New Jersey Administrative Code
Title 7 - ENVIRONMENTAL PROTECTION
Chapter 18 - REGULATIONS GOVERNING THE CERTIFICATION OF LABORATORIES AND ENVIRONMENTAL MEASUREMENTS
Subchapter 5 - CHEMICAL TESTING
Section 7:18-5.2 - Requirements for environmental laboratory equipment and instruments
Universal Citation: NJ Admin Code 7:18-5.2
Current through Register Vol. 56, No. 18, September 16, 2024
(a) The supervisor shall have control over the equipment and instruments used in chemical testing. The laboratory shall use only equipment and instruments that meets the applicable requirements listed in (a)1 through 17 below, the applicable requirements of N.J.A.C. 7:18-3, and the requirements of the applicable DSAM.
1. Spectrophotometers (other than atomic
absorption spectrophotometers) shall meet the following requirements:
i. The spectral range shall be at least 400 to 700
nanometers (nm). The maximum spectral bandwidth shall be no more than 20
nm;
ii. Wavelength accuracy shall be
within +/-2.5 nm; and
iii.Spectrophotometers shall employ a cell path
length permitting a linear calibration of the instrument in the anticipated
concentration range consistent with the DSAM.
2. Filter photometers or colorimeters shall meet
the following requirements:
i. Filter photometers
or colorimeters shall have filters that isolate various radiant energy bands in the
400 to 700 nm range. The filters shall have a bandwidth between 10 and 70 nm;
and
ii. Filter photometers and
colorimeters shall employ a cell path length permitting a linear calibration of the
instrument in the anticipated concentration range consistent with the
DSAM.
3. Atomic absorption
spectrophotometers shall meet the following requirements:
i. Atomic absorption spectrophotometers shall be
single or multiple channel, single or double beam instruments having a grating
monochromator, photomultiplier detector, adjustable slits, and provisions for
interfacing with an analog/digital chart recorder/printer or a computer data
system;
ii. If used, a computer data
system shall perform analog-digital conversions with integration, storage, and
output. The laboratory shall produce completed header information for the computer
system to define the unique sample, blank or standard run; date/time of analysis;
analyst; parameter(s) concentrations, and or absorbance values;
iii. The instruments shall be operated with the
fuel and oxidant gases specified by the analytical method;
iv. Instruments used to analyze metals as hydrides
shall:
(1) Have a hydride generator that meets the
specifications of the applicable DSAM; and
(2) Be able to meet the temperature and background
correction requirements of the applicable DSAM.
4. For mercury analysis, a mercury analyzer or an
atomic absorption spectrophotometer used for mercury analysis shall meet the
following requirements:
i. The laboratory shall
operate the instruments used for cold-vapor mercury analysis using the lamps
specified by the applicable DSAM;
ii.
The laboratory shall use absorption cells that measure at least 10 centimeters (cm)
and have 2.5 cm quartz end windows, or their equivalent;
iii. The laboratory shall use a vapor flow system
including an air pump delivering one liter per minute, a heated drying unit or a
tube containing 20 grams of magnesium perchlorate, and an aeration tube with coarse
glass-frit; and
iv. Because of the toxic
nature of mercury vapor, the laboratory shall take precautions to avoid subjecting
individuals to inhalation of the vapor. Therefore, when the samples are analyzed,
the released mercury vapor shall be passed through an absorbing media, such as equal
volumes of 0.1 N potassium permanganate (KMnO[4]) and 10 percent sulfuric acid
(H[2]SO[4]), or 0.025 percent iodine in a three percent potassium iodide (KI)
solution, or specially treated charcoal that will absorb mercury vapor.
5. Inductively coupled plasma (ICP)
spectrometers shall meet the following requirements:
i. The laboratory's ICP instruments shall be
computer-controlled;
ii. The system
shall be capable of background correction; and
iii. The system shall include a computer data
system that performs analog-digital conversions with integration, storage, and
output. The laboratory shall produce completed header information for the computer
system to define the unique sample, blank or standard run; the date and time of
instrumental analysis; the analyst; and the parameter or parameters for which the
sample is being analyzed.
6.
Inductively coupled plasma/mass spectrometers (ICP/MS) shall meet the requirements
applicable to ICP spectrometers under (a)5 above. The laboratory shall operate
ICP/MS instrumentation using the mass spectrometer ionization conditions, scan
range, and scan rate defined by the applicable DSAM, and shall meet the tuning
criteria, initial and continuing calibration, quality assurance, and quality control
requirements of the applicable DSAM.
7.
Transmission electron microscopes and associated energy dispersive X-ray analyzers
shall meet the requirements of the applicable DSAM.
8. Gas chromatographs shall meet the following
requirements:
i. GC Column ovens shall be capable
of isothermal temperature control;
ii.
Injection systems, columns, and carrier gas flow control conditions shall meet the
requirements of the applicable DSAM;
iii. Detectors shall meet the requirements of the
applicable DSAM;
iv. Chromatograms shall
be recorded with a strip chart recorder and integrator or combined
recorder/integrator or computer data system; and
v. The original hard copy of all chromatograms
shall meet the requirements of (a)14 below.
