New Jersey Administrative Code
Title 7 - ENVIRONMENTAL PROTECTION
Chapter 18 - REGULATIONS GOVERNING THE CERTIFICATION OF LABORATORIES AND ENVIRONMENTAL MEASUREMENTS
Subchapter 7 - TOXICITY TESTING
Section 7:18-7.3 - Laboratory equipment, instruments and materials
Universal Citation: NJ Admin Code 7:18-7.3
Current through Register Vol. 56, No. 24, December 18, 2024
(a) A laboratory performing toxicity tests shall have, on the premises and under the control of the laboratory supervisor, equipment and instruments that satisfy the requirements of (a)1 through 14 below and N.J.A.C. 7:18-3.3.
1. For materials used in the construction of
toxicity testing systems, test organism culturing systems, and sample
collection, handling, and transport systems:
i. The laboratory shall use only materials
listed as "Approved" in Table 7.3 below for the type of test organism in
question.
Table 7.3 |
Materials for constructing toxicity testing systems, |
test organism culturing systems, and |
sample collection, handling and transport systems |
Test Organisms |
Material | Vertebrate | Invertebrate |
Glass, borosilicate, tempered, or soda lime | Approved | Approved |
Stainless steel, # 304 or 316 | Approved | Approved |
Medical grade or food contact silicone, sealant, | Approved | Approved |
tubing, and stoppers | ||
Perfluorocarbon plastics | Approved | Approved |
Polyethylene, white or clear | Approved | Approved |
Polypropylene | Approved | Approved |
Polycarbonate | Approved | Approved |
Polystyrene | Approved | Approved |
Acrylic | Approved | Approved |
Tygon (R), clear or black | Approved | Not Approved |
(except for | ||
Mysids) | ||
Nylon | Approved | Approved |
Fiberglass | Approved | Approved |
Potable water or food contact grade polyvinyl | Approved | Approved |
chloride | ||
Rubber, Neoprene and Gum Latex | Not Appro | Not Approved |
ved | ||
Ceramic (Aluminum Oxide) | Approved | Approved |
ii. The
laboratory shall use glass, stainless steel, ceramic and perfluorocarbon
plastics whenever possible for components that come in contact with wastewater
samples;
iii. If the laboratory
uses silicone, polyethylene, polypropylene, nylon, Tygon(R), polycarbonate and
polystyrene plastics for a component that comes in contact with wastewater
samples, it shall either discard the component after a single use, or
demonstrate that the component can be decontaminated, without significant
degradation, by one or more cleaning procedures listed in
N.J.A.C.
7:18-7.4(c). To demonstrate
that the component can be decontaminated, the laboratory shall:
(1) Clean the component in accordance with
the applicable procedures under
N.J.A.C.
7:18-7.4(c) after using the
component to conduct a compliance toxicity test;
(2) Remove the component, taking an adequate
sample of each type of material being used;
(3) Segregate each type of material into a
separate container, just large enough to completely immerse the materials in
laboratory pure water. The laboratory shall have cleaned the container using
the procedure under
N.J.A.C.
7:18-7.4(c) appropriate to
the test organism used;
(4) Soak
the component in laboratory pure water for 24 hours;
(5) Decant a sufficient volume of water from
each container (or groups of containers of like materials) to analyze for the
organic compounds, metals and trace elements listed in
N.J.A.C.
7:18-7.4(b)1;
(6) Perform an analysis for each type of
material for which the laboratory seeks approval; and
(7) Forward the analytical results to the
Department. The Department shall approve the use of the material only if the
analytical results show that there is no significant degradation of the
material, or cross-over of contamination.
iv. The use of polyvinylchloride, fiberglass,
and acrylics shall only be for holding, acclimating, and rearing system
components and for dilution water storage and delivery system components.
