New Jersey Administrative Code
Title 7 - ENVIRONMENTAL PROTECTION
Chapter 18 - REGULATIONS GOVERNING THE CERTIFICATION OF LABORATORIES AND ENVIRONMENTAL MEASUREMENTS
Subchapter 7 - TOXICITY TESTING
Section 7:18-7.3 - Laboratory equipment, instruments and materials
Universal Citation: NJ Admin Code 7:18-7.3
Current through Register Vol. 56, No. 18, September 16, 2024
(a) A laboratory performing toxicity tests shall have, on the premises and under the control of the laboratory supervisor, equipment and instruments that satisfy the requirements of (a)1 through 14 below and 7:18-3.3.
1. For materials used in the construction of
toxicity testing systems, test organism culturing systems, and sample collection,
handling, and transport systems:
i. The laboratory
shall use only materials listed as "Approved" in Table 7.3 below for the type of
test organism in question.
Table 7.3 |
Materials for constructing toxicity testing systems, |
test organism culturing systems, and |
sample collection, handling and transport systems |
Test Organisms |
Material | Vertebrate | Invertebrate |
Glass, borosilicate, tempered, or soda lime | Approved | Approved |
Stainless steel, # 304 or 316 | Approved | Approved |
Medical grade or food contact silicone, sealant, | Approved | Approved |
tubing, and stoppers | ||
Perfluorocarbon plastics | Approved | Approved |
Polyethylene, white or clear | Approved | Approved |
Polypropylene | Approved | Approved |
Polycarbonate | Approved | Approved |
Polystyrene | Approved | Approved |
Acrylic | Approved | Approved |
Tygon (R), clear or black | Approved | Not Approved |
(except for | ||
Mysids) | ||
Nylon | Approved | Approved |
Fiberglass | Approved | Approved |
Potable water or food contact grade polyvinyl | Approved | Approved |
chloride | ||
Rubber, Neoprene and Gum Latex | Not Appro | Not Approved |
ved | ||
Ceramic (Aluminum Oxide) | Approved | Approved |
ii. The laboratory
shall use glass, stainless steel, ceramic and perfluorocarbon plastics whenever
possible for components that come in contact with wastewater samples;
iii. If the laboratory uses silicone,
polyethylene, polypropylene, nylon, Tygon(R), polycarbonate and polystyrene plastics
for a component that comes in contact with wastewater samples, it shall either
discard the component after a single use, or demonstrate that the component can be
decontaminated, without significant degradation, by one or more cleaning procedures
listed in 7:18-7.4(c). To
demonstrate that the component can be decontaminated, the laboratory shall:
(1) Clean the component in accordance with the
applicable procedures under
7:18-7.4(c) after
using the component to conduct a compliance toxicity test;
(2) Remove the component, taking an adequate
sample of each type of material being used;
(3) Segregate each type of material into a
separate container, just large enough to completely immerse the materials in
laboratory pure water. The laboratory shall have cleaned the container using the
procedure under 7:18-7.4(c)
appropriate to the test organism used;
(4) Soak the component in laboratory pure water
for 24 hours;
(5) Decant a sufficient
volume of water from each container (or groups of containers of like materials) to
analyze for the organic compounds, metals and trace elements listed in
7:18-7.4(b)1;
(6) Perform an analysis for each type of material
for which the laboratory seeks approval; and
(7) Forward the analytical results to the
Department. The Department shall approve the use of the material only if the
analytical results show that there is no significant degradation of the material, or
cross-over of contamination.
iv. The use of polyvinylchloride, fiberglass, and
acrylics shall only be for holding, acclimating, and rearing system components and
for dilution water storage and delivery system components. Before use, the
laboratory shall test every batch of these materials for toxicity to the pertinent
test organisms. The laboratory shall retain the documentation of such
tests;
v. The laboratory shall not use
Tygon(R) for components used in an invertebrate testing, holding, acclimating or
rearing system except for Mysids. If the laboratory uses bakelite components in an
invertebrate testing, holding, acclimating or rearing system, and if that bakelite
is heated to sterilization temperatures, the laboratory shall not allow any other
system components to come in contact with either the bakelite or the fumes arising
from the bakelite;
vi. The laboratory
shall not use in toxicity testing any material that is not listed in Table 7.3,
without first obtaining the Department's written approval. To obtain the
Department's approval, the laboratory shall test the material's toxicity to the
pertinent test organisms and submit documentation of the testing to the Department.
