Current through Register No. 40, October 3, 2024
(a) Each
facility location authorized to use a therapeutic radiation machine in
accordance with He-P 4047.07 shall possess,
operable and calibrated in accordance with
He-P
4047.08, portable monitoring equipment to include as a
minimum a portable radiation measurement survey instrument capable of measuring
dose rates over the range 10 µSv (1 mrem) per hour to 10 mSv (1000 mrem)
per hour.
(b) Leakage radiation
outside the maximum useful beam in photon and electron modes shall meet the
following requirements:
(1) The absorbed dose
due to leakage radiation (excluding neutrons) at any point outside the maximum
sized useful beam, but within a circular plane of radius 2 meters which is
perpendicular to and centered on the central axis of the useful beam at the
nominal treatment distance or patient plane, shall not exceed a maximum of 0.2
percent and an average of 0.1 percent of the absorbed dose on the central axis
of the beam at the nominal treatment distance when measurements are averaged
over an area not exceeding 100 square centimeters at a minimum of 16 points
uniformly distributed in the plane;
(2) Except for the area defined in
He-P
4047.07(b) (1) , the absorbed dose
due to leakage radiation (excluding neutrons) at 1 meter from the electron path
between the electron source and the target or electron window shall not exceed
0.5 percent of the absorbed dose on the central axis of the beam at the nominal
treatment distance when measurements are averaged over an area not exceeding
100 square centimeters;
(3) The
neutron absorbed dose outside the useful beam shall be in compliance with
International Electrotechnical Commission (IEC) Document 60601-2-1, as amended,
which is incorporated by reference and included in Appendix A;
(4) For each therapeutic radiation machine,
the registrant shall determine, or obtain from the manufacturer, the leakage
radiation existing at the positions specified in
He-P
4047.07(b) (1) -(3) for the
specified operating conditions; and
(5) Records on leakage radiation measurements
shall be maintained after installation for inspection by DHHS/RHS.
(c) Leakage radiation through beam
limiting devices shall be as follows:
(1) For
photon radiation, all adjustable or interchangeable beam limiting devices shall
attenuate the useful beam such that at the nominal treatment distance, the
maximum absorbed dose anywhere in the area shielded by the beam limiting
device(s) shall not exceed 5 percent of the maximum absorbed dose on the
central axis of the useful beam measured in a 100 square centimeter radiation
field or maximum available field size if less than 100 square
centimeters;
(2) For electron
radiation, all adjustable or interchangeable electron applicators shall
attenuate the radiation, including but not limited to photon radiation
generated by electrons incident on the beam limiting device and electron
applicator and other parts of the radiation head, such that the absorbed dose
in a plane perpendicular to the central axis of the useful beam at the nominal
treatment distance shall not exceed:
a. For
points beyond a line 7 centimeters outside the periphery of the useful beam, a
maximum limit of 2 percent and average of 0.5 percent of the absorbed dose on
the central axis of the useful beam at the nominal treatment distance;
and
b. For points beyond a line 2
centimeters up to 7 centimeters outside the periphery of the useful beam, a
maximum limit of 10 percent of the absorbed dose on the central axis of the
useful beam at the nominal treatment distance;
(3) Measurements of leakage radiation through
the beam limiting for photon radiation devices shall be:
a. Made with the beam limiting devices
closed;
b. Made with any residual
aperture blocked by at least 2 tenth value layers of suitable absorbing
material;
c. Measured independently
at the depth of maximum dose for each set of overlapping beam limiting device;
and
d. The depth of maximum dose
made using a radiation detector of area not exceeding 10 square
centimeters;
(4)
Measurements of leakage radiation through the electron applicators shall:
a. Be made with the electron beam directed
into the air;
b. Use a radiation
detector of area up to but not exceeding 1 square centimeter suitably protected
against radiation which has been scattered from material beyond the radiation
detector; and
c. Be made using one
centimeter of water equivalent build up material; and
(5) Leakage radiation through beam limiting
devices shall be determined for photon radiation and for electron radiation in
radiation therapy machines which operate in both modes.
(d) Filters and wedges used in therapeutic
radiation machines shall meet the following requirements:
(1) Each wedge filter which is removable from
the system shall be clearly marked with an identification number;
(2) Each removable wedge filter shall have
the nominal wedge angle appear on the wedge or wedge tray if permanently
mounted to the tray;
(3) If the
wedge or wedge tray is significantly damaged, the wedge transmission factor
shall be redetermined;
(4) If the
absorbed dose rate information required by
He-P
4047.07(i) relates exclusively to
operation with a field flattening filter or beam scattering foil in place, such
foil or filter shall be removable only by the use of tools; and
(5) For equipment which utilizes a system of
wedge filters, interchangeable field flattening filters, or interchangeable
beam scattering foils:
a. Irradiation shall
not be possible until a selection of a filter or a positive selection to use
"no filter" has been made at the treatment control panel, either manually or
automatically;
b. An interlock
system shall be provided to prevent irradiation if the filter selected is not
in the correct position;
c. A
display shall be provided at the treatment control panel showing the wedge
filter(s), interchangeable field flattening filter(s), and/or interchangeable
beam scattering foil(s) in use; and
d. An interlock shall be provided to prevent
irradiation if any filter and/or beam scattering foil selection operation
carried out in the treatment room does not agree with the filter and/or beam
scattering foil selection operation carried out at the treatment control
panel.
