Subchapter B - INJURY CONTROL
Section .0300 - BREATH ALCOHOL TEST REGULATIONS
Section 41B .0322 - INTOXIMETERS: MODEL INTOX EC/IR II AND MODEL INTOX EC/IR II (ENHANCED WITH SERIAL NUMBER 10,000 OR HIGHER)
Current through Register Vol. 39, No. 6, September 16, 2024
The following operational procedures shall be followed when administering a test using the Intoximeters, Model Intox EC/IR II and Model Intox EC/IR II (Enhanced with serial number 10,000 or higher):
(1) Ensure instrument displays time and date;
(2) Ensure observation period requirements have been met;
(3) Initiate breath test sequence;
(4) Enter information as prompted by the instrument;
(5) Verify instrument accuracy;
(6) When "PLEASE BLOW" appears, collect breath sample;
(7) When "PLEASE BLOW" appears, collect breath sample; and
(8) Print test record.
If the alcohol concentrations differ by more than 0.02, a third or fourth breath sample shall be collected when "PLEASE BLOW" appears. Sequential tests shall be administered as soon as feasible by repeating steps (1) through (8), as applicable.