Current through all regulations passed and filed through September 16, 2024
[Comment: For dates of non-regulatory government publications,
publications of recognized organizations and associations, federal rules, and
federal statutory provisions referenced in this rule, see rule
3745-300-15 of the
Administrative Code titled "Incorporation by reference - voluntary action
program."]
(A)
Authority of a certified laboratory to perform
analyses.
(1) A certified laboratory produces
certified data under affidavit in accordance with rule
3745-300-13 of the
Administrative Code only when the analyses are performed within the
laboratory's current certification. Certification may limit the analysis of
certain environmental media, as indicated on the laboratory's certificate
issued by Ohio EPA under this chapter.
(2) In order to produce certified data to
support a voluntary action under this chapter and Chapter 3746. of the Revised
Code, the following shall occur:
(a) The
certified laboratory shall be certified for each
analyte, parameter group, and method at the time the laboratory performs the
analyses.
(b) The
certified laboratory's analyses shall remain
consistent with the laboratory's standard operating procedures (SOPs) and
quality assurance program plan (QAPP).
(c) The certified laboratory's SOPs and QAPP used to produce
certified data shall be consistent with all of the following:
(i) Requirements in the published method
published or endorsed by U.S. EPA.
(ii) The applicable minimum requirements in
paragraph
(D) of this rule.
(iii) Any additional requirements specified
during approval for analysis using performance-based methods under paragraph
(B)(1)(c) of this rule.
(iv) Where
there is a conflict between paragraph (A)(2)(c)(i) of this rule and paragraph
(A)(2)(c)(ii) of this rule or paragraph (A)(2)(c)(iii) of this rule,
requirements consistent with paragraph (A)(2)(c)(ii) of this rule or paragraph
(A)(2)(c)(iii) of this rule take precedence.
(3) Certification pursuant to this rule is
applicable to analyses performed in support of a voluntary action, including
but not limited to the issuance of a no further action letter under this
chapter and Chapter 3746. of the Revised Code. Certification pursuant to this
rule does not constitute certification under any other state or federal
laboratory certification or accreditation program.
(4) For certification obtained that relies on
third-party accreditation, the certified laboratory shall maintain in good
standing the accreditation provided in paragraphs (B)(1)(d) and (B)(1)(e) of
this rule.
(5) A volunteer may
request that a certified laboratory analyze the constituents of a hazardous
substance when a chemical testing method or technology does not exist to
measure the concentration of the hazardous substance. When a hazardous
substance is comprised of more than one constituent, the certified laboratory
shall obtain certification for each constituent, even if the constituent is not
listed as a hazardous substance.
(6) If a certified laboratory no longer
intends to retain certification, the certified laboratory may return the
certificate with a notice to Ohio EPA that indicates the certified laboratory's
intent to withdraw from certification. Upon withdrawal, the laboratory shall
not report data as certified data under this chapter.
(B) Methods for the analysis of analytes or
parameter groups are the following:
(1) A
laboratory
certified pursuant to this
rule for any method used for the analysis of any analyte or parameter
group that meets the following criteria, except as provided in paragraph (B)(2)
of this rule:
(a) Certification is restricted
to hazardous substances or petroleum. If a chemical testing method or
technology does not exist to measure the concentration of the hazardous
substance, then certification to test for constituents of the hazardous
substance may be granted pursuant to paragraph (A)(5) of this rule.
(b) Chemical testing methods. A
certified laboratory may apply for certification
for any chemical testing method published or endorsed by U.S. EPA. A
certified laboratory shall use a published or
endorsed method only in the manner for which that method is designed.
(c) Performance-based methods. At the request
of a laboratory, Ohio EPA may evaluate whether to certify a laboratory to
perform analyses using a performance-based method. A laboratory shall
demonstrate the laboratory's ability to perform the method using a proficiency
testing sample as provided in paragraph (C)(1) of this rule, if available, and
in accordance with the application requirements in paragraph (D)(4) of this
rule.
(d) Asbestos accreditations.
A laboratory
holding certification for the analysis of
asbestos shall have current accreditation in at least one of the following
programs:
(i) American industrial hygiene
association, asbestos analysts registry.
(ii) National institute of standards
technology, national voluntary laboratory accreditation program for asbestos
fiber analysis.
(iii) "The NELAC
Institute" (TNI) recognized accreditation body.
(e)
Accreditations for any constituent other than
asbestos. A laboratory that
holds certification
for
analysis of any constituent other than asbestos may rely on current
accreditation from an accreditation
body that is recognized by TNI.
(2) Testing for characteristic hazardous
waste or for radioactive materials is not included for certification under this
rule.
[Comment: Ohio EPA coordinates with the Ohio department of
health for the Ohio department of health's review of any release of radioactive
materials or substances.]
(3) Testing for sediment toxicity is not
included for certification under this rule. Testing for sediment toxicity shall
be performed in accordance with paragraph (F)(4)(b) of rule
3745-300-09 of the
Administrative Code.
(C)
Proficiency testing program.
(1) Use of
proficiency testing samples; requirement to purchase proficiency testing
samples through proficiency testing providers:
(a) Unless otherwise exempt from this
requirement, a laboratory that
holds certification under this rule shall do the
following:
(i) Analyze proficiency testing
samples representative of the analytes or parameter groups for
certification.
(ii) Receive from
the proficiency testing provider acceptable proficiency testing results
pursuant to the criteria of this rule.
(b) The certified laboratory shall order proficiency testing
samples from a proficiency testing provider that is approved to produce the
proficiency testing samples and evaluate the proficiency testing
results.
(c) For the purposes of
this rule, a certified laboratory shall analyze a
proficiency testing sample that was formulated and evaluated using the criteria
established by TNI.
(i) If the laboratory
holds certification for testing aqueous samples, the
laboratory shall analyze a non-potable water sample or, if a non-potable water
sample is not available, a drinking water sample.
(ii) If the laboratory
holds certification for testing solid matrix samples,
the certified laboratory shall analyze a solid
matrix sample or, if a solid matrix sample is not available, an aqueous
sample.
(d) Paragraph
(C)(1) of this rule does not apply to certifications for asbestos or to the
circumstances provided in paragraph (C)(2) of this rule.
(2) Criteria for analysis of proficiency
testing samples, exceptions, and waivers. To demonstrate compliance with this
rule for any analyte or parameter group a certified laboratory shall analyze proficiency testing
samples, which a proficiency testing provider prepared and evaluated using TNI
criteria, except as follows:
(a) When a
non-potable water proficiency testing sample is not available for an analyte or
parameter group for which the laboratory
holds certification, proficiency
testing samples prepared and evaluated based on drinking water criteria may be
used instead.
(b) For any analyte or
parameter group for which TNI has not published any proficiency testing
criteria or for which proficiency testing samples are not available, Ohio EPA
may waive the proficiency testing sample analysis requirement in paragraph
(C)(2) of this rule. In the case of a waiver, certification for the analyte or
parameter group shall be limited to the use of a performance-based method as
described in paragraph (B)(1)(c) of this rule.
(c) For any analyte or
parameter group for which the certified
laboratory holds
NELAC accreditation in good standing pursuant to
paragraph (B)(1)(e) of this rule, analysis of proficiency testing samples is
not required unless Ohio EPA determines that proficiency testing is
required.
(3) Use of
existing proficiency testing results. A certified
laboratory may use the proficiency testing results obtained for another state
or federal certification or accreditation program to demonstrate compliance
with this rule, provided that the proficiency testing samples comply with this
rule.
(4) Analysis of proficiency
testing samples.
(a) The
certified laboratory shall analyze proficiency
testing samples that include the analyte or parameter group which corresponds
to the scope of the laboratory's certification .
(b) The certified laboratory may analyze a proficiency testing
sample on more than one technology to demonstrate proficiency for an analyte or
parameter group and method. For example, a laboratory that
holds certification for volatile organic compounds by
gas chromatography and mass spectrometry may analyze the same proficiency
testing sample on both technologies.
(c) Analysis of proficiency testing samples
shall be conducted in accordance with the certified laboratory's SOPs and QAPP identified in
paragraph (D) of this rule.
(d) The
ordering and analysis of proficiency testing samples is based on a technology.
To comply with this rule, a certified laboratory
shall order a proficiency testing sample based on the technology that is
representative of the certification. For example, to encompass the scope of a
certification for volatile organic compounds, the certified laboratory shall ensure that the proficiency
testing sample contains both aromatics and halocarbons.
(5) Reporting and time lines for proficiency
testing studies are as follows:
(a) Reporting
proficiency testing results. A laboratory that is certified
for
multiple technologies for an analyte or parameter group shall analyze and
report proficiency testing results for each technology, and may use the same
proficiency testing sample. For example, the same volatile organic compound
proficiency testing sample may be analyzed on gas chromatography and mass
spectrometry with a separate result reported for each technology.