(e)
The registrant shall determine during acceptance testing, or obtain from the
manufacturer, data sufficient to ensure that X ray stray radiation in the
useful electron beam, absorbed dose at the surface during X-ray irradiation and
stray neutron radiation in the useful X-ray beam are in compliance with
International Electrotechnical Commission (IEC) Document 60601-2-1 which is
incorporated by reference and included in Appendix A.
(f) All therapeutic radiation machines
subject to He-P 4047.07 shall meet the
following requirements:
(1) All therapeutic
radiation machines shall be provided with:
a.
Redundant beam monitoring systems which have sensors fixed in the useful beam
during treatment to indicate the dose monitor unit rate;
b. At least 2 independently powered
integrating dose meters if manufactured after July 1, 1998; and
c. At least one radiation detector
incorporated into a useful beam monitoring system, if manufactured before July
1, 1998; and
(2) The
detector and the system into which that detector is incorporated shall meet the
following requirements:
a. Each detector
shall be removable only with tools;
b. If the detector is movable, it shall be
interlocked to prevent incorrect positioning;
c. Each detector shall form part of a beam
monitoring system from whose readings in dose monitor units the absorbed dose
at a reference point can be calculated;
d. Each beam monitoring system shall be
capable of independently monitoring, interrupting, and terminating
irradiation;
e. The design of the
beam monitoring systems shall ensure that the:
1. Malfunctioning of one system shall not
affect the correct functioning of the other system(s); and
2. Failure of either system shall terminate
irradiation or prevent the initiation of radiation; and
f. Each beam monitoring system shall have a
legible display at the treatment control panel which shall:
1. Maintain a reading until intentionally
reset;
2. Have only one scale and
no electrical or mechanical scale multiplying factors;
3. Utilize a design such that increasing dose
is displayed by increasing numbers; and
4. In the event of power failure, the beam
monitoring information displayed at the control panel at the time of failure
shall be retrievable in at least one system for a 20-minute period of
time.
(g) Bent-beam linear accelerators shall be
provided with auxiliary device(s) to monitor beam symmetry which:
(1) Shall be able to detect field asymmetry
greater than 10 percent; and
(2)
Shall be configured to terminate irradiation if the specifications above cannot
be maintained.
(h)
Selection and display of dose monitor units shall be as follows:
(1) Irradiation shall not be possible until a
new selection of a number of dose monitor units has been made at the treatment
control panel;
(2) The pre-selected
number of dose monitor units shall be displayed at the treatment control panel
until reset manually for the next irradiation;
(3) After termination of irradiation, it
shall be necessary to reset the dosimeter display before subsequent treatment
can be initiated; and
(4) After
termination of irradiation, it shall be necessary for the operator to reset the
pre-selected dose monitor units before irradiation can be initiated.
(i) A system shall be provided
from whose readings the air kerma rate or absorbed dose rate a reference point
can be calculated, and which meets the following requirements:
(1) The dose monitor unit rate shall be
displayed at the treatment control panel;
(2) If the equipment can deliver under any
conditions an air kerma rate or absorbed dose rate at the nominal treatment
distance more than twice the maximum value specified by the manufacturer, a
device shall be provided which terminates irradiation when the air kerma rate
or absorbed dose rate exceeds a value twice the specified maximum;
(3) The dose rate at which the irradiation
will be terminated shall be a record maintained by the registrant;
(4) If the equipment can deliver under any
fault condition(s) an air kerma rate or absorbed dose rate at the nominal
treatment distance more than 10 times the maximum value specified by the
manufacturer, a device shall be provided to prevent the air kerma rate or
absorbed dose rate anywhere in the radiation field from exceeding twice the
specified maximum value and to terminate irradiation if the excess absorbed
dose at the nominal treatment distance exceeds 4 Gy (400 rad);
(5) For each therapeutic radiation machine,
the registrant shall determine, or obtain from the manufacturer, the maximum
value(s) for the specified operating conditions; and
(6) Records of maximum value(s) shall be
maintained at the installation for inspection by DHHS/RHS.