Before use, the laboratory shall test every batch of these materials for
toxicity to the pertinent test organisms. The laboratory shall retain the
documentation of such tests;
v. The
laboratory shall not use Tygon(R) for components used in an invertebrate
testing, holding, acclimating or rearing system except for Mysids. If the
laboratory uses bakelite components in an invertebrate testing, holding,
acclimating or rearing system, and if that bakelite is heated to sterilization
temperatures, the laboratory shall not allow any other system components to
come in contact with either the bakelite or the fumes arising from the
bakelite;
vi. The laboratory shall
not use in toxicity testing any material that is not listed in Table 7.3,
without first obtaining the Department's written approval. To obtain the
Department's approval, the laboratory shall test the material's toxicity to the
pertinent test organisms and submit documentation of the testing to the
Department. The Department shall approve the material only if the documentation
demonstrates that the material does not exhibit toxic or subtoxic effects (that
is, decreased brood size in invertebrate test organisms) to the test organisms;
and
vii. Except for materials
labeled and sold as either, "medical grade" or "food grade," the laboratory
shall clean all new materials before using them. The laboratory shall follow
the following cleaning procedure:
(1) Wash
the materials with a solution consisting of a detergent and hot tap water.
Prepare the solution according to the detergent manufacturer's instructions. Be
sure that the detergent is of a type that leaves no toxic residue;
(2) Rinse the materials well with hot tap
water to remove all traces of detergent;
(3) If the material is all-glass laboratory
ware or perfluorocarbon plastic material, and has a capacity less than or equal
to four liters, then soak glassware in 10 percent hydrochloric acid (HCl) for
at least one hour to remove heavy metal contamination. If the material is
all-glass laboratory ware or perfluorocarbon plastic material, and has a
capacity greater than four liters, then rinse it at least twice with 10 percent
HCl. After soaking or rinsing with acid, rinse twice or more with laboratory
pure water to remove all traces of acid; and
(4) If the material is perfluorocarbon
plastic, rinse it twice with full strength acetone, then rinse it at least
twice with laboratory pure water and air or oven dry it.
2. For flow through toxicity
tests, the laboratory shall use a dilutor system for the accurate measuring,
mixing, and delivery of sample and dilution water to the exposure chambers.
Detailed descriptions of dilutor systems allowable are found in Standard
Methods, 16th edition, and in EPA Acute Methods #027F-1993. The laboratory
shall use a dilutor system that:
i. Provides
an adequate supply of dilution water to maintain 24 hours of continuous
operation. The system shall obtain the supply from a dilution water reservoir,
or by direct continuous pumping from the source of the water;
ii. Is capable of metering the flow of
dilution water and sample into a mixing chamber for the determination of
concentrations. The system shall use a constant head box or metering pumps to
meter the flow of dilution water and sample;
iii. Uses mixing chambers to ensure complete
mixing of dilution water and sample before dispensing solutions into the
exposure chambers;
iv. Uses
separate delivery tubes to transmit the dilution water and sample from the flow
splitters, after the mixing chambers, to each of the replicate exposure
chambers;
v. Provides a flow rate
through the exposure chambers that results in at least five 90 percent water
volume changes every 24 hours, and that is sufficient to maintain dissolved
oxygen in the exposure chambers in accordance with
N.J.A.C.
7:18-7.5(h);
vi. Provides a flow rate through the exposure
chambers that does not vary by more than +/-10 percent among all exposure
chambers or +/-five percent within any given exposure chamber throughout the
duration of the test;
vii.
Maintains the test concentration in each exposure chamber within +/-five
percent of the starting concentration for the duration of the test;
viii. Should be designed to maintain a
constant temperature in the exposure chambers within +/-two degrees Celsius of
the specified test temperature;
ix.
Is designed to curtail automatically the delivery of the sample to the mixing
chambers if the supply of dilution water to the mixing chamber is
interrupted;
x. Is designed to
prevent the test organisms from entering the overflow outlets in the exposure
chambers;
xi. Is capable of
maintaining at least five separate effluent dilutions and a control containing
dilution water with replicate exposure chambers; and
xii. Has had its exposure chamber flow rate,
exposure chamber effluent concentration accuracy, and test solution
temperatures checked and calibrated initially and at least once per day for the
duration of the test, including the last day of the test. The laboratory shall
keep records of these calibrations in accordance with
N.J.A.C.
7:18-7.7(i), and make them
available to the Department during an inspection of the laboratory.
3. The laboratory shall use
holding, acclimating and culturing chambers that:
i. Are constructed of non-toxic materials
that satisfy the requirements of (a)1 above;
ii. Include devices for temperature control,
or are located in a temperature-controlled room;
iii. Are constructed for ease of cleaning and
the prevention of waste material build-up; and
iv. If used for vertebrate species, are
shielded from outside disturbances. The laboratory may shield the chamber
either by isolating it in a low-traffic area, or by shielding it individually.