The Department shall approve the material only if the documentation demonstrates
that the material does not exhibit toxic or subtoxic effects (that is, decreased
brood size in invertebrate test organisms) to the test organisms; and
vii. Except for materials labeled and sold as
either, "medical grade" or "food grade," the laboratory shall clean all new
materials before using them. The laboratory shall follow the following cleaning
procedure:
(1) Wash the materials with a solution
consisting of a detergent and hot tap water. Prepare the solution according to the
detergent manufacturer's instructions. Be sure that the detergent is of a type that
leaves no toxic residue;
(2) Rinse the
materials well with hot tap water to remove all traces of detergent;
(3) If the material is all-glass laboratory ware
or perfluorocarbon plastic material, and has a capacity less than or equal to four
liters, then soak glassware in 10 percent hydrochloric acid (HCl) for at least one
hour to remove heavy metal contamination. If the material is all-glass laboratory
ware or perfluorocarbon plastic material, and has a capacity greater than four
liters, then rinse it at least twice with 10 percent HCl. After soaking or rinsing
with acid, rinse twice or more with laboratory pure water to remove all traces of
acid; and
(4) If the material is
perfluorocarbon plastic, rinse it twice with full strength acetone, then rinse it at
least twice with laboratory pure water and air or oven dry it.
2. For flow through toxicity tests, the
laboratory shall use a dilutor system for the accurate measuring, mixing, and
delivery of sample and dilution water to the exposure chambers. Detailed
descriptions of dilutor systems allowable are found in Standard Methods, 16th
edition, and in EPA Acute Methods #027F-1993. The laboratory shall use a dilutor
system that:
i. Provides an adequate supply of
dilution water to maintain 24 hours of continuous operation. The system shall obtain
the supply from a dilution water reservoir, or by direct continuous pumping from the
source of the water;
ii. Is capable of
metering the flow of dilution water and sample into a mixing chamber for the
determination of concentrations. The system shall use a constant head box or
metering pumps to meter the flow of dilution water and sample;
iii. Uses mixing chambers to ensure complete
mixing of dilution water and sample before dispensing solutions into the exposure
chambers;
iv. Uses separate delivery
tubes to transmit the dilution water and sample from the flow splitters, after the
mixing chambers, to each of the replicate exposure chambers;
v. Provides a flow rate through the exposure
chambers that results in at least five 90 percent water volume changes every 24
hours, and that is sufficient to maintain dissolved oxygen in the exposure chambers
in accordance with
7:18-7.5(h);
vi. Provides a flow rate through the exposure
chambers that does not vary by more than +/-10 percent among all exposure chambers
or +/-five percent within any given exposure chamber throughout the duration of the
test;
vii. Maintains the test
concentration in each exposure chamber within +/-five percent of the starting
concentration for the duration of the test;
viii. Should be designed to maintain a constant
temperature in the exposure chambers within +/-two degrees Celsius of the specified
test temperature;
ix. Is designed to
curtail automatically the delivery of the sample to the mixing chambers if the
supply of dilution water to the mixing chamber is interrupted;
x. Is designed to prevent the test organisms from
entering the overflow outlets in the exposure chambers;
xi. Is capable of maintaining at least five
separate effluent dilutions and a control containing dilution water with replicate
exposure chambers; and
xii. Has had its
exposure chamber flow rate, exposure chamber effluent concentration accuracy, and
test solution temperatures checked and calibrated initially and at least once per
day for the duration of the test, including the last day of the test. The laboratory
shall keep records of these calibrations in accordance with
7:18-7.7(i), and make
them available to the Department during an inspection of the laboratory.
3. The laboratory shall use holding,
acclimating and culturing chambers that:
i. Are
constructed of non-toxic materials that satisfy the requirements of (a)1
above;
ii. Include devices for
temperature control, or are located in a temperature-controlled room;
iii. Are constructed for ease of cleaning and the
prevention of waste material build-up; and
iv. If used for vertebrate species, are shielded
from outside disturbances. The laboratory may shield the chamber either by isolating
it in a low-traffic area, or by shielding it individually. If the materials used to
shield a chamber individually will contact the culture media, the laboratory shall
use materials that satisfy the requirements of (a)1 above.