(j) Termination of irradiation by the beam
monitoring system or systems during stationary beam radiation therapy shall be
as follows:
(1) Each primary system shall
terminate irradiation when the pre-selected number of dose monitor units has
been detected by the system;
(2) If
the original design of the equipment includes a secondary dose monitoring
system, that system shall be capable of terminating irradiation when not more
than 15 percent or 40 dose monitor units, above the pre-selected number of dose
monitor units, set at the control panel, has been detected by the secondary
dose monitoring system; and
(3) An
indicator on the control panel shall show which monitoring system has
terminated radiation.
(k) It shall be possible to terminate
irradiation and equipment movement or go from an interruption condition to
termination condition at any time from the operator's position at the treatment
control panel.
(l) If a therapeutic
radiation machine has an interrupt mode:
(1)
It shall be possible to interrupt irradiation and equipment movements at any
time from the treatment control panel;
(2) Following an interruption it shall be
possible to restart irradiation by operator action without any re-selection of
operating conditions; and
(3) If
any change is made of a pre-selected value during an interruption, irradiation
and equipment movements shall be automatically terminated.
(m) A suitable irradiation control device
shall be provided to terminate the irradiation after a pre-set time interval as
follows:
(1) A timer shall be provided which
has a display at the treatment control panel;
(2) The time provided shall have a pre-set
time selector and an elapsed time indicator;
(3) The timer shall be a cumulative timer
which activates with an indication of "BEAM-ON" and retains its reading after
irradiation is interrupted or terminated;
(4) After irradiation is terminated and
before irradiation can be re-initiated, it shall be necessary to reset the
elapsed time indicator; and
(5) The
timer shall terminate irradiation when a pre-selected time has elapsed, if the
dose monitoring systems have not previously terminated irradiation.
(n) Equipment capable of both
x-ray therapy and electron therapy shall meet the following additional
requirements:
(1) Irradiation shall not be
possible until a selection of radiation type (x-rays or electrons) has been
made at the treatment control panel;
(2) The radiation type selected shall be
displayed at the treatment control panel before and during
irradiation;
(3) An interlock
system shall be provided to ensure that the equipment can principally emit only
the radiation type which has been selected;
(4) An interlock system shall be provided to
prevent irradiation with x-rays, except to obtain an image, when electron
applicators are fitted;
(5) An
interlock system shall be provided to prevent irradiation with electrons when
accessories specific for x-ray therapy are fitted; and
(6) An interlock system shall be provided to
prevent irradiation if any selected operations carried out in the treatment
room do not agree with the selected operations carried out at the treatment
control panel.
(o)
Equipment capable of generating radiation beams of different energies shall
meet the following requirements:
(1)
Irradiation shall not be possible until a selection of energy has been made at
the treatment control panel;
(2)
The nominal energy value selected shall be displayed at the treatment control
panel until reset manually for the next irradiation;
(3) After termination of irradiation, it
shall be necessary to reset the nominal energy value selected before subsequent
treatment can be initiated;
(4)
Irradiation shall not be possible until the appropriate flattening filter or
scattering foil for the selected energy is in its proper location;
and
(5) The selection of energy
shall be in compliance with International Electrotechnical Commission (IEC)
Document 60601-2-1, which is incorporated by reference and included in Appendix
A.
(p) Therapeutic
radiation machines capable of both stationary beam radiation therapy and moving
beam radiation therapy shall meet the following requirements:
(1) Irradiation shall not be possible until a
selection of stationary beam radiation therapy or moving beam radiation therapy
has been made at the treatment control panel;
(2) The mode of operation shall be displayed
at the treatment control panel;
(3)
An interlock system shall be provided to ensure that the equipment can operate
only in the mode which has been selected;
(4) An interlock system shall be provided to
prevent irradiation if any selected parameter in the treatment room does not
agree with the selected parameter at the treatment control panel;
(5) Moving beam irradiation therapy shall be
controlled to obtain the selected relationships between incremental dose
monitor units and incremental movement as follows:
a. An interlock system shall be provided to
terminate irradiation if the number of dose monitor units delivered in any 10
degrees of rotation or one cm of linear motion differs by more than 20 percent
from the selected value;
b. Where
angle terminates the irradiation in moving beam radiation therapy, the dose
monitor units delivered shall differ by less than 5 percent from the dose
monitor unit value selected;
c. An
interlock shall be provided to prevent motion of more than 5 degrees or 1 cm
beyond the selected limits during moving beam radiation therapy;
d. An interlock shall be provided to require
that a selection of direction be made at the treatment control panel in all
units which are capable of both clockwise and counter-clockwise moving beam
radiation therapy; and
e. Moving
beam radiation therapy shall be controlled with both primary position sensors
and secondary position sensors to obtain the selected relationships between
incremental dose monitor units and incremental movement;
(6) Where the beam monitor system terminates
the irradiation in moving beam radiation therapy, the termination of
irradiation shall be as required by
He-P
4047.07(j) ; and
(7) An interlock system shall be provided to
terminate irradiation if movement:
a. Occurs
during stationary beam radiation therapy; or
b. Does not start or stops during moving beam
radiation therapy unless stoppage is a pre-planned function.
