Ohio Administrative Code
Title 3701:1 - Radiation Control
Chapter 3701:1-58 - Medical Use of Radioactive Materials
Section 3701:1-58-27 - Requirements for possession of sealed sources and brachytherapy sources

Universal Citation: OH Admin Code 3701:1-58-27

Current through all regulations passed and filed through March 18, 2024

(A) A licensee in possession of any sealed source or brachytherapy source shall follow the radiation safety and handling instructions supplied by the manufacturer.

(B) A licensee in possession of a sealed source shall:

(1) Test the source for leakage before its first use unless the licensee has a certificate from the supplier indicating that the source was tested within six months before transfer to the licensee; and

(2) Test the source for leakage at intervals not to exceed six months or at other intervals approved by the director, United States nuclear regulatory commission, or an agreement state.

(C) To satisfy the leak test requirements of this rule, the licensee shall measure the sample so that the leak test can detect the presence of one hundred eighty-five becquerels (0.005microcurie) of radioactive material in the sample.

(D) A licensee shall retain leak test records in accordance with paragraph (A) of rule 3701:1-58-80 of the Administrative Code.

(E) If the leak test reveals the presence of one hundred eighty-five becquerels (0.005

microcurie or more of removable contamination, the licensee shall:
(1) Immediately withdraw the sealed source from use and store, dispose, or cause it to be repaired in accordance with the requirements in Chapter 3701:1-38 of the Administrative Code, and

(2) File a report within five days of the leak test in accordance with rule 3701:1-58-103 of the Administrative Code.

(F) A licensee need not perform a leak test on the following sources:

(1) Sources containing only radioactive material with a half-life of less than thirty days;

(2) Sources containing only radioactive material as a gas;

(3) Sources containing 3.7 megabecquerels (one hundred microcuries) or less of beta or gamma-emitting material or 0.37megabecquerel (ten microcuries) or less of alpha-emitting material;

(4) Seeds of iridium-192 encased in nylon ribbon; and

(5) Sources stored and not being used. However, the licensee shall test each such source for leakage before any use or transfer unless it has been leak tested within six months before the date of use or transfer.

(G) A licensee in possession of sealed sources or brachytherapy sources, except for gamma stereotactic radiosurgery sources, shall conduct a semi-annual physical inventory of all such sources in its possession. The licensee shall retain each inventory record in accordance with paragraph (B) of rule 3701:1-58-80 of the Administrative Code.

Five Year Review (FYR) Dates: 06/12/2015 and 06/01/2020
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 8/15/2005, 10/4/10

Disclaimer: These regulations may not be the most recent version. Ohio may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.