New York Codes, Rules and Regulations
Title 10 - DEPARTMENT OF HEALTH
Chapter I - State Sanitary Code
Part 16 - Ionizing Radiation
Radiation Equipment
Section 16.61 - Therapy equipment operating at potentials of 60 kvp and below
Universal Citation: 10 NY Comp Codes Rules and Regs ยง 16.61
Current through Register Vol. 46, No. 39, September 25, 2024
(a) Equipment. All provisions of subdivision (a) of section 16.60 of this Part shall apply except that the leakage five centimeters from the surface of the tube housing shall not exceed 0.1 roentgen per hour.
(b) Conditions for operation of equipment.
(1) The output of the X-ray generator shall
be calibrated prior to the use of the apparatus for treating humans.
Calibration shall be performed by a person qualified pursuant to section
16.122(f)
of this Part and experienced in the calibration of such units. The method of
calibration used shall ensure accurate delivery of the prescribed dose under
all conditions of use. Calibrations shall be made at least annually.
Recalibration, however, shall be made after each tube replacement and after any
changes or replacements in the generating apparatus which could change the
X-ray output.
(2) If the tube must
be hand-held during irradiation, the operator shall wear protective gloves and
a protective apron of no less than 0.5 mm lead equivalent.
(3) The operator shall be able to observe and
communicate with the patient during irradiation.
(4) Equipment having an output of more than
1,000 roentgens per minute at any accessible place shall not be left unattended
without the power being shut off at the main disconnect switch in addition to
the control panel switch.
(5) When
operating equipment constructed with beryllium or other low filtration windows
the operator shall insure that the useful beam is blocked at all times except
when actually being used.
Disclaimer: These regulations may not be the most recent version. New York may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.