If the materials used to shield a chamber individually will contact the culture
media, the laboratory shall use materials that satisfy the requirements of (a)1
above.
4. The laboratory
shall use test chambers that:
i. Can
accommodate the testing of fish species in containers with a test solution at
least five centimeters (cm) deep;
ii. If fabricated from non-seamless stainless
steel, have welded seams rather than soldered seams;
iii. If fabricated from lead-free glass, are
made in one piece or made with the use of clear silicone adhesive, of the type
approved by the manufacturer for use in aquaria, to bond the seams. The
laboratory shall expose as little of the silicone adhesive to the test solution
as possible. Extra beads of adhesive shall be placed only on the outside of
containers; and
iv. Are designed to
keep the surface areas as small as possible in relation to their volume, in
order to limit sorption to the vessel walls. Containers to be used with
flow-through tests shall be designed to keep the liquid surface area/volume
ratio as small as possible in order to reduce loss of volatile
substances.
5. A
laboratory shall have and use a balance that:
i. Satisfies the requirements of
N.J.A.C.
7:18-3.3(a)2;
ii. Has a range of at least 0 to 40
grams;
iii. Is readable within 0.1
grams;
iv. Provides reproducibility
of at least +/-0.05 grams.
6. Laboratories performing acute toxicity
testing shall have and use one or more pH meters that satisfy the requirements
of N.J.A.C.
7:18-3.3(a)3.
7. Laboratories performing acute toxicity
testing shall have and use one or more conductivity instruments that satisfy
the requirements of
N.J.A.C.
7:18-3.3(a)6.
8. Laboratories performing acute toxicity
testing shall have and use one or more dissolved oxygen meters that satisfy the
requirements of
N.J.A.C.
7:18-5.2(a)17.
9. Laboratories performing tests with
Cladoceran, shall have the following equipment:
i. A light meter capable of measuring in Lux
or footcandles in the range of at least 0 to 200 footcandles;
ii. Medicine droppers or pipettes with 1.0 to
3.0 mm bores;
iii. Borosilicate
glass beakers with covers, or test chambers made of another approved material
under (a)1 above; and
iv. All
testing equipment to be constructed with materials as approved for
invertebrates in (a)1 above.
10. A laboratory shall have a refrigerator
that is capable of storing the required sample volumes and that satisfies the
requirements of
N.J.A.C.
7:18-3.3(a)7.
11. Laboratories performing zooplankton or
macrocrustacean toxicity tests shall have and use a low-power magnification
device, for working with invertebrate species.
12. A laboratory shall use only glassware,
plasticware and metal utensils that satisfy the requirements of
N.J.A.C.
7:18-3.3(a)8. The laboratory
shall use plasticware only if it is made of inert, nontoxic materials approved
under (a)1 above. When manually establishing test solutions, the laboratory
shall use Class "A" volumetric flasks or graduated cylinders, calibrated "to
deliver."
13. Dilution water sample
containers used by the laboratory for discrete samples shall meet the following
requirements:
i. The laboratory shall use
only wide-mouthed containers equipped either with stoppers, screw caps or an
equivalent closure;
ii. The
laboratory shall use only containers and cap liners constructed of materials
approved under (a)1 above; and
iii.
The laboratory shall clean each container after each use, in accordance with
N.J.A.C.
7:18-7.4(c).
14. A laboratory performing
discrete effluent sampling shall use containers meeting the following
requirements:
i. The laboratory shall use
either wide-mouthed glass containers, disposable unplasticized plastic
containers, or disposable unplasticized plastic liners for containers that are
leakproof and constructed of materials meeting the requirements of
(a)1;
ii. The laboratory shall not
reuse containers made of materials listed in (a)1ii above unless they have been
cleaned in accordance with
N.J.A.C.
7:18-7.4(c);
iii. The laboratory shall discard after one
use any containers made of materials specified in (a)1iii above, and not
cleaned and reused unless the laboratory has demonstrated pursuant to (a)1iii
above that the container can be decontaminated without significant
degradation;
iv. Container closures
shall be leakproof and constructed of materials meeting the requirements of
(a)1 above; and
v. The laboratory
shall store containers in a manner that prevents contamination.
Disclaimer: These regulations may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.