New Mexico Administrative Code
Title 20 - ENVIRONMENTAL PROTECTION
Chapter 3 - RADIATION PROTECTION
Part 6 - X-RAYS IN HEALING ARTS
Section 20.3.6.603 - GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC X-RAY SYSTEMS
Current through Register Vol. 35, No. 18, September 24, 2024
In addition to other requirements of this Subpart, all diagnostic x-ray systems shall meet the following requirements:
A. Warning Label: The control panel containing the main power switch shall bear the warning statement, legible and accessible to view: "WARNING: This X-ray Unit May Be Dangerous To Patient and Operator Unless Safe Exposure Factors and Operating Instructions Are Observed";
B. Battery Charge Indicator: On battery-powered generators, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation;
C. Leakage Radiation from the Diagnostic Source Assembly: The leakage radiation from the diagnostic source assembly measured at a distance of 1 m in any direction from the source shall not exceed 100 mR (1 mSv) in 1 hour when the x-ray tube is operated at its leakage technique factor. Compliance shall be determined by measurements averaged over an area of 100 sq cm (39.37 inches) with no linear dimension greater than 20 cm (7.87 inches); and
D. Radiation from Components other than the Diagnostic Source Assembly: The radiation emitted by a component other than the diagnostic source assembly shall not exceed 2 mR (2 mSv) in 1 hour at 5 cm from any accessible surface of the component when it is operated in an assembled x-ray system under any conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of 100 sq cm (39.37 inches) with no linear dimension greater than 20 cm (7.87 inches).
E. Beam Quality:
TABLE I
Design operating range Measured potential Half-value layer
(Kilovolts peak) (Kilovolts peak) aluminum) (Millimeters of)
Below 50 30 0.3
40 0.4
49 0.5
50 to 70 50 1.2
60 1.3
70 1.5
Above 70 71 2.1
80 2.3
90 2.5
100 2.7
110 3.0
120 3.2
130 3.5
140 3.8
150 4.1
TABLE II
FILTRATION REQUIRED vs. OPERATING VOLTAGE
Total Filtration (inherent plus added)
Operating Voltage (millimeters aluminum
(kVp)equivalent)
Below 50 0.5 mm
50 - 70 1.5 mm
Above 70 2.5 mm
F. Multiple Tubes: Where two or more radiographic tubes are controlled by one exposure switch, the tube or tubes which have been selected shall be clearly indicated prior to initiation of the exposure. This indication shall be both on the x-ray control and at or near the tube housing assembly which has been selected.
G. Mechanical Support of Tube Head: The tube housing assembly supports shall be adjusted such that the tube housing assembly will remain stable during an exposure unless the tube housing movement is a designed function of the x-ray system.
H. Technique Indicators: