Current through Register Vol. 35, No. 24, December 23, 2024
All fluoroscopic x-ray systems shall meet the following
requirements.
A. Limitation of Useful
Beam:
(1) The fluoroscopic tube shall not
produce x-rays unless the primary protective barrier is in position to
intercept the entire useful beam at all times.
(2) The entire cross-section of the useful
beam shall be intercepted by the primary protective barrier of the fluoroscopic
image assembly at any SID.
(3)
Limitation to the Imaging Surface.
(a)
Non-Image-Intensified Fluoroscopy and Spot Filming: The x-ray field produced by
non-image-intensified fluoroscopic equipment shall not extend beyond the entire
visible area of the image receptor. This requirement applies to field size
during both fluoroscopic procedures and spot-filming procedures.
(b) Image-Intensified Fluoroscopy and Spot
Filming.
(i) During fluoroscopic or
spot-filming procedures, neither the length nor the width of the x-ray field in
the plane of the image receptor shall exceed the visible area of the image
receptor by more than 3 percent of the SID. The sum of the excess length and
the excess width shall be no greater than 4 percent of the SID.
(ii) Compliance shall be determined with the
beam axis perpendicular to the image receptor. For rectangular x-ray fields
used with circular image reception, the error in alignment shall be determined
along the length and width dimensions of the x-ray field which pass through the
center of the visible area of the image receptor.
B. Activation of the
Fluoroscopic Tube: X-ray production in the fluoroscopic mode shall be
controlled by a device which requires continuous pressure by the fluoroscopist
for the entire time of any exposure. When recording serial fluoroscopic images,
the fluoroscopist shall be able to terminate the x-ray exposure(s) at any time,
but means may be provided to permit completion of any single exposure of the
series in process.
C. Entrance
Exposure Rate Allowable Limits:
(1) The
exposure rate measured at the point where the center of the useful beam enters
the patient shall not exceed 10 R per minute, (2.58 mC/kg) except during
recording of fluoroscopic images or when provided with optional high-level
control.
(2) When provided with
optional high-level control, the equipment shall not be operable at any
combination of tube potential and current which will result in an exposure rate
in excess of 5 Rem (1.29 mC/kg) per minute at the point where the center of the
useful beam enters the patient unless the high-level control is activated. A
continuous signal audible to the fluoroscopist shall indicate that the
high-level control is being employed.
(3) Measuring Compliance of Entrance Exposure
Rate Limits: Compliance with 604.C [Subsection C., Section 604 of
20.3.6.604 NMAC] shall be
determined by:
(a) removing movable grids and
compression devices from the useful beam during the measurements;
(b) if the source is below the table, express
exposure rate, 1 cm above the tabletop or cradle;
(c) express exposure rate, if the source is
above the table, 30 cm above the tabletop with the end of the beam-limiting
device or spacer positioned as closely as possible to the point of measurement;
and
(d) in a C-arm type of
fluoroscope, the exposure rate shall be measured, 30 cm (11.81 inches) from the
input surface of the fluoroscopic imaging assembly.
(4) Periodic measurement of entrance exposure
rate limits:
(a) periodic measurements of the
exposure rate shall be made by a qualified expert. An adequate period for such
measurements shall be annually or after any maintenance of the system which
might affect the exposure rate.
(b)
results of these measurements shall be posted where any fluoroscopist may have
ready access to such results while using that fluoroscope and in the record
required in 602.A.2.e [Subparagraph (e), Paragraph (2), Subsection A., Section
602 of 20.3.6.602 NMAC]. Results of the
measurements shall include the maximum possible Rem/per minute, (1.29 mC/kg) as
well as the physical factors used to determine all data; the name of the person
performing the measurements; and the date the measurements were
performed.
(c) Use of monitoring
devices (e.g., commercially available film badges, thermoluminescent
dosimeters, or low-energy dosimeters) may be used to perform the test, provided
the measurements are made as noted in 604.C.4.d [Subparagraph (d), Paragraph
(4), Subsection C. Section 604 of
20.3.6.604 NMAC].
(d) Conditions of measurement:
(i) the measurement shall be made under the
conditions that satisfy the requirements of 604.C.3 [Paragraph (3), Subsection
C., Section 604 of
20.3.6.604 NMAC];
(ii) the kVp shall be the peak Kv that the
x-ray system is capable of producing;
(iii) the high-level control, if present,
shall not be activated;
(iv) the
x-ray systems that do not incorporate automatic exposure control (automatic
brightness control, etc.) shall have sufficient material (e.g., lead or lead
equivalence) placed in the useful beam to produce the maximum milliamperage of
the x-ray system; and
(v) x-ray
systems that incorporate automatic exposure control shall utilize the maximum
milliamperage of the x-ray system. Materials (e.g. an attenuation block) may be
placed in the useful beam to protect the imaging system.