(b)
Ohio EPA may use periodic performance testing to assess a
certified laboratory's ability to perform testing
under this rule. Upon request, certified
laboratories shall analyze proficiency testing samples and shall report the
results to Ohio EPA.
(6)
Proficiency testing reports submitted to Ohio EPA shall
include the
following:
(a) Name of proficiency
testing provider.
(b)
Certified laboratory name and address.
(c)
Opening and closing dates of the proficiency testing study.
(d)
Date proficiency testing report was issued.
(e) Analyte or
parameter group with units, reported value, assigned value, and acceptance
limits.
(f) Performance
evaluation by proficiency testing provider.
(g) Technology code
or method description.
(h) Sample matrix
type.
(D) Minimum
requirements for the QAPP and SOPs. Each certified laboratory shall have a
written QAPP and written SOPs for every method and procedure used by the
certified laboratory to produce certified data,
and shall keep these documents at the certified
laboratory for use by laboratory personnel. The QAPP and SOPs shall comply with
paragraph (A)(2)(c)(iii) of this rule and the methods published or endorsed by
U.S. EPA, except as specified in this rule.
(1) At a minimum, the QAPP shall include
provisions that require the certified laboratory
reports issued in compliance with this chapter to contain the following:
(a) An accompanying affidavit that complies
with paragraph (P) of rule
3745-300-13 of the
Administrative Code.
(b) At a
minimum, a case narrative that includes the following:
(i) Discussion of any issues that impact the
quality of the data with sample receipt, sample process, or sample
analysis.
(ii) Discussion of any
potential bias in sample results, as appropriate.
(c) A report of the analytical results
determined by the methods indicated on the certified laboratory's certificate.
(d) A report of the quality control sample
results and indication of whether applicable criteria were met.
(e) A copy of the chain of custody that
accompanied the samples to the certified
laboratory.
(f) At a minimum, a
copy of the sample receipt form that records the following:
(i) Temperature of samples upon receipt by
the certified laboratory, if the method requires
monitoring.
(ii) Date and time the
samples were received by the certified
laboratory.
(iii) Notation of
whether holding times specified in the SOPs for sampling preparation and
analysis were exceeded.
(iv) Any
exceptions or special instructions for sample handling, analysis, or
reporting.
(v) Notation of whether
samples include appropriate labeling, such as the date and time of sample
collection and a sample identification notation.
(vi) Notation of whether sample containers
contain appropriate sample preservatives, if applicable.
(vii) Description of the general condition of
sample containers, including whether any containers were damaged or improperly
filled.
(2)
Data interpretation and reporting requirements. To ensure quality data
interpretation and quality reporting of certified
laboratory results, the QAPP or SOPs shall include, at a minimum, the
following:
(a) SOPs shall include information
regarding how the qualitative and quantitative analyses are performed and
interpreted by the analysts.
(b)
Certified laboratories shall report solid samples on a
dry weight basis, unless otherwise dictated by the method or when inadequate
sample volume limits the laboratory's ability to determine dry weight. The
moisture content also shall be reported, when applicable. SOPs shall include
processes and calculations for this purpose.
(c) Prior to issuance of the
certified laboratory reports as certified data,
certified laboratories shall complete peer review
of applicable calibration, calibration verification, quality assurance, and
quality control results, as well as sample laboratory results.
(d) SOPs for methods that include manual
integration of chromatographic data to ensure that manual integrations are
performed in a consistent and technically justifiable manner for standards,
samples, and quality control solutions. The SOPs shall contain the following
minimum requirements:
(i) Examples of proper
and improper manual integrations.
(ii) Procedures to manually adjust data to
ensure that obvious inaccuracies in automated integrations are corrected and
that reported results accurately reflect the information contained in the
analytical data.
(iii) Both
original and modified chromatograms, including the chromatographic peaks and
baselines, shall be peer reviewed.
(iv) The certified laboratory shall retain copies of the
original and modified chromatograms. These shall be made available to Ohio EPA,
or other interested parties, upon request.
(e) As appropriate, the QAPP or SOPs shall
contain a provision that the certified laboratory
shall narrate potential bias in sample results if the requirements in the SOP
cannot be met, including, but not limited to, the following:
(i) Failure to meet required holding
times.
(ii) Improper sample
preservation.
(iii) Inability to
perform corrective actions for calibration, calibration verification, or
quality control outliers.
(iv)
Insufficient sample amount.
(f) For dual column analysis, the following
apply:
(i) Results shall only be reported if
the analyte is detected in both columns.
