New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4047 - USE OF RADIATION MACHINES: THERAPEUTIC RADIATION MACHINES
Section He-P 4047.08 - Calibration of Survey Instruments

Universal Citation: NH Admin Rules He-P 4047.08

Current through Register No. 40, October 3, 2024

(a) The registrant shall ensure that the survey instruments used to show compliance with He-P 4047 have been calibrated as follows:

(1) Before the first use;

(2) At intervals not to exceed 12 months; and

(3) Following repair.

(b) To satisfy the requirements of He-P 4047.08(a) , the registrant shall:

(1) Calibrate all required scale readings up to 10 mSv (1000 mrem) per hour with an appropriate radiation source that is traceable to the National Institute of Standards and Technology (NIST); and

(2) Calibrate at least 2 points on each scale to be calibrated. These points shall be approximately 1/3 to 2/3 full-scale.

(c) To satisfy the requirements of He-P 4047.08(b) , the registrant shall:

(1) Consider a point as calibrated if the indicated dose rate differs from the calculated dose rate by not more than 10 percent; and

(2) Consider a point as calibrated if the indicated dose rate differs from the calculated dose rate by not more than 20 percent if a correction factor or graph is conspicuously attached to the instrument.

(d) The registrant shall retain a record of each calibration required in He-P 4047.08(a) for 3 years.

(e) The record required in He-P 4047.08(d) shall include:

(1) A description of the calibration procedure; and

(2) A description of:
a. The source used;

b. The certified dose rates from the source;

c. The rates indicated by the instrument being calibrated;

d. The correction factors deduced from the calibration data;

e. The signature of the individual who performed the calibration; and

f. The date of calibration.

(f) The registrant may obtain the services of individuals registered or licensed by DHHS/RHS, the U.S. Nuclear Regulatory Commission, an Agreement State, or a Licensing State to perform calibrations of survey instruments.

#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07

Disclaimer: These regulations may not be the most recent version. New Hampshire may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.