Section 15 .0604 - GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC SYSTEMS
Current through Register Vol. 39, No. 6, September 16, 2024
(a) In addition to other requirements of this Section, all diagnostic x-ray systems shall meet the following requirements:
X-Ray Tube Voltage |
(kilovolt peak) |
Minimum HVL (millimeters of Aluminum) |
Minimum HVL (millimeters of Aluminum) |
Designed operating range |
Measured Operating Potential |
Specified Dental Systems |
Other X-ray Systems |
Below 50------------ |
30 |
1.5 |
0.3 |
40 |
1.5 |
0.4 |
|
49 |
1.5 |
0.5 |
|
50 to 70------------ |
50 |
1.5 |
1.2 |
60 |
1.5 |
1.2 |
|
70 |
1.5 |
1.5 |
|
Above 70------------ |
71 |
2.1 |
2.1 |
80 |
2.3 |
2.3 |
|
90 |
2.5 |
2.5 |
|
100 |
2.7 |
2.7 |
|
110 |
3.0 |
3.0 |
|
120 |
3.2 |
3.2 |
|
130 |
3.5 |
3.5 |
|
140 |
3.8 |
3.8 |
|
150 |
4.1 |
4.1 |
If it is necessary to determine such half-value layer at an x-ray tube potential which is not listed in the table, linear interpolation or extrapolation may be made. Positive means shall be provided to insure that at least the minimum filtration needed to achieve the above beam quality requirements is in the useful beam during each exposure.
Filtration Required versus Operating Voltage |
|
Operating Voltage (kVp) |
Minimum total filtration (inherent plus added) |
(millimeters aluminum equivalent) |
|
Below 50 |
0.5 millimeters |
50 - 70 |
1.5 millimeters |
Above 70 |
2.5 millimeters |
(b) Structural Shielding