(ii) If the certified laboratory does not designate a primary
column, then the higher result shall be reported unless a matrix interference
is causing the elevated concentration.
(iii) If the certified laboratory designates a primary column, then
the results from the primary column shall be reported unless matrix
interference is present.
(iv) In
cases where matrix interference is present, the lower result, or both results,
shall be included in the analytical report.
(g) Analytes reported as certified data shall
meet all calibration, calibration verification, and quality control criteria.
If analytes do not meet criteria, then the analytes shall not be reported as
certified data except as provided in paragraph
(D)(2)(e)(iii) or
(D)(4)(c)
of this rule. This applies to all analytes including, but not limited to, the
following:
(i) Analytes traditionally known as
poor performers or common laboratory contaminants.
(ii) Analytes that may meet method criteria
through provisions of marginal exceedance.
(h)
Certified
laboratories shall not provide certified data for tentatively identified
compounds.
(i) As appropriate, the
QAPP or SOPs shall require the certified
laboratory to report as certified data only for
analytes specified in the method, unless the laboratory's certification
specifically allows reporting of additional analytes for the method. If a
certified laboratory report includes analytes
that are not specified in the method and are not specifically allowed by the
laboratory's certification, the laboratory shall identify the analytes as
exceptions to the certified data attested to in the affidavit that is issued
with the certified laboratory report.
(j) If a certified laboratory's certification relies on
maintaining third-party accreditation in good standing, the laboratory shall
not report certified data under this rule when the relied-upon third-party
accreditation is not in good standing with the issuing accreditation
body.
(k) Samples for metals
analysis that are filtered prior to digestion shall be reported as dissolved
metals. Unfiltered samples or samples that are filtered after digestion may be
reported as total metals.
(3) Instrument calibration requirements. To
ensure the quality of the data to be analyzed, the QAPP or SOPs shall comply
with the approved method and shall include the following minimum instrument
calibration requirements:
(a) At a minimum,
SOPs shall include all of the following calibration requirements:
(i) Information about the frequency of
initial calibration and calibration verification.
(ii) Criteria to evaluate results of initial
calibration and calibration verification, including calibration
blanks.
(iii) Without exception,
corrective actions the analyst shall follow for initial calibration,
calibration verification, and calibration blanks when these standards do not
meet the criteria required by paragraph (G)(3)(a)(ii) of this rule, as
applicable.
(iv) A prohibition
against forcing the initial calibration curve through the origin.
(v) A prohibition against use of the zero
point in an initial calibration curve, unless specified by the method or there
are instrument limitations.
(vi)
Nonlinear initial calibrations (e.g., quadratic calibration model) may be used
but are restricted to compounds that have historically exhibited a nonlinear
response.
(vii) Nonlinear initial
calibration models shall not be used to extend the calibration range for
compounds that normally exhibit a linear response.
(viii) The lowest standard concentration used
for initial calibration shall be at or below the certified laboratory's practical quantitation
limit.
(b) Standard
operation procedures shall specify initial calibration models as follows:
(i) For quadratic calibration models, a
minimum number of standard concentrations is six.
(ii) Unless otherwise specified by the
method, for all other calibration models, the minimum number of standard
concentrations is five.
(iii) If
more than the minimum number of standard concentrations is used, only the
lowest or highest standard concentrations may be omitted from the calibration
model as long as the minimum number of standard concentrations from paragraph
(G)(3)(b)(i) or (G)(3)(b)(ii) of this rule are retained for use.
(c) Calibration solutions shall
meet the following minimum requirements:
(i)
Unless the method allows for use of a different solution, the same solution
used to prepare the initial calibration standards shall be used to prepare the
continuing calibration verification standard.
(ii) Unless use of the same solution is
specifically allowed by the method, when an initial calibration verification
standard is included in a method, a different solution other than the one used
to prepare the calibration curve shall be used.
(iii) A prohibition on the use of expired
standards or spiking solutions, except for the analysis of air
samples.
(iv) For analysis of air
samples, expired standards or spiking solutions may be used if revalidated
against an unexpired reference material or if recertified by the vendor. The
certified laboratory shall keep on file the
documentation of such revalidation or recertification.
(v) Retention time marker solutions shall be
used for petroleum analysis. These solutions shall be analyzed before the
instrument is calibrated.
(d) For all dual column analysis, the
calibration criteria required by paragraphs (G)(3)(a) to (G)(3)(c) of this rule
shall be met on the column used to report data as certified.