4. The laboratory shall use test chambers that:
i. Can accommodate the testing of fish species in
containers with a test solution at least five centimeters (cm) deep;
ii. If fabricated from non-seamless stainless
steel, have welded seams rather than soldered seams;
iii. If fabricated from lead-free glass, are made
in one piece or made with the use of clear silicone adhesive, of the type approved
by the manufacturer for use in aquaria, to bond the seams. The laboratory shall
expose as little of the silicone adhesive to the test solution as possible. Extra
beads of adhesive shall be placed only on the outside of containers; and
iv. Are designed to keep the surface areas as
small as possible in relation to their volume, in order to limit sorption to the
vessel walls. Containers to be used with flow-through tests shall be designed to
keep the liquid surface area/volume ratio as small as possible in order to reduce
loss of volatile substances.
5. A laboratory shall have and use a balance that:
i. Satisfies the requirements of
7:18-3.3(a)2;
ii. Has a range of at least 0 to 40
grams;
iii. Is readable within 0.1
grams;
iv. Provides reproducibility of
at least +/-0.05 grams.
6.
Laboratories performing acute toxicity testing shall have and use one or more pH
meters that satisfy the requirements of
7:18-3.3(a)3.
7. Laboratories performing acute toxicity testing
shall have and use one or more conductivity instruments that satisfy the
requirements of 7:18-3.3(a)6.
8. Laboratories performing acute toxicity testing
shall have and use one or more dissolved oxygen meters that satisfy the requirements
of 7:18-5.2(a)17.
9. Laboratories performing tests with Cladoceran,
shall have the following equipment:
i. A light
meter capable of measuring in Lux or footcandles in the range of at least 0 to 200
footcandles;
ii. Medicine droppers or
pipettes with 1.0 to 3.0 mm bores;
iii.
Borosilicate glass beakers with covers, or test chambers made of another approved
material under (a)1 above; and
iv. All
testing equipment to be constructed with materials as approved for invertebrates in
(a)1 above.
10. A laboratory
shall have a refrigerator that is capable of storing the required sample volumes and
that satisfies the requirements of
7:18-3.3(a)7.
11. Laboratories performing zooplankton or
macrocrustacean toxicity tests shall have and use a low-power magnification device,
for working with invertebrate species.
12. A laboratory shall use only glassware,
plasticware and metal utensils that satisfy the requirements of
7:18-3.3(a)8. The
laboratory shall use plasticware only if it is made of inert, nontoxic materials
approved under (a)1 above. When manually establishing test solutions, the laboratory
shall use Class "A" volumetric flasks or graduated cylinders, calibrated "to
deliver."
13. Dilution water sample
containers used by the laboratory for discrete samples shall meet the following
requirements:
i. The laboratory shall use only
wide-mouthed containers equipped either with stoppers, screw caps or an equivalent
closure;
ii. The laboratory shall use
only containers and cap liners constructed of materials approved under (a)1 above;
and
iii. The laboratory shall clean each
container after each use, in accordance with
7:18-7.4(c).
14. A laboratory performing discrete
effluent sampling shall use containers meeting the following requirements:
i. The laboratory shall use either wide-mouthed
glass containers, disposable unplasticized plastic containers, or disposable
unplasticized plastic liners for containers that are leakproof and constructed of
materials meeting the requirements of (a)1;
ii. The laboratory shall not reuse containers made
of materials listed in (a)1ii above unless they have been cleaned in accordance with
7:18-7.4(c);
iii. The laboratory shall discard after one use
any containers made of materials specified in (a)1iii above, and not cleaned and
reused unless the laboratory has demonstrated pursuant to (a)1iii above that the
container can be decontaminated without significant degradation;
iv. Container closures shall be leakproof and
constructed of materials meeting the requirements of (a)1 above; and
v. The laboratory shall store containers in a
manner that prevents contamination.
Disclaimer: These regulations may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.