New York Codes, Rules and Regulations
Title 10 - DEPARTMENT OF HEALTH
Chapter I - State Sanitary Code
Part 16 - Ionizing Radiation
Radiation Equipment
Section 16.59 - Use of Computed Tomography Equipment
Universal Citation: 10 NY Comp Codes Rules and Regs ยง 16.59
Current through Register Vol. 46, No. 39, September 25, 2024
(a) Definitions
(1) "Computed tomography (CT)" scan and
"computerized axial tomography (CAT)" scan refer to an imaging procedure that
uses x-rays to create cross-sectional images of the human body.
(2) "Computed tomography dose index" (CTDI)
means the integral of the dose profile along a line perpendicular to the
tomographic plane divided by the product of the nominal tomographic section
thickness and the number of tomograms produced in a single scan where the dose
profile is centered around z = 0 and for a multiple tomogram system, the scan
increment between adjacent scans is nT;
z = position along a line perpendicular to the tomographic plane;
D (z) = Dose at position z;
T = Nominal tomographic section thickness;
n = Number of tomograms produced in a single scan
(3) "CT x-ray system" is
technology that is used to perform CT scans and includes but is not limited to,
a control panel, image display device, gantry, x-ray tube, collimating device
with filters, high voltage transformer and a data acquisition system.
(4) "CT scanner" refers to technology used to
perform and interpret CT scans and includes, but is not limited to, a control
panel, gantry, high voltage generator, x-ray tube, table and display devices
that are used for image interpretation.
(5) "CTDI100" is the
dose measurement made with a 16cm diameter (head/pediatric body) or a 32cm
diameter (body) acrylic phantom. The measurements are made utilizing a 100mm
long pencil ionization chamber. Readings are made with the ion chamber in both
the center (axial or central dose) position and near surface slots of the
phantom (the peripheral dose).
(6)
"CTDIw", the weighted or blended dose, is calculated by
adding together two-thirds of the CTDI100 peripheral
dose with one-third of the CTDI100 axial or center dose.
(CTDIw = 2/3 CTDI100 peripheral +
1/3 CTDI100 axial or center). CTDIw represents an
average dose in the x and y planes.
(7) "CTDIvol"
represents the integrated dose over the total volume that is irradiated,
CTDIvol = (1/PITCH) x (CTDIw),
where "Pitch" is defined as the table travel per rotation divided by the
collimation of the x-ray beam. CTDIvol represents the
average dose in the x, y and z planes.
(8) "CT conditions of operation" means all
selectable parameters governing the operation of a CT x-ray system including
nominal tomographic section thickness, filtration, and the technique
factors.
(9) "CT dosimetry phantom"
means the phantom used for determination of the dose delivered by a CT x-ray
system. The phantom shall be a right circular cylinder of
polymethyl-methacrylate of density 1.19±0.01 grams per cubic centimeter. The
phantom shall be at least 14 centimeters in length and shall have diameters of
32.0 centimeters for testing any CT system designed to image any section of the
body (whole body scanners) and 16.0 centimeters for any system designed to
image the head (head scanners) or for any whole body scanner operated in the
head scanning mode. The phantom shall provide means for the placement of a
dosimeter(s) along its axis of rotation and along a line parallel to the axis
of rotation 1.0 centimeter from the outer surface and within the
phantom.
(10) "Dose length product"
(DLP) is defined as the CTDIvol times the irradiated
length of the body for the whole series of images that are taken during a CT
scan.
(11) "Picture Archiving and
Communication System (PACS)" is a medical imaging technology that provides
access to and storage for medical images from multiple modalities. It is
comprised of an image acquisition system, display, network and data storage or
archiving system.
(12) "Reference
plane" means a plane which is displaced from and parallel to the tomographic
plane.
(13) "Scan" means the
complete process of collecting x-ray transmission data for the production of a
tomogram or a series of tomograms.
(14) "Scan increment" means the amount of
relative displacement of the patient with respect to the CT x-ray system
between successive scans measured along the direction of such
displacement.
(15) "Technique"
means the settings selected on the control panel of the equipment and may
include the position of the x-ray tube, image intensifier and
patient.
(16) "Technique chart"
means a chart that lists the standard settings and positions for a given
technique.
