New York Codes, Rules and Regulations
Title 10 - DEPARTMENT OF HEALTH
Chapter I - State Sanitary Code
Part 16 - Ionizing Radiation
Radiation Equipment
Section 16.60 - Therapy equipment operated at potentials up to 10 million volts.see section 16.61 for special requirements for equipment operating below 60 kvp
Universal Citation: 10 NY Comp Codes Rules and Regs ยง 16.60
Current through Register Vol. 46, No. 39, September 25, 2024
(a) Equipment.
(1) The protective tube housing shall be of
therapeutic type.
(2) Fixed
diaphragms or cones used to restrict the useful beam shall be so constructed as
to provide the same degree of protection as is required of the tube
housing.
(3) Adjustable or
removable beam limiting diaphragms or cones shall not transmit more than five
percent of the useful beam at the maximum kilovoltage and with the maximum
treatment filter.
(4) The filter
system shall be so arranged as to minimize the possibility of error in filter
selection and alignment. The filter slot shall be so constructed that the
radiation escaping through it does not produce an exposure exceeding one
roentgen per hour at one meter, or if the patient is likely to be exposed to
radiation escaping from the slot, 30 roentgens per hour at two inches (5
centimeters) from the external opening. Each removable filter shall be marked
with its thickness and material.
(5) It shall be possible for the person
operating the controls of the therapeutic equipment to determine from the
operating position what filters are in place in the equipment. Filters shall be
so mounted as to prevent their movement during the treatment.
(6) The X-ray tube shall be secured so that
it cannot move in respect to the housing aperture. A mark on the exterior of
the tube housing is recommended to indicate the focal spot.
(7) Adequate devices shall be provided to
secure the tube housing during stationary portal treatment.
(8) An easily discernible indicator which
shows whether or not X-rays are being produced shall be on the control
panel.
(9) A suitable exposure
control device (e.g., an automatic timer, exposure meter or dose meter) shall
be provided to terminate the exposure automatically after a preset time
interval or preset exposure or dose limit. Means shall be provided for the
operator to terminate the exposure at any time.
(10) Equipment utilizing shutters to control
the useful beam shall have a shutter position indicator on the control
panel.
(b) Structural shielding.
(1) The protective barriers for
all therapy equipment operating at voltages above 60 kVp shall be fixed to the
wall of the building.
(2) The
control panel for all therapy equipment operating at voltages above 150 kVp
installed after January 1, 1963, shall be located outside the treatment
room.
(c) Conditions for operation of equipment.
(1) The output of the
X-ray generator shall be calibrated prior to the use of the apparatus for
treating humans. Calibration shall be performed by a person qualified pursuant
to section
16.122(f)
of this Part. The method of calibration used shall be in accordance with
procedures recommended by the American Association of Physicists in Medicine
for the energy and type of radiation employed. Calibration shall be made at
least annually. Recalibration, however, shall be made after each tube
replacement and after any changes or replacements in the generating apparatus,
or changes or updates in computer programs which govern or interact with the
functions of the machine and which could change the X-ray output.
(2) No person other than the patient, shall
be permitted to remain within the X-ray therapy room while the X-ray generator
is in operation.
(3) Every entrance
to an X-ray therapy room in which equipment is capable of operating above 150
kV shall be protected by interlocks to insure that during the production of
X-rays no person can enter the therapy room without turning off the radiation
equipment. They shall be so arranged that irradiation equipment cannot be
started again without manually resetting the controls.
(4) Windows, mirror systems, or
closed-circuit television viewing screens shall be provided to permit
continuous observation of the patient during irradiation and shall be so
located that the operator may see the patient and the control panel from the
same position.
(5) Provision shall
be made for oral communication with the patient.
(6) In addition to the interlocking controls,
there shall be installed signals which are readily observable or discernible
near the outside of all access doors to indicate the production of X-rays.
[FN11] See section 16.61 for special requirements for equipment operating below 60 kVp.
Disclaimer: These regulations may not be the most recent version. New York may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.