(4) Quality control. To ensure
reliable data, the QAPP or SOPs shall comply with the approved method and shall
include the following:
(a) Identify all
reagents, standards, and spiking solutions to be used in sample preparation and
analysis.
(b) Define criteria for
the quality control solutions or provide reference as to where the information
is available. When criteria are not met for all associated quality control
solutions, including but not limited to the method blank and spiked laboratory
control solutions, or when surrogate recoveries or internal standard recoveries
fail to meet the defined criteria in samples or quality control solutions,
corrective actions shall occur, except as provided in paragraph (G)(2)(e)(iii)
or (G)(4)(c) of this rule. Upon re-analysis of the failed quality control
solution once, appropriate corrective actions may include re-preparation of the
entire batch, including re-digestion, re-distillation, or
re-extraction.
(c) When surrogates,
internal standards, method blanks, calibration verification solutions, or
spiked laboratory control solutions are biased high and the associated samples
are non-detect for the outlying analytes, corrective actions need not be taken,
and reporting of certified data is acceptable.
(d) Include procedures to prepare initial,
continuing, and calibration verification standard solutions and calibration
blank solutions.
(e) Include
procedures to prepare samples, including the weight or volume of the
media.
(f) Calibration verification
solutions shall contain all target analytes, except for the analysis of
polychlorinated biphenyls. For polychlorinated biphenyl analysis, a spike mix
that contains aroclors 1016 and 1260 is sufficient to represent the range of
aroclors specified in the method.
(g) SOPs shall include information regarding
quality control solutions, including all of the following:
(i) Frequency of analysis.
(ii) Weight or volume of the media
used.
(iii) Criteria used to
evaluate results.
(h)
Quality control solutions shall be treated in the same manner as samples,
including handling, preservation, preparation, and equipment use.
(i) All detections in the method blank equal
to or greater than the reporting limit require corrective actions as specified
in paragraph (G)(4)(b) of this rule.
(j) All calibration verification standards
required by the method and evaluated for per cent recovery, as defined by the
methods, shall be reported based on the true value of the standard.
(k) To report dual column analysis data as
certified, criteria required by paragraph (G)(4)(b) of this rule for quality
control solutions shall be met on the column used.
(5) Preparation of samples. To ensure the
quality of the samples to be analyzed, the QAPP or SOPs shall comply with the
approved method and shall include the following minimum requirements for
preparation of samples:
(a) Identify
requirements for sample preservation, storage, holding times (including
beginning and ending times), and the proper sample collection container,
including the following:
(i) If requirements
from the approved method do not specify requirements for sample preservation,
storage, holding times, and the proper sample collection container, the
certified laboratory shall include such
requirements in the QAPP or SOPs, as applicable.
(ii) Holding times described in the approved
method shall not be increased by alternate preservation techniques or by
alternate demonstrations.
(iii) Air
samples from "Tedlar" bags shall not be reported as certified data. The
transfer of air samples from "Tedlar" bags to a canister for air analysis shall
be prohibited.
(b)
Identify equipment used for sample preparation, and identify diluents used for
all dilutions.
(c) Identify
requirements to be followed for holding times for extracted, digested, or
distilled samples, and the storage requirements and the proper storage
containers for each.
(d) Include
details to ensure that sample preparation specifications for digestion,
distillation, clean-up, and extraction shall meet the final volume for
analysis, either volumetrically measured or otherwise verified to meet
volumetric specifications.
(e) If
the preparatory batch standards (i.e., initial calibration or calibration
verification standards) for digestion, distillation, or extraction are
processed with the sample and the batch standards fail either quality control
criteria or calibration criteria, upon re-analysis of the failed quality
control solutions or calibration solutions once, the entire batch shall be
prepared again. The corrective action shall occur except as provided in
paragraph (G)(2)(e)(iii) or (G)(4)(c) of this rule.
(f) Analysis of non-aqueous samples for
volatile organic compounds shall utilize a closed-system purge-and-trap process
consistent with "SW-846" method 5035 or method 5035A, unless the analytical
method pre-dates December 1996.
(g)
For organic extraction methods that include instruction for drying solid matrix
samples, surrogates or any other spiking compounds shall be added with the
drying agent and into the homogenous mixture of sample or quality control
sample. Surrogates or any other spiking compounds may not be added via the
extraction solvent to samples and the associated quality control
solutions.
(6) Analysis
of samples. To ensure the quality of the samples to be analyzed, the QAPP or
SOPs shall comply with the approved method and shall include the following
minimum requirements:
(a) Identify equipment
and instrumentation used for analysis of samples.
