North Carolina Administrative Code
Title 10A - HEALTH AND HUMAN SERVICES
Chapter 15 - RADIATION PROTECTION
Section .0900 - REQUIREMENTS FOR PARTICLE ACCELERATORS
Section 15 .0908 - OPERATING PROCEDURES
Current through Register Vol. 39, No. 6, September 16, 2024
(a) Particle accelerators, when not in operation, shall be secured to prevent unauthorized use.
(b) Only a switch on the accelerator control console shall be routinely used to turn the accelerator beam "on" and "off". The safety interlock system shall not be used to turn off the accelerator beam except in an emergency.
(c) All safety and warning devices, including interlocks shall be checked for proper operability at least every six months unless more frequent checks are required by the agency. Results of such tests shall be maintained for two years at the accelerator facility for inspection by the agency.
(d) Electrical circuit diagrams of the accelerator, and the associated interlock systems, shall be kept current and maintained for inspection by the agency.
(e) If, for any reason, it is necessary to intentionally bypass a safety interlock or interlocks, such action shall be:
(f) A copy of the current operating and the emergency procedures shall be maintained at the accelerator control panel.
Authority
G.S.
104E-7;
Eff. February 1,
1980;
Transferred and Recodified from
15A NCAC
11.0908 Eff. February 1,
2015.