New Jersey Administrative Code
Title 7 - ENVIRONMENTAL PROTECTION
Chapter 18 - REGULATIONS GOVERNING THE CERTIFICATION OF LABORATORIES AND ENVIRONMENTAL MEASUREMENTS
Subchapter 5 - CHEMICAL TESTING
Section 7:18-5.2 - Requirements for environmental laboratory equipment and instruments
Universal Citation: NJ Admin Code 7:18-5.2
Current through Register Vol. 56, No. 24, December 18, 2024
(a) The supervisor shall have control over the equipment and instruments used in chemical testing. The laboratory shall use only equipment and instruments that meets the applicable requirements listed in (a)1 through 17 below, the applicable requirements of N.J.A.C. 7:18-3, and the requirements of the applicable DSAM.
1. Spectrophotometers (other
than atomic absorption spectrophotometers) shall meet the following
requirements:
i. The spectral range shall be
at least 400 to 700 nanometers (nm). The maximum spectral bandwidth shall be no
more than 20 nm;
ii. Wavelength
accuracy shall be within +/-2.5 nm; and
iii.Spectrophotometers shall employ a cell
path length permitting a linear calibration of the instrument in the
anticipated concentration range consistent with the DSAM.
2. Filter photometers or colorimeters shall
meet the following requirements:
i. Filter
photometers or colorimeters shall have filters that isolate various radiant
energy bands in the 400 to 700 nm range. The filters shall have a bandwidth
between 10 and 70 nm; and
ii.
Filter photometers and colorimeters shall employ a cell path length permitting
a linear calibration of the instrument in the anticipated concentration range
consistent with the DSAM.
3. Atomic absorption spectrophotometers shall
meet the following requirements:
i. Atomic
absorption spectrophotometers shall be single or multiple channel, single or
double beam instruments having a grating monochromator, photomultiplier
detector, adjustable slits, and provisions for interfacing with an
analog/digital chart recorder/printer or a computer data system;
ii. If used, a computer data system shall
perform analog-digital conversions with integration, storage, and output. The
laboratory shall produce completed header information for the computer system
to define the unique sample, blank or standard run; date/time of analysis;
analyst; parameter(s) concentrations, and or absorbance values;
iii. The instruments shall be operated with
the fuel and oxidant gases specified by the analytical method;
iv. Instruments used to analyze metals as
hydrides shall:
(1) Have a hydride generator
that meets the specifications of the applicable DSAM; and
(2) Be able to meet the temperature and
background correction requirements of the applicable DSAM.
4. For mercury analysis, a mercury
analyzer or an atomic absorption spectrophotometer used for mercury analysis
shall meet the following requirements:
i. The
laboratory shall operate the instruments used for cold-vapor mercury analysis
using the lamps specified by the applicable DSAM;
ii. The laboratory shall use absorption cells
that measure at least 10 centimeters (cm) and have 2.5 cm quartz end windows,
or their equivalent;
iii. The
laboratory shall use a vapor flow system including an air pump delivering one
liter per minute, a heated drying unit or a tube containing 20 grams of
magnesium perchlorate, and an aeration tube with coarse glass-frit;
and
iv. Because of the toxic nature
of mercury vapor, the laboratory shall take precautions to avoid subjecting
individuals to inhalation of the vapor. Therefore, when the samples are
analyzed, the released mercury vapor shall be passed through an absorbing
media, such as equal volumes of 0.1 N potassium permanganate (KMnO[4]) and 10
percent sulfuric acid (H[2]SO[4]), or 0.025 percent iodine in a three percent
potassium iodide (KI) solution, or specially treated charcoal that will absorb
mercury vapor.
5.
Inductively coupled plasma (ICP) spectrometers shall meet the following
requirements:
i. The laboratory's ICP
instruments shall be computer-controlled;
ii. The system shall be capable of background
correction; and
iii. The system
shall include a computer data system that performs analog-digital conversions
with integration, storage, and output. The laboratory shall produce completed
header information for the computer system to define the unique sample, blank
or standard run; the date and time of instrumental analysis; the analyst; and
the parameter or parameters for which the sample is being analyzed.
6. Inductively coupled plasma/mass
spectrometers (ICP/MS) shall meet the requirements applicable to ICP
spectrometers under (a)5 above. The laboratory shall operate ICP/MS
instrumentation using the mass spectrometer ionization conditions, scan range,
and scan rate defined by the applicable DSAM, and shall meet the tuning
criteria, initial and continuing calibration, quality assurance, and quality
control requirements of the applicable DSAM.
7. Transmission electron microscopes and
associated energy dispersive X-ray analyzers shall meet the requirements of the
applicable DSAM.
8. Gas
chromatographs shall meet the following requirements:
i. GC Column ovens shall be capable of
isothermal temperature control;
ii.
Injection systems, columns, and carrier gas flow control conditions shall meet
the requirements of the applicable DSAM;
iii. Detectors shall meet the requirements of
the applicable DSAM;
iv.
Chromatograms shall be recorded with a strip chart recorder and integrator or
combined recorder/integrator or computer data system; and
v. The original hard copy of all
chromatograms shall meet the requirements of (a)14 below.