(b) For organic analysis methods, samples
with failing internal standard or surrogate criteria require re-analysis of the
samples. Dilutions shall be made only if matrix interference is present.
Dilutions shall not be made for the sole purpose to meet, or attempt to meet,
internal standard or surrogate criteria.
(c) When mass spectrometry methods are used,
designate the primary and secondary ions used for identification of
compounds.
(d)
Certified laboratories that report selective ion
monitoring data shall include operating procedures for selective ion monitoring
analysis within the associated SOP.
(7) At a minimum, the written QAPP shall meet
any necessary requirements in paragraphs (G)(1) to (G)(6) of this rule and
shall include provisions that describe the following:
(a) Procedures that require proper citation
and use of method numbers, including the appropriate revision suffix, if
applicable, shall be consistently identified and included on instrument
printouts,
logbooks, analytical reports, and any other
laboratory documents. All method numbers plus the revision suffix, if
applicable, shall correlate with the method number and revision suffix on the
certificate issued under this rule.
(b) SOPs requirements shall be reviewed for
potential updates at least once every two years.
(c) Describe storage requirements of samples
during all phases of analysis.
(d)
Provide details for the expiration of stock standards, solutions, and all
working standards and solutions, or cross-reference to the location of that
information.
(e) Identify how the
certified laboratory shall establish quality
control acceptance limits for the analysis of samples.
(f) Identify how the certified laboratory shall manage waste in accordance
with all applicable federal, state, and local requirements.
(g) Include a provision to address initial
and periodic training for personnel in sample receipt, preparation, analysis,
and data interpretation and review:
(i)
Certified laboratory personnel shall review the QAPP
and applicable SOPs which relate to the tasks associated with laboratory
personnel's duties at the laboratory. Laboratory personnel shall sign
documentation that acknowledges review of the documents.
(ii) The certified laboratory shall maintain training records
and documentation that the laboratory personnel reviewed the appropriate
documents.
(E) Standards of
performance and conduct to maintain certification. To maintain certification
under this rule, a certified laboratory shall do the following:
(1) Produce results as certified data
pursuant to paragraph (A) of this rule when the certified laboratory is requested to provide data in
support of a voluntary action under this chapter or Chapter 3746. of the
Revised Code.
(2) Disclose when the
certified laboratory does not hold certification
for a requested analyte, parameter group, or method included in a request for
analysis. After this disclosure, if the requester still requests the analysis
to be performed, the certified laboratory shall
specify in the affidavit that accompanies the analytical report the analytes,
parameter groups, or methods for which the laboratory is not providing
certified data.
(3) Comply with the
methods for which the laboratory is certified.
(4) Notify Ohio EPA in writing within thirty
days after any of the following:
(a) A change
in management personnel or quality assurance personnel.
(b) A change in certified laboratory operations that affects the
laboratory's ability to perform analyses pursuant to this rule.
(c) A change in name or ownership of the
certified laboratory.
(d) A relocation of the
certified laboratory, in whole or in part, or a
change of address of the laboratory.
(e) Anything that results in the loss of
accreditation, temporarily or permanently, that is relied upon for
certification under paragraph (A)(4) of this rule for any analytes, parameter
groups, or methods for which the certified
laboratory holds certification.
(5) Perform acceptably on each
certified laboratory audit conducted pursuant to
this rule, and address in a timely manner the deficiencies that are identified
by Ohio EPA.
(6) Perform analyses
in accordance with the certified laboratory's
QAPP and SOPs that are consistent with paragraph
(D) of this rule
when the laboratory produces certified data.
(7) Disclose when the certified laboratory cannot quantify at or below an
applicable standard specified in a request for analysis as follows:
(a) The certified laboratory shall provide
certified data that detects chemicals of concern in environmental media at or
below the applicable standards, unless the certified laboratory discloses that the laboratory is
incapable of achieving an applicable standard under the laboratory's
certification.
(b) Unless the
certified laboratory is otherwise informed of the need for a lower applicable
standard, the certified laboratory shall quantify at or below the single
chemical generic numerical standards in appendices A and B to rule
3745-300-08 of the
Administrative Code.
(c) If a
certified laboratory that performs analyses in support of a no further action
letter but is not capable of detecting the chemicals of concern in
environmental media at or below the applicable standards, the
laboratory shall notify, in
writing, the person who requests the analysis that the laboratory cannot
quantify at or below an applicable standard using a method for which the
laboratory is currently certified. The certified
laboratory may disclose this information in the analytical report or by other
means.
