopist's position.
(f)
Source-to-Skin distance. The SSD shall not be less than:
(1) 38 centimeters on all stationary
fluoroscopes,
(2) 30 centimeters on
all mobile fluoroscopes, and
(3) 20
centimeters for image intensified fluoroscopes used for specific surgical
application.
(g)
Fluoroscopic timer.
(1) Means
shall be provided to preset the cumulative on-time of the fluoroscopic x-ray
tube. The maximum cumulative time of the timing device shall not exceed 5
minutes without resetting.
(2) A
signal audible to the fluoroscopist shall indicate the completion of any preset
cumulative on-time. Such signal will continue to sound while x-rays are
produced until the timing device is reset.
(h)
Radiation therapy simulation
systems. If a radiation therapy simulation system complies with (1) and
(2) of this subsection, then it is exempt from all the requirements of
310:281-11-13(a), (c), (d), and (g) provided that:
(1) such systems are designed and used in
such a manner that no individual other than the patient is in the x-ray room
during periods of time when the system is producing x-rays; and
(2) systems which do not meet the
requirements of 310:281- 11-13(g) are provided with a means of indicating the
cumulative time that an individual patient has been exposed to x-rays.
Procedures shall require in such cases that the timer be reset between
examinations.
Added at 8 Ok Reg 3503,
eff 8-15-91 (emergency); Added at 9 Ok Reg 1643, eff 5-29-92; Amended at 21 Ok
Reg 1037, eff 5-13-04; Amended at 27 Ok Reg 2512, eff
7-25-10