9. Gas chromatograph/mass spectrometers
(GC/MS) shall meet the requirements for gas chromatographs in (a)8 above, and
the requirements for mass spectrometers under (a)13 below.
10. High performance liquid chromatographs
(HPLC) shall meet the following requirements:
i. Isocratic and/or linear gradient elution
chromatography shall be used;
ii.
Fluorescence, UV or electrochemical detectors shall be used, as required by the
applicable DSAM;
iii. Reverse-phase
or other columns shall be used as prescribed by the applicable DSAM;
iv. Chromatograms shall be recorded with a
strip chart recorder and integrator or combined recorder/integrator or computer
data system; and
v. The original
hard copy of all chromatograms shall meet the requirements of (a)14
below.
11. High
performance liquid chromatograph/mass spectrometers (HPLC/MS) shall satisfy the
requirements for high performance liquid chromatographs in (a)10i, iii and iv
above, and the requirements for mass spectrometers under (a)13 below.
12. Ion chromatographs shall meet the
requirements defined by the DSAM including the following requirements:
i. Suppressor and separator or other columns
shall be used as required by the applicable DSAM;
ii. Conductivity or other detector shall be
used as required by the applicable DSAM;
iii. Chromatograms shall be recorded with a
strip chart recorder and integrator or combined recorder/integrator or computer
data system; and
iv. The original
hard copy of all chromatograms shall meet the requirements of (a)14
below.
13. Mass
spectrometers under (a)9 and 11 above shall meet the following requirements:
i. Mass spectrometer instrumentation shall be
operated using the ionization conditions, scan range, and scan rate, and shall
meet the tuning criteria, initial and continuing calibration, quality assurance
and quality control requirements of the applicable DSAM;
ii. The mass spectrometer shall have a
computer data system for performing qualitative identifications and
quantitative calculations for target compounds. It shall be capable of
identifying and semi-quantitating "non-target" or tentatively identified
compounds (TICs). The software shall use retention time and mass spectral
comparisons for qualitative identifications. The software shall use a formula
defined in the applicable DSAM to calculate quantitative results of target
compounds;
iii. The computer data
system shall be capable of performing a mass spectra search against the NIST
library or other USEPA-approved mass spectral library. The data system shall
rank and present the best three qualitative identification mass spectral
matches. If a parameter cannot be specifically identified, but its compound
class can be determined by mass spectral matching, its compound class shall be
reported. If the compound class is indeterminate, the parameter shall be
reported as an unknown. Semi-quantitative results for a non-target TIC shall be
estimated by assuming that its concentration is proportional to that of the
nearest internal standard;
iv. The
laboratory's GC/MS analyst or supervisor shall independently confirm all
software qualitative identifications for found parameters; and
v. The original hard copy of all
chromatograms shall meet the requirements of (a)14 below.
14. The laboratory shall have the analyst
sign the original hard copy of all chromatograms, analog or digital, prepared
using any of the types of equipment listed in (a)8 through 12 above. In the
original hard copy, the laboratory shall include a table setting forth all of
the following information:
i. Identification
of the sample, blank or standard;
ii. The date and time of the
analysis;
iii. The run number;
and
iv. Peak identification, by
number, by retention time, or by name. The peak identification shall include
internal standards, surrogates, and sample components.
15. Auto-analyzer equipment shall meet the
requirements defined by the automated methods of the DSAMs including the
following requirements:
i. The spectral range
shall be at least a minimum of 400 to 700 nm. The maximum spectral bandwidth
shall be no more than 20 nm;
ii.
Wavelength accuracy shall be within +/-2.5 nm; and
iii. The laboratory shall have the analyst
sign the original hard copy of all outputs. In all outputs, analog or digital,
the laboratory shall include a table setting forth the following information:
identification of the sample, blank or standard; the date and time of analysis;
the run number; and peak identification.
16. Any burets used for titration shall be
Class "A" burets, and need not be calibrated before use.
17. Dissolved oxygen (DO) meters with
membrane electrodes shall meet the following requirements:
i. Dissolved oxygen measurements shall be
accurate to within +/-0.3 mg dissolved oxygen per liter (DO/L) and shall be
precise to within +/-0.15 mg DO/L; and
ii. Meters shall be capable of compensation
for temperature.
18. At
least annually, the laboratory shall check salinity meters equipped with
conductivity cells having platinum electrodes. The check shall cover the range
of interest using at least five concentrations of a standard potassium chloride
solution. Conductivity cells not having platinum electrodes shall be checked
against a conductivity meter equipped with platinum electrodes. The laboratory
shall perform this check annually. The laboratory shall record the raw data,
cell constant, and results in a log book, with each entry signed and dated by
the analyst.
19. The laboratory
shall have documented procedures for the calibration and verification of air
sampling equipment such as pumps, meter boxes, critical orifices, flow
measurement devices and continuous analyzers, if this equipment is used or
supplied by the laboratory.
20. All
air sampling canisters shall be internally passivated by the SUMMA
electropolish process, as set forth in the methodologies referenced at
N.J.A.C.
7:18-1.5(a)5, or other EPA
approved process.
Disclaimer: These regulations may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.