(q) In addition to
shielding adequate to meet requirements of
He-P
4047.09, the following design requirements are made
for therapeutic radiation machines operating above 500 kV:
(1) All protective barriers shall be fixed,
except for access doors to the treatment room or movable beam
interceptors;
(2) The control panel
shall:
a. Be located outside the treatment
room;
b. Provide an indication of
whether electrical power is available at the control panel and if activation of
the radiation is possible;
c.
Provide an indication or whether radiation is being produced; and
d. Include an access control (locking) device
which will prevent unauthorized use of the therapeutic radiation
machine;
(3) Windows,
mirrors, closed-circuit television, or an equivalent viewing system shall be
provided to permit continuous observation of the patient following positioning
and during irradiation and shall be so located that the operator may observe
the patient from the treatment control panel;
(4) The therapeutic radiation machine shall
not be used for patient irradiation unless at least one viewing system is
operational;
(5) Provision shall be
made for continuous two-way aural communication between the patient and the
operator at the control panel;
(6)
The therapeutic radiation machine shall not be used for irradiation of patients
unless continuous two-way aural communication is possible;
(7) Treatment room entrances shall be
provided with warning lights in a readily observable position near the outside
of all access doors, which will indicate when the useful beam is "ON" and when
it is "OFF";
(8) Interlocks shall
be provided such that all access controls are activated before treatment can be
initiated or continued;
(9) If the
radiation beam is interrupted by any access control, it shall not be possible
to restore the machine to operation without resetting the access control and
re-initiating irradiation by manual action at the control panel;
(10) If the shielding material in any
protective barrier requires the presence of a beam interceptor to ensure
compliance with He-P4020.13, interlocks shall be provided to prevent the
production of radiation, unless the beam interceptor is in place, whenever the
useful beam is directed at the designated barrier(s);
(11) At least one emergency power cutoff
switch in addition to the termination switch shall be located in the radiation
therapy room and shall terminate all equipment electrical power including
radiation and mechanical motion;
(12) All emergency power cutoff switches
shall include a manual reset so that the therapeutic radiation machine cannot
be restarted from the unit's control console without resetting the emergency
cutoff switch;
(13) All safety
interlocks shall be designed so that any defect or component failure in the
safety interlock system prevents or terminates operation of the therapeutic
radiation machine; and
(14) Surveys
for residual activity shall be conducted on all therapeutic radiation machines
capable of generating photon and electron energies above 10 MV prior to
machining, removing, or working on therapeutic radiation machine components
which may have become activated due to photo-neutron production.
(r) The services of the radiation
therapy physicist shall be required in facilities having therapeutic radiation
machines with energies of 500 kV and above.
(s) The radiation therapy physicist required
in He-P
4047.07(r) shall be responsible for:
(1) Full calibration(s) required by
He-P
4047.07(u) ;
(2) Protection surveys required by
He-P
4047.05;
(3) Supervision and review of
dosimetry;
(4) Beam data
acquisition and transfer for computerized dosimetry and supervision of its
use;
(5) Quality assurance,
including quality assurance check review required by
He-P
4047.07(v) (6) ;
(6) Consultation with the authorized user in
treatment planning, as needed; and
(7) Performing calculation and assessments
regarding misadministrations.
(t) The following operating procedures shall
be required:
(1) If the radiation therapy
physicist is not a full-time employee of the registrant, the operating
procedures shall specifically address how the radiation therapy physicist is to
be contacted for problems or emergencies, as well as the specific actions, if
any, to be taken until the radiation therapy physicist can be
contacted;
(2) No individual, other
than the patient, shall be in the treatment room during treatment or during any
irradiation for testing or calibration purposes;
(3) Therapeutic radiation machines shall not
be made available for medical use unless the requirements of
He-P
4047.05 and
He-P
4047.07(u) and (v) have been
met;
(4) Therapeutic radiation
machines, when not in operation, shall be secured to prevent unauthorized
use;
(5) When adjustable beam
limiting devices are used, the position and shape of the radiation field shall
be indicated by a light field;
(6)
If a patient must be held in position during treatment, mechanical supporting
or restraining devices shall be used; and
(7) A copy of the current operating and
emergency procedures shall be maintained at the therapeutic radiation machine
control console.
(u)
Acceptance testing, commissioning, and full calibration measurements shall be
as follows:
(1) Acceptance testing,
commissioning, and full calibration of a therapeutic radiation machine shall be
performed by, or under the direct supervision of, a radiation therapy
physicist;
(2) