(A) A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during source replacement , to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed five years for each teletherapy unit and shall not exceed seven years for each gamma stereotactic radiosurgery unit.

(B) This inspection and servicing may only be performed by persons specifically licensed to do so by the director, the United States nuclear regulatory commission, or an agreement state.

(C) A licensee shall keep a record of the inspection and servicing in accordance with rule 3701:1-58-100 of the Administrative Code.

Disclaimer: These regulations may not be the most recent version. Ohio may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.