(a) Each AMU shall be equipped with a litter for the transport of stretcher bound patients. The litter shall be at least 72 inches long (when flat) and at least 20 inches wide. The litter shall have a commercially manufactured stretcher mattress. There shall be clean linens on the litter.

(b) The litter shall have three sets of two-inch wide patient restraints with quick release buckles positioned at the chest, waist and knees. The quick release buckles may be of the "slide-through" or "metal to metal" type. Velcro-type closures shall not be utilized.

(c) The litter shall be restrained by a litter fastener whenever the aircraft is in motion. The litter fastener shall be securely fastened to the aircraft, shall be installed under a FAA Supplemental Type Certificate and shall meet the same "g" requirements as those contained in FAR Part 27.561, "Airworthiness Standards: Normal Category Rotorcraft/ Emergency Landing Conditions: General" or FAR Part 29.561, "Airworthiness Standards: Transport Category Rotorcraft/Emergency Landing Conditions: General" for seats.

Disclaimer: These regulations may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.