New Hampshire Code of Administrative Rules
Saf - Department of Safety
Subtitle Saf-C - Commissioner, Department of Safety
Chapter Saf-C 6300 - BREATH ALCOHOL PROCEDURE USING THE INTOXILYZER 5000 INSTRUMENT
Part Saf-C 6306 - INDEPENDENT LABORATORY ACCREDITATION
Section Saf-C 6306.04 - Captured Sample Analysis Report
Universal Citation: NH Admin Rules Saf-C 6306.04
Current through Register No. 40, October 3, 2024
(a) The report of the analysis of a captured sample performed by an independent laboratory pursuant to RSA 265-A:7 shall include:
(1) The serial number of the captured sample
bag;
(2) The serial number of the
approved instrument from which the breath sample was captured;
(3) The subject's reported value identified
on the printout from the approved instrument in accordance with
Saf-C
6302.04;
(4) The results of the captured sample
analyzed in duplicate in accordance with Saf-C 6305.02(a)(5) and (6);
(5) The independent laboratory's reported
value as determined in accordance with Saf-C 6305.03(a); and
(6) The name of the person conducting the
analysis.
Disclaimer: These regulations may not be the most recent version. New Hampshire may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.