Ohio Administrative Code
Title 3701:1 - Radiation Control
Chapter 3701:1-58 - Medical Use of Radioactive Materials
Section 3701:1-58-66 - Periodic spot-checks for gamma stereotactic radiosurgery units

Universal Citation: OH Admin Code 3701:1-58-66

Current through all regulations passed and filed through March 18, 2024

(A) A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform spot-checks of each gamma stereotactic radiosurgery facility and on each unit:

(1) Monthly;

(2) Before the first use of the unit on a given day; and

(3) After each source installation.

(B) A licensee shall:

(1) Perform the measurements required by paragraph (A) of this rule in accordance with written procedures established by the authorized medical physicist. That individual need not actually perform the spot-check measurements.

(2) Have the authorized medical physicist review the results of each spot-check within fifteen days. The authorized medical physicist shall notify the licensee as soon as possible in writing of the results of each spot-check.

(C) To satisfy the requirements of paragraph (A)(1) of this rule, spot-checks must, at a minimum:

(1) Assure proper operation of:
(a) Treatment table retraction mechanism, using backup battery power or hydraulic backups with the unit off;

(b) Helmet microswitches;

(c) Emergency timing circuits; and

(d) Stereotactic frames and localizing devices (trunnions).

(2) Determine:
(a) The output for one typical set of operating conditions measured with the dosimetry system described in paragraph (B) of rule 3701:1-58-60;

(b) The difference between the measurement made in paragraph (C)(2)(a) of this rule and the anticipated output, expressed as a percentage of the anticipated output (i.e., the value obtained at last full calibration corrected mathematically for radioactive decay);

(c) Source output against computer calculation;

(d) Timer accuracy and linearity over the range of use;

(e) On-off error; and

(f) Trunnion centricity.

(D) To satisfy the requirements of paragraphs (A)(2) and (A)(3) of this rule, spot-checks must assure proper operation of:

(1) Electrical interlocks at each gamma stereotactic radiosurgery room entrance;

(2) Source exposure indicator lights on the gamma stereotactic radiosurgery unit, on the control console, and in the facility;

(3) Viewing and intercom systems;

(4) Timer termination;

(5) Radiation monitors used to indicate room exposures; and

(6) Emergency off buttons.

(E) A licensee shall arrange for the repair of any system identified in paragraph (C) of this rule that is not operating properly as soon as possible.

(F) If the results of the checks required in paragraph (D) of this rule indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.

(G) A licensee shall retain a record of each check required by paragraphs (C) and (D) of this rule and a copy of the procedures required by paragraph (B) of this rule in accordance with rule 3701:1-58-97 of the Administrative Code.

Five Year Review (FYR) Dates: 06/12/2015 and 06/01/2020
Promulgated Under: 119.03
Statutory Authority: 3743.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 8/15/2005

Disclaimer: These regulations may not be the most recent version. Ohio may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.