New Mexico Administrative Code
Title 20 - ENVIRONMENTAL PROTECTION
Chapter 3 - RADIATION PROTECTION
Part 6 - X-RAYS IN HEALING ARTS
Section 20.3.6.607 - THERAPEUTIC X-RAY INSTALLATIONS
Universal Citation: 20 NM Admin Code 20.3.6.607
Current through Register Vol. 35, No. 18, September 24, 2024
A. Equipment:
(1) The protective tube housing
shall be of therapeutic type.
(2)
Permanent diaphragms or cones for collimating the useful beam shall afford the
same degree of protection as the tube housing. Adjustable or removable
beam-defining diaphragms or cones shall transmit not more than five percent of
the useful beam at the maximum kilovoltage and with maximum treatment
filter.
(3) Filters shall be
secured in place to prevent them from dropping out during treatment. The filter
slot shall be so constructed that the radiation escaping through it does not
exceed 1 R (0.258 mC/kg) per hour at 1 m (3.28 feet), or, if the radiation from
the slot is accessible to the patient, 30 R (7.74 mC/kg) per hour at 5 cm (1.96
inches) from the external opening. Each removable filter shall be marked with
its thickness and material.
(4) A
filter indication system shall be used on all therapy machines using changeable
filters. It shall be designed so as to permit easy recognition of any added
filter in place. It shall indicate, from the control panel, the presence or
absence of any filter.
(5) The
x-ray tube shall be so mounted that it cannot turn or slide with respect to the
housing aperture.
(6) Means shall
be provided to immobilize the tube housing during stationary portal
treatment.
(7) A device (e.g., an
automatic timer exposure meter or dose meter) shall be provided to terminate
the exposure after a preset time interval or preset exposure of dose limit.
Means shall be provided for the operator to terminate the exposure at any
time.
(8) Equipment utilizing
shutters to control the useful beam shall have a shutter position indicator on
the control panel.
(9) The control
panel shall include a device (usually an ammeter) which will give positive
indication of the production of x-rays whenever the x-ray tube is
energized.
B. Structural Shielding:
(1) All walls, floors, and
ceilings that can be struck by the useful beam shall be provided with primary
barriers to the height of the ceiling. Low-voltage superficial therapy units
only require a height of 2.1 m (6.88 feet).
(2) All walls, floors, and ceilings that,
because of restrictions in the orientation of the useful beam, cannot be struck
by the useful beam shall be provided with secondary barriers to a minimum
height of 2.1 m (6.88 feet).
(3)
With equipment operating at voltages above 50 kVp, the required barriers shall
be an integral part of the building.
(4) With equipment operating above 150 kVp,
the control panel shall be within a protective booth equipped with an
interlocked door, or located outside the treatment room.
(5) Interlocks shall be provided for x-ray
therapy equipment capable of operating above 150 kVp so that, when any door of
the treatment room is opened, either the machine will shut off automatically or
the radiation level within the room will be reduced to an average of not more
than 2 mR (20 mSv) per hour and a maximum of 10 mR (100 mSv) per hour at a
distance of 1 m (3.28 feet) in any direction from the target. After such
shutoff or reduction in output, it shall be possible to restore the machine to
full operation only from the control panel.
(6) Windows, mirror systems, or
closed-circuit television viewing screen shall be provided to permit continuous
observation of the patient during irradiation and shall be so located that the
operator may see the patient and the control panel from the same
position.
(7) Provision shall be
made for oral communication with the patient from the control room.
(8) Treatment rooms to which access is
possible through more than one entrance shall be provided with flashing warning
lights in a readily observable position near the outside of all access doors,
which will indicate when the useful beam is "on".
C. Operating Procedure:
(1) All new facilities, and existing
facilities not previously surveyed, shall have a protection survey made by, or
under the direction of a qualified expert. This also shall be done after any
change in the facility which might produce a radiation hazard. The expert shall
report his findings in writing to the person in charge of the facility and a
copy of this report shall be transmitted to the Department.
(2) The facility shall be operated in
compliance with any limitations indicated by the protection survey.
(3) When a patient must be held in position
for radiation therapy, mechanical supporting or restraining devices shall be
used whenever feasible, If the patient must be held by an individual, that
individual shall be adequately protected and shall be positioned so that no
part of the body will be struck by the useful beam and that the body is as far
as possible from the edge of the useful beam. The exposure of any individual
used for this purpose shall be monitored.
(4) The output of each therapeutic x-ray
machine shall be calibrated by, or under the direction of, a qualified expert.
The calibration shall be repeated after any change in or replacement of
components of the x-ray generating equipment which could cause a change in
x-ray output. Check calibrations shall be made at least once a year thereafter.
Records of calibration shall be maintained by the registrant.
Disclaimer: These regulations may not be the most recent version. New Mexico may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.