(17) "Tomogram" is an
image of a tissue plane or section of tissue.
(18) "Tomographic plane" means that geometric
plane which the manufacturer identified as corresponding to the output
tomogram.
(19) "Tomographic
section" means the volume of an object whose x-ray attenuation properties are
imaged in a tomogram.
(b) CT X-Ray System Equipment Requirements.
(1) Each control panel and gantry of a CT
x-ray system shall include visual signals that indicate to the operator of the
CT x-ray system whenever x-rays are being produced and when x-ray production is
terminated, and, if applicable, whether the shutter is open or
closed.
(2) Each CT x-ray system
shall be equipped with a control that allows the operator of the CT x-ray
system to terminate the x-ray exposure at any time during a scan, or series of
scans, when the exposure time is greater than one-half second
duration.
(3) Each CT x-ray system
shall be designed such that the CT conditions of operation to be used during a
scan or a scan sequence are indicated prior to the initiation of a scan or a
scan sequence.
(4) Each CT x-ray
system shall include a clearly and conspicuously labeled emergency shutoff
button or switch.
(5) Premature
termination of the x-ray exposure by the operator shall necessitate resetting
of the CT conditions of operation by the operator prior to the initiation of
another scan.
(c) Patient communication and viewing requirements.
(1) Each CT x-ray system shall be equipped to
allow two-way aural communication between the patient and the operator at the
control panel.
(2) Each CT x-ray
system shall be equipped with windows, mirrors, closed-circuit television, or
an equivalent to permit continuous visual observation of the patient during CT
scanning by the CT operator from the control panel.
(3) When the primary viewing system is by
electronic means, an alternate viewing system (which may be electronic) shall
be available for use in the event of failure of the primary viewing
system.
(d) Calibration.
(1) Each registrant shall ensure that the
calibration of the radiation output of each CT x-ray system that it operates is
performed by, or under the direction of, a licensed medical
physicist.
(2) Each registrant
shall maintain and make available for review by the Department, on the premises
of its radiation installation where a CT x-ray system is located written
procedures for the appropriate calibration of the CT x-ray system.
(3) After initial installation, the CT x-ray
system shall be calibrated prior to its use on human beings and recalibrated at
least within every 14 months thereafter. Any change or replacement of
components of a CT x-ray system which could cause a change in the radiation
output will require a recalibration within 30 days of component installation by
a licensed medical physicist operating within their scope of
practice.
(4) The calibration of
the radiation output of a CT x-ray system shall be performed with a calibrated
dosimetry system. This system shall have been calibrated either by the National
Institute of Standards and Technology (NIST) or by an American Association of
Physicists in Medicine (AAPM) Accredited Dosimetry Calibration Laboratory
(ADCL) and traceable to NIST. The calibration shall have been performed within
the previous 24 months and after any servicing that might have affected system
calibration.
(5) CT dosimetry
phantom(s) shall be used in determining the radiation output of each CT x-ray
system. Such phantom(s) shall meet the following specifications and conditions
of use:
(i) Any effects on the doses measured
because of the removal of phantom material to accommodate dosimeters shall be
accounted for through appropriate corrections to the reported data or included
in the statement of maximum deviation for the values obtained using the
phantom; and
(ii) All dose
measurements shall be performed with the CT dosimetry phantom placed on the
patient couch or support device without additional attenuation materials
present.
(iii) The requirements of
subparagraphs (i) and (ii) of this paragraph can also be met by using an
alternative method of radiation measurement and calculation published in the
peer-reviewed scientific literature and acceptable to the Department.
(6) Records of calibrations
performed shall be maintained for a period of three (3) years at the radiation
installation where the CT is located.
(e) Quality Assurance Testing
(1) Each registrant shall maintain a Quality
Assurance (QA) manual that shall contain written procedures for all testing and
shall meet the requirements specified in this section and section
16.23(a)(1).
The CT Quality Assurance procedures shall have been developed under the
direction of a licensed medical physicist or radiologist.
(2) The QA procedures shall incorporate the
use of one or more image quality dosimetry phantoms or the phantom supplied by
the original equipment manufacturer which have the capability of providing an
indication of contrast scale, noise, nominal tomographic section thickness, the
resolution capability of the system for low and high contrast objects, and
measuring the mean CT Number for water or other reference material. All of
these image quality parameters shall be evaluated at least annually by a
licensed medical physicist.
