North Carolina Administrative Code
Title 10A - HEALTH AND HUMAN SERVICES
Chapter 15 - RADIATION PROTECTION
Section .0800 - REQUIREMENTS FOR NON-HUMAN USE OF RADIATION GENERATING DEVICES
Section 15 .0803 - EQUIPMENT REQUIREMENTS
Current through Register Vol. 39, No. 6, September 16, 2024
(a) Certified cabinet x-ray systems shall meet the requirements of 21 CFR 1020.40 as incorporated by reference in Rule .0117(a)(3) of this Chapter.
(b) All certified and certifiable cabinet x-ray systems shall:
(c) Open-beam analytical RGD systems shall be equipped with a safety device that prevents the entry of any portion of an individual's body into the primary x-ray beam path that causes the beam to be shut off upon entry into its path.
(d) Open-beam analytical RGDs shall be provided with a visible and legible indication of:
(e) Warning devices on open-beam analytical RGDs shall be labeled so that their purpose is identified. On open-beam analytical RGDs installed after February 1, 1980, warning devices and lights shall have fail-safe characteristics.
(f) Unused ports on radiation source housings for open-beam RGDs shall be secured in the closed position in a manner that will prevent unintended opening.
(g) Each port on the radiation source housing on open-beam analytical RGDs installed after February 1, 1980 and designed to accommodate interchangeable components shall be equipped with a shutter that cannot be opened unless a collimator or a component coupling is connected to the port.
(h) Portable open-beam
analytical RGDs that shall be manufactured to be used hand-held without safety
devices are exempt from the requirements of Paragraph (c) of this Rule and
shall be constructed according to International Standard IEC 62495 that is
incorporated by reference and includes subsequent amendments. This standard can
be downloaded for one hundred twenty-one dollars ($121.00) at the following
website:
(i) A registrant may apply to the agency, as defined in Rule.0104 of this Chapter, for an exemption from the requirement of a safety device. This request shall include:
(j) Analytical RGDs shall be provided with a visible and legible label(s) bearing the radiation symbol and the words:
(k) Warning lights labeled with the words "X-RAYS ON," or other words having similar meaning, shall be located:
(l) Warning lights shall activate when the x-ray tube is energized.
(m) Each x-ray tube housing shall be:
(n) Each x-ray generator shall be supplied with a protection cabinet which limits leakage radiation measured at a distance of five centimeters from its surface such that it is not capable of producing an exposure in excess of 0.25 mrem/2.5µSv in one hour.
(o) Permanent radiographic installations and industrial radiography RGDs shall comply with the requirements of Rule.0807 of this Section.
Authority
G.S.
104E-7;
Eff. February 1,
1980;
Transferred and Recodified from
15A NCAC
11.0803 Eff. February 1, 2015;
Amended Eff. October 1, 2015;
Pursuant to
G.S.
150B-21.3A, rule is necessary without
substantive public interest Eff. June 22,
2019.