Section .0300 - BREATH ALCOHOL TEST REGULATIONS
Section 41B .0322 - INTOXIMETERS: MODEL INTOX EC/IR II AND MODEL INTOX EC/IR II (ENHANCED WITH SERIAL NUMBER 10,000 OR HIGHER)

Universal Citation: 10A NC Admin Code 41B .0322

Current through Register Vol. 39, No. 6, September 16, 2024

The following operational procedures shall be followed when administering a test using the Intoximeters, Model Intox EC/IR II and Model Intox EC/IR II (Enhanced with serial number 10,000 or higher):

(1) Ensure instrument displays time and date;

(2) Ensure observation period requirements have been met;

(3) Initiate breath test sequence;

(4) Enter information as prompted by the instrument;

(5) Verify instrument accuracy;

(6) When "PLEASE BLOW" appears, collect breath sample;

(7) When "PLEASE BLOW" appears, collect breath sample; and

(8) Print test record.

If the alcohol concentrations differ by more than 0.02, a third or fourth breath sample shall be collected when "PLEASE BLOW" appears. Sequential tests shall be administered as soon as feasible by repeating steps (1) through (8), as applicable.

Authority G.S. 15A-534.2(d); 20-139.1(b);
Eff. November 1, 2007;
Pursuant to G.S. 150B-21.3A, rule is necessary without substantive public interest Eff. January 9, 2018;
Amended Eff. April 1, 2020.

Disclaimer: These regulations may not be the most recent version. North Carolina may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.