(3)
Written records of the QA checks performed by the registrant shall be
maintained for review by the Department for a period of at least three (3)
years.
(4) QA checks shall include
the following:
(a) Images obtained with the
CT dosimetry phantom(s) using the same processing mode and CT conditions of
operation as are used to perform calibrations. The images shall be retained as
photographic copies or as electronic copies stored within the CT x-ray system
or stored on the PACS.
(b) Dose
assessment for the most common CT examinations that are performed on the system
for which reference levels have been published by the American College of
Radiology (ACR), the American Association of Physicists in Medicine (AAPM) or
the National Council on Radiation Protection and Measurements (NCRP) for
pediatric heads, pediatric abdomens, adult heads and adult abdomens.
(c) An evaluation of image quality.
(f) Operating Procedures and Policies
(1) The CT x-ray
system shall not be operated on a human being except by a physician or by a
radiologic technologist licensed pursuant to Article 35 of the Public Health
Law who has been specifically trained in its operation.
(2) The registrant shall ensure that each CT
x-ray system has a radiation protection survey or other measurement and
assessment of exposure to persons in controlled and non-controlled areas made
at the time of installation. Additional radiation protection surveys shall be
done after any change in the radiation installation or equipment which might
cause a significant increase in radiation hazard.
(3) Each CT x-ray system shall have available
at the control panel written information regarding the operation and
calibration of the CT x-ray system. Such information shall include:
(i) Dates of the latest calibration and QC
checks and the location within the facility where the results of those tests
may be obtained;
(ii) Instructions
on the use of the CT dosimetry phantom(s) including a schedule of QC tests that
are appropriate for the system as determined by the manufacturer, allowable
variations for the indicated parameters, and the results of at least the most
recent spot checks conducted on the system;
(iii) A current set of default protocols are
available at the control panel (either electronically or as a document) which
specifies for each routine examination the CT conditions of operation and the
slice thickness, spacing between slices and/or pitch;
(iv) A list of techniques optimized for the
body part being imaged to obtain a quality image and to ensure that the lowest
amount of radiation is used as consistent with good medical practice.
(4) If the QC testing on the CT
x-ray system identifies that a system operating parameter has exceeded a
tolerance as specified in the Quality Assurance manual, use of the CT x-ray
system on patients shall be limited to those exceptions permitted by
established written instructions of the licensed medical physicist or
radiologist. Upon completion of corrective action, the QC testing shall be
repeated to verify that the system is back within tolerance.
(5) Commencing one (1) year after the
effective date of these regulations, each registrant performing CT scans on
human beings shall ensure that for each scan, the radiation dose delivered by
the scanner to a reference phantom or the dose received by the patient is saved
and recorded. The dose delivered shall be recorded as Computed Tomography Dose
Index volume (CTDIvol), dose length product (DLP) or
other dosimetry metric published in the peer reviewed scientific literature and
acceptable to the Department. The dose received by a patient shall be recorded
as organ dose or other dosimetry metric published in the peer reviewed
scientific literature and acceptable to the Department.
(6) The displayed dose shall be verified on
an annual basis by or under the supervision of a licensed medical physicist to
ensure that the equipment manufacturer's displayed dose is within 20% of the
measured dose.
(7) Eighteen months
after the effective date of these regulations, each current registrant that
performs diagnostic CT scans on human beings shall be accredited by a
nationally recognized accreditation program that is acceptable to the
Department. A facility performing CT that loses their existing accreditation or
a registrant or licensee that fails to obtain accreditation must report this
fact within 30 days to the Department. After the effective date of these
regulations new licensees or registrants will have 18 months to become
accredited, but must demonstrate that they have initiated the accreditation
process within 90 days of the start of operations.
(8) Each registrant that performs CT scans on
human beings shall establish and implement a policy and a procedure to ensure
that:
(i) a request for a CT scan originates
from a physician or other authorized health care practitioner familiar with the
patient's clinical condition; and
(ii) the request includes sufficient
information to demonstrate the medical indication for the CT examination and
allow for the proper performance and interpretation of the CT scan.
Disclaimer: These regulations may not be the most recent version. New York may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.