D. Barrier Transmitted
Radiation Rate Limits:
(1) The exposure rate
due to transmission through the primary protective barrier with the attenuation
block in the useful beam combined with radiation from the image intensifier, if
provided, shall not exceed 2 mR (0.516 mC/kg) per hour at 10 cm (3.93 inches)
from any surface of the fluoroscopic imaging assembly beyond the plane of the
image receptor for each roentgen per minute of entrance exposure
rate.
(2) Measuring compliance of
barrier transmission:
(a) The exposure rate
due to transmission through the primary protective barrier combined with
radiation from the image intensifier shall be determined by measurements
averaged over an area of 100 sq cm with no linear dimension greater than 20 cm
(7.87 inches).
(b) If the source is
below the tabletop, the measurement shall be made with the input surface of the
fluoroscopic imaging assembly, positioned 30 cm (11.81 inches) above the
tabletop.
(c) If the source is
above the tabletop and the SID is variable, the measurement shall be made with
the end of the beam-limiting device or spacer as close to the tabletop as it
can be placed, provided that it shall not be closer than 30 cm (11.81
inches).
(d) Movable grids and
compression devices shall be removed from the useful beam during the
measurement.
(e) The attenuation
block shall be positioned in the useful beam 10 cm (3.93 inches) from the point
of measurement of entrance exposure rate and between this point and the input
surface of the fluoroscopic imaging assembly.
E. Indication of Potential and Current:
During fluoroscopy and cinefluorography, x-ray tube potential and current shall
be continuously indicated.
F.
Source-Skin Distance: The source to skin distance shall not be less than:
(1) 38 cm (14.96 inches) on stationary
fluoroscopes installed after March 10, 1989;
(2) 35.5 cm (13.98 inches) on stationary
fluoroscopes which are in operation prior to March 10, 1989;
(3) 30 cm (11.81 inches) on all mobile
fluoroscopes; and
(4) 20 cm (7.87
inches) for image intensified fluoroscopes used for specific surgical
application. The users operating manual must provide precautionary measures to
be adhered to during the use of this device.
G. Fluoroscopic Timer:
(1) Means shall be provided to preset the
cumulative on-time of the fluoroscopic tube. The maximum cumulative time of the
timing device shall not exceed 5 minutes without resetting.
(2) A signal audible to the fluoroscopist
shall indicate the completion of any preset cumulative on-time. Such signal
shall continue to sound while x-rays are produced until the timing device is
reset.
H. Mobile
Fluoroscopes: In addition to the other requirements of 604 [Section 604 of
20.3.6.604 NMAC], mobile
fluoroscopes shall provide intensified imaging.
I. Control of Scattered Radiation:
(1) Fluoroscopic table designs when combined
with procedures utilized shall be such that no unprotected part of any staff or
ancillary person's body shall be exposed to unattenuated scattered radiation
which originates from under the table. The attenuation required shall be not
less than 0.25 mm lead equivalent.
(2) Equipment configuration when combined
with procedures shall be such that no portion of any staff or ancillary
person's body, except the extremities, shall be exposed to the unattenuated
scattered radiation emanating from above the tabletop unless that individual:
(a) is at least 120 cm (47.24 inches) from
the center of the useful beam; or
(b) the radiation has passed through not less
than 0.25 mm lead equivalent material (e.g., drapes, Bucky-slot cover, sliding
or folding panel, or self supporting curtains) in addition to any lead
equivalency provided by the protective apron referred to in 602 A.1.e.(2)[Item
(ii), Subparagraph (e), Paragraph (1), Subsection A., Section 602 of
20.3.6.602 NMAC]; and
(c) Exceptions to 604.I.2 [Paragrph (2),
Subsection I., Section 604 of
20.3.6.604 NMAC] may be made in
some special procedures where a sterile field will not permit the use of the
normal protective barriers. Where the use of the prefitted sterilized cover for
the barriers is practical, the Department shall not permit such
exception.