9. Gas chromatograph/mass spectrometers (GC/MS)
shall meet the requirements for gas chromatographs in (a)8 above, and the
requirements for mass spectrometers under (a)13 below.
10. High performance liquid chromatographs (HPLC)
shall meet the following requirements:
i.
Isocratic and/or linear gradient elution chromatography shall be used;
ii. Fluorescence, UV or electrochemical detectors
shall be used, as required by the applicable DSAM;
iii. Reverse-phase or other columns shall be used
as prescribed by the applicable DSAM;
iv. Chromatograms shall be recorded with a strip
chart recorder and integrator or combined recorder/integrator or computer data
system; and
v. The original hard copy of
all chromatograms shall meet the requirements of (a)14 below.
11. High performance liquid chromatograph/mass
spectrometers (HPLC/MS) shall satisfy the requirements for high performance liquid
chromatographs in (a)10i, iii and iv above, and the requirements for mass
spectrometers under (a)13 below.
12. Ion
chromatographs shall meet the requirements defined by the DSAM including the
following requirements:
i. Suppressor and
separator or other columns shall be used as required by the applicable
DSAM;
ii. Conductivity or other detector
shall be used as required by the applicable DSAM;
iii. Chromatograms shall be recorded with a strip
chart recorder and integrator or combined recorder/integrator or computer data
system; and
iv. The original hard copy
of all chromatograms shall meet the requirements of (a)14 below.
13. Mass spectrometers under (a)9 and
11 above shall meet the following requirements:
i.
Mass spectrometer instrumentation shall be operated using the ionization conditions,
scan range, and scan rate, and shall meet the tuning criteria, initial and
continuing calibration, quality assurance and quality control requirements of the
applicable DSAM;
ii. The mass
spectrometer shall have a computer data system for performing qualitative
identifications and quantitative calculations for target compounds. It shall be
capable of identifying and semi-quantitating "non-target" or tentatively identified
compounds (TICs). The software shall use retention time and mass spectral
comparisons for qualitative identifications. The software shall use a formula
defined in the applicable DSAM to calculate quantitative results of target
compounds;
iii. The computer data system
shall be capable of performing a mass spectra search against the NIST library or
other USEPA-approved mass spectral library. The data system shall rank and present
the best three qualitative identification mass spectral matches. If a parameter
cannot be specifically identified, but its compound class can be determined by mass
spectral matching, its compound class shall be reported. If the compound class is
indeterminate, the parameter shall be reported as an unknown. Semi-quantitative
results for a non-target TIC shall be estimated by assuming that its concentration
is proportional to that of the nearest internal standard;
iv. The laboratory's GC/MS analyst or supervisor
shall independently confirm all software qualitative identifications for found
parameters; and
v. The original hard
copy of all chromatograms shall meet the requirements of (a)14 below.
14. The laboratory shall have the
analyst sign the original hard copy of all chromatograms, analog or digital,
prepared using any of the types of equipment listed in (a)8 through 12 above. In the
original hard copy, the laboratory shall include a table setting forth all of the
following information:
i. Identification of the
sample, blank or standard;
ii. The date
and time of the analysis;
iii. The run
number; and
iv. Peak identification, by
number, by retention time, or by name. The peak identification shall include
internal standards, surrogates, and sample components.
15. Auto-analyzer equipment shall meet the
requirements defined by the automated methods of the DSAMs including the following
requirements:
i. The spectral range shall be at
least a minimum of 400 to 700 nm. The maximum spectral bandwidth shall be no more
than 20 nm;
ii. Wavelength accuracy
shall be within +/-2.5 nm; and
iii. The
laboratory shall have the analyst sign the original hard copy of all outputs. In all
outputs, analog or digital, the laboratory shall include a table setting forth the
following information: identification of the sample, blank or standard; the date and
time of analysis; the run number; and peak identification.
16. Any burets used for titration shall be Class
"A" burets, and need not be calibrated before use.
17. Dissolved oxygen (DO) meters with membrane
electrodes shall meet the following requirements:
i. Dissolved oxygen measurements shall be accurate
to within +/-0.3 mg dissolved oxygen per liter (DO/L) and shall be precise to within
+/-0.15 mg DO/L; and
ii. Meters shall be
capable of compensation for temperature.
18. At least annually, the laboratory shall check
salinity meters equipped with conductivity cells having platinum electrodes. The
check shall cover the range of interest using at least five concentrations of a
standard potassium chloride solution. Conductivity cells not having platinum
electrodes shall be checked against a conductivity meter equipped with platinum
electrodes. The laboratory shall perform this check annually. The laboratory shall
record the raw data, cell constant, and results in a log book, with each entry
signed and dated by the analyst.
19. The
laboratory shall have documented procedures for the calibration and verification of
air sampling equipment such as pumps, meter boxes, critical orifices, flow
measurement devices and continuous analyzers, if this equipment is used or supplied
by the laboratory.
20. All air sampling
canisters shall be internally passivated by the SUMMA electropolish process, as set
forth in the methodologies referenced at
7:18-1.5(a)5, or other
EPA approved process.
Disclaimer: These regulations may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.