(8) Not falsify
any information on any application, SOP, QAPP, or any proficiency testing
result, or any certified data used in support of a no further action letter, or
any other submittal to Ohio EPA.
(9) Not perform analyses in support of a
request for a no further action letter for which the certified laboratory has a conflict of
interest.
(10) Provide Ohio EPA
access to the certified laboratory's facility and
documents, data, or information related to any voluntary action, or laboratory
certification, in order to determine compliance with this chapter and Chapter
3746. of the Revised Code.
(11)
Promptly and completely respond to all document and data requests made by the
director under this chapter and Chapter 3746. of the Revised Code.
(12) As required by
this rule and rule
3745-300-13 of the
Administrative Code, submit by affidavit all information, data, documents, and
reports for use in support of a request for a no further action
letter.
(13) Conduct
laboratory operations in compliance with all applicable federal and state laws,
regulations and rules, including but not limited to, requirements for
management and disposal of samples that meet the definition of "hazardous
waste" in rule
3745-51-03 of the Administrative
Code and other hazardous wastes stored on property in compliance with Chapters
3745-52 and
3745-65 of the Administrative
Code.
(14) Maintain in good
standing any third-party accreditations relied upon for certification.
(F)
All documentation provided
to Ohio EPA in accordance with this rule shall be submitted to Ohio EPA in a
format prescribed by Ohio EPA.
(G) Laboratory audits.
(1) At Ohio EPA's discretion, Ohio EPA shall
audit certified laboratories to determine
compliance with this rule .
Laboratory audits may consist of a review of either documents or other
information submitted to Ohio EPA. Laboratory audits may include an on-site
visit to the laboratory to review the laboratory's operations and to evaluate
the laboratory's facility and personnel.
(a) At
any time and for any purpose, Ohio EPA shall evaluate a certified laboratory to
determine a laboratory's compliance with the laboratory's obligations as a
certified laboratory under this rule and the laboratory's ability to produce
certified data in accordance with this rule. Ohio EPA may conduct this
evaluation for any reason, including but not limited to, the following:
(i) When there is a change in laboratory
personnel, management personnel, operational procedures, or other functional
issue.
(ii)
If Ohio EPA receives a complaint regarding the certified laboratory's performance.
(2) In order to
determine compliance with this rule, an audit of a certified laboratory may include, but is not limited
to, the following:
(a) Review of the
certified laboratory's SOPs, QAPP, analytical
reports and associated data, affidavits, and other documents.
(b) On-site visit and review of the
certified laboratory's sample receiving area,
waste storage area, analytical testing areas, and other pertinent areas of the
certified laboratory. During the visit, Ohio EPA
may review the following:
(i)
Logbooks.
(ii) Sample storage procedures.
(iii) Instrumentation set-up and software
programs.
(iv) Equipment
calibration and maintenance procedures.
(v) Data review procedures.
(vi) Record filing and storage.
(vii) Project management and communication
procedures.
(viii) Data reporting
procedures.
(ix) Record
files.
(x) Any other information or
area of the laboratory deemed appropriate by Ohio EPA.
(c) Interviews of laboratory personnel to
determine knowledge of personnel who perform the analyses.
(d) Review of any other documentation that
Ohio EPA considers appropriate, including, if applicable, review of any
documents related to third-party accreditation relied upon for
certification.
(e) Review of
performance testing results, as required by paragraph (C) of this
rule.
(f) Evaluation of whether any
violations of this rule are material in a laboratory's ability to report
reliable, defensible, and representative data that satisfies the requirements
for certified data under this rule.
(3) Ohio EPA shall prepare an audit report
that indicates any deficiencies that are identified during the audit that
require corrective actions by the laboratory. Failure to address the
deficiencies in a timely manner may result in suspension or revocation of a
laboratory's certification.
(4) If Ohio EPA identifies any deficiencies
during a certified laboratory audit, the
laboratory shall correct those deficiencies to Ohio EPA's satisfaction
.
(H) Laboratory
certifications:
(1)
The certification, issued by the director, may limit the
analysis of certain environmental media.
(2) The certification automatically expires
three years after the date of issuance, unless the laboratory's certification
is suspended
or revoked, prior to the certification's expiration.
(3) The certification
applies only to the individual certified
laboratory facility identified in the certificate.
(4) The effective
certificate shall be displayed in a prominent location in the
certified laboratory.
