New Jersey Administrative Code
Title 7 - ENVIRONMENTAL PROTECTION
Chapter 18 - REGULATIONS GOVERNING THE CERTIFICATION OF LABORATORIES AND ENVIRONMENTAL MEASUREMENTS
Subchapter 4 - MICROBIOLOGICAL TESTING
Section 7:18-4.4 - Requirements for general environmental laboratory practices
Universal Citation: NJ Admin Code 7:18-4.4
Current through Register Vol. 56, No. 24, December 18, 2024
(a) A laboratory performing microbiological analysis shall practice and meet the requirements listed in (a)1 through 4 below.
1. The laboratory shall follow sterilization
procedures meeting the following requirements:
i. The times for autoclaving materials at 121
degrees Celsius are listed below. Except for membrane filters and pads and
carbohydrate-containing media, indicated times are minimal times which may
necessitate adjustment depending upon volumes, containers, and loads;
Material | Time (Minutes) |
Membrane filters and pads | 10 |
Carbohydrate-containing media (lauryl tryptose, | 12-15 |
brilliant green lactose bile broth, etc.) | |
Contaminated materials and discarded tests | 30 |
Membrane filter assemblies (wrapped), sample | 15 |
collection bottles (empty), individual glassware | |
items | |
Rinse water | 15 |
Dilution water blanks | 15 |
ii. Membrane
filter assemblies shall be sterilized at the start of the first filtration
series, either by autoclaving in accordance with (a)1i above, or by two minutes
of exposure in an ultraviolet sterilizer unit. The laboratory shall not use the
ultraviolet sterilizer unit if its use affects the validity of the results. The
laboratory shall test the ultraviolet lamps quarterly with a light meter and a
spread plate irradiation test. The laboratory shall not reuse a filtration unit
without sterilizing it if 30 minutes or more has elapsed since the last sample
was filtered; and
iii. If glassware
is sterilized in a hot air oven, the temperature shall be held at 170 degrees
Celsius for a minimum of two hours.
2. The laboratory shall use only laboratory
pure water that:
i. Has been tested by a
certified environmental laboratory certified to perform the required chemical
analysis for testing of the laboratory pure water; and
ii. Meets the requirements in Table 4.1 at
N.J.A.C.
7:18-4.5(c)4.
3. The laboratory shall use only
rinse water and dilution water that meets the following requirements:
i. Stock buffer solution shall be prepared in
accordance with the DSAM or EPA Microbiological Methods, using laboratory pure
water;
ii. Stock buffer shall be
either autoclaved or filter-sterilized, and must be labeled, dated, and stored
at one to five degrees Celsius;
iii. The stored buffer solution shall be free
of turbidity; and
iv. Rinse and
dilution water shall be prepared by adding 1.25 milliliters of stock buffer
solution and five milliliters of magnesium chloride solution (81.1 grams of
magnesium chloride hexahydrate) per liter of laboratory pure water, and the
final pH shall be 7.2 +/- 0.1.
4. The laboratory shall use only media that
is prepared and stored in accordance with the following requirements:
i. All media shall be prepared according to
the procedures for media preparation set out in the DSAM. However, lactose
broth shall not be used;
ii.
Dehydrated media containers shall be kept tightly closed and stored in a cool,
dry location, to prevent discoloration and caking. Laboratories shall not use
discolored or caked dehydrated media;
iii. Dissolution of the media using
laboratory pure water shall be completed before dispensing to culture tubes or
bottles;
iv. The membrane filter
broth and agar media shall be heated in a boiling water bath or they may be
heated on a hot plate if constantly agitated until completely
dissolved;
v. MF broths shall be
stored and refrigerated no longer than 96 hours and poured MF agar plates shall
be stored, refrigerated and used within two weeks;
vi. MPN media prepared in tubes with
loose-fitting caps shall be used within one week, but if MPN media are
refrigerated after sterilization, they shall be incubated overnight at 35
degrees Celsius to confirm usability, and tubes showing growth or gas bubbles
shall be discarded;
vii. Media in
screw cap containers may be held up to three months, provided that the media
are stored in an enclosed area so that no light may enter and evaporation does
not exceed 10 percent of the original volume; in addition, commercially
prepared liquid and agar media supplies may be used; and
viii. Ampule media shall be stored at one to
five degrees Celsius (34 to 41 degrees Fahrenheit), and storage time shall be
limited to the manufacturer's expiration date.
Disclaimer: These regulations may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.