Ohio Administrative Code
Title 3701:1 - Radiation Control
Chapter 3701:1-58 - Medical Use of Radioactive Materials
Section 3701:1-58-23 - Possession, use, and calibration of instruments used to measure the activity of unsealed radioactive material
Current through all regulations passed and filed through March 18, 2024
(A) For direct measurements performed in accordance with rule 3701:1-58-25 of the Administrative Code, a licensee shall possess and use instrumentation to measure the activity of unsealed radioactive material before it is administered to each patient or human research subject.
(B) A licensee shall calibrate the instrumentation required in paragraph (A) of this rule in accordance with nationally recognized standards or the manufacturer's instructions.
(C) A licensee shall retain a record of each instrument calibration required by this rule in accordance with rule 3701:1-58-77 of the Administrative Code.