(5) If a laboratory's
certification is revised, the revised certification supersedes any prior
certification.
(I) Retention of
documents and data.
(1) A
certified laboratory shall maintain all documents
and data prepared or acquired in connection with a voluntary action for a
period of at least ten years after the date that the laboratory's analyses were
submitted to a certified professional or volunteer.
(2) The certified laboratory may retain the documents and data
using any available technology, provided that the laboratory can readily
retrieve the documents and data in legible condition when retrieval is
requested by Ohio EPA during the ten-year retention period.
(3) If a certified laboratory does not intend to retain such
documents and data after ten years, the laboratory shall notify Ohio EPA of
such intent, and shall provide Ohio EPA the opportunity to obtain the documents
and data.
(4) The documents and
data shall be retained until the notice described in paragraph
(I)(3)
of this rule is provided to Ohio EPA, and Ohio EPA notifies the
certified laboratory in writing whether Ohio EPA
shall obtain the documents and data.
(5) Notification of Ohio EPA pursuant to this
paragraph is not required as long as a certified
laboratory continues to retain all documents and data.
(6) Failure to provide documents or data
requested by Ohio EPA may result in permanent revocation of the laboratory's
certification in accordance with paragraph
(L)(3) of this
rule.
(J) Out-of-state
laboratories.
(1) As a condition of
certification under this rule, certified
laboratories located outside the state of Ohio consent to service of
process and to personal jurisdiction of any Ohio court or the Ohio
environmental review appeals commission in proceedings that adjudicate any
rights or obligations under this chapter and Chapter 3746. of the Revised Code,
or in which the cause of action involves, in whole or in part, the laboratory's
performance under this chapter or Chapter 3746. of the Revised Code.
(2) Out-of-state certified laboratories consent to Ohio EPA's right of
entry for inspection or investigation, and to the service of administrative
warrants, inspection warrants, or other appropriate search warrants as a
condition of certification under this rule.
(K)
Appeal of certification determinations. The denial, suspension, or revocation of any
laboratory certification is a final action of the director, which is subject to
the procedure for appeal provided in Chapter 3745. of the Revised
Code.
(L) Revocation or
suspension of certification.
(1) The director
may revoke or suspend a laboratory's certification issued pursuant to this
rule, for a period to be determined by the director, upon finding that a
laboratory failed to comply with paragraph
(E) of this rule,
except as provided in paragraphs
(L)(2) and
(L)(3) of this
rule.
(2) The director may
permanently revoke a laboratory's certification if the laboratory falsifies any
information in connection with the laboratory's certification or any voluntary
action, in violation of paragraph
(E)(8) of this rule.
(3) The director shall permanently revoke a
laboratory's certification if the laboratory does not comply with a request for
documents and data, in violation of paragraph
(E)(11) of
this rule.
(4) If a laboratory's
certification relies upon maintaining third-party accreditation in good
standing, the director may revoke or suspend a laboratory's certification upon
finding that the laboratory's third-party accreditation is no longer maintained
in good standing.
(5) Upon
revocation or suspension of certification, the laboratory shall promptly return
to Ohio EPA the certificate to which the revocation or suspension
applies.
(M) Procedure to
request reinstatement of certification.
(1)
Procedures to request reinstatement of certification after a suspension period
are as follows:
(a) A suspended laboratory
may request to reinstate the laboratory's certification for a suspension issued
because of the laboratory's failure to comply with paragraphs
(E)(1)
to
(E)(7),
(E)(9) to
(E)(10), and
(E)(12) to
(E)(14) of this rule.
(b) After the suspension period, the
laboratory may request reinstatement of the laboratory's certification by
providing
a
written request for
reinstatement and any documentation to demonstrate that the laboratory resolved
all findings which resulted in the suspension.
(2) Pursuant to rule
3745-300-03 of the
Administrative Code, the certified laboratory is
required to pay any costs incurred by Ohio EPA to review requests for
reinstatement.
(3) If Ohio EPA
conducts a certified laboratory audit of the
laboratory the laboratory shall do one of the following:
(a) Perform acceptably on the
audit.
(b) Prior to reinstatement
of the laboratory's certification, shall correct any deficiencies that are
identified during the audit. Laboratory audits shall be conducted consistent
with paragraph
(G) of this rule.
(4) A
certification shall expire
on
the date listed in the original issuance.
(N)
Procedures to request modifications to certifications.
(1) A laboratory shall request a modification
to the laboratory's certificate to reflect changes in company name or
address through use of a cover letter
.