B. Radiation Exposure Control Devices:
(1) Timers: Means shall be provided to
terminate the exposure at a preset time interval, preset product of current and
time, a preset number of pulses, or a preset radiation exposure to the image
receptor. In addition:
(a) termination of
exposure shall cause automatic resetting of the timer to its initial setting or
to zero; and
(b) it shall not be
possible to make an exposure when the timer is set to a zero or off position if
either position is provided.
(2) X-ray control (exposure switch):
(a) A control shall be incorporated into each
x-ray system such that an exposure can be terminated at any time except for:
(i) exposure of one-half second or less;
or
(ii) during serial radiography
when means shall be provided to permit completion of any single exposure of the
series in process.
(b)
Each x-ray control shall:
(i) for stationary
x-ray systems be permanently mounted in a protected area so that the operator
is required to remain in that protected area during the entire exposure;
or
(ii) mobile and portable x-ray
systems which are:
1) used for greater than
one week in one location (one room or suite) shall meet the requirements of
605.B.2.b.(1) [Item (i), Subparagraph (b), Paragraph (2), Subsection B.,
Section 605 of 20.3.6.605 NMAC] above;
2) used for more than 1 hour and less than 1
week at one location (one room, or suite) shall meet the requirement of
605.B.2.b.(2)(a) [Item (ii), Subparagraph (b), Paragraph (2), Subsection B.,
Section 605 of 20.3.6.605 NMAC] or be provided with 1.98 m (6.5 feet) high
protective barrier which is placed at least 1.83 m (6 feet) from the tube
housing assembly and at least 1.83 m (6 feet) from the patient; or
3) used to make an exposure(s) of only one
patient at the use location shall meet the requirement of 605.B.2.b.(2)(a)
[Item (ii), Subparagraph (b), Paragraph (2), Subsection B., Section 605 of
20.3.6.605 NMAC]or 605.B.2.b.(2)(b) [Item(ii), Subparagraph (b), Paragraph (2),
Subsection B., Section 605 of 20.3.6.605 NMAC] or be provided with a method of
control which will permit the operator to be at least 3.66 m (12 feet) from the
tube head assembly during an exposure.
(iii) the x-ray control shall provide visual
indication observable at or from the operator's protected position whenever
x-rays are produced. In addition, a signal audible to the operator shall
indicate that the exposure has terminated.
(3) Automatic Exposure Controls (Phototimer):
When an automatic exposure control is provided:
(a) indication shall be made on the control
panel when this mode of operation is selected;
(b) when the x-ray tube potential is equal to
or greater than 50 kVp, the minimum exposure time for field emission equipment
rated for pulsed operation shall be equal to or less than a time interval
equivalent to two pulses;
(c) the
minimum exposure time for all equipment other than specified in 605.B.3.b
[Subparagraph (b), Paragraph (3), Subsection B., Section 605 of 20.3.6.605
NMAC] shall be equal to or less than 1/60 second or a time interval required to
deliver 5 mAs, whichever is greater;
(d) Either the product of peak x-ray tube
potential, current, and exposure time shall be limited to not more than 60 kWs
per exposure or the product of x-ray tube current and exposure time shall be
limited to not more than 600 mAs per exposure except when the x-ray tube
potential is less than 50 kVp in which case the product of x-ray tube current
and exposure time shall be limited to not more than 2000 mAs per exposure;
and
(e) A visible signal shall
indicate when an exposure has been terminated at the limits described in
605.B.3.d [Subparagraph (d), Paragraph (3), Subsection B., Section 605 of
20.3.6.605 NMAC], and manual resetting shall be required before further
automatically timed exposures can be made. [5-3-95 ]
(4) Reproducibility: With a timer setting of
0.5 seconds or less, the average exposure period (T) shall be greater than or
equal to 5 times the maximum exposure period (Tmax) minus the minimum exposure
period (Tmin) when 4 timer tests are performed: T 5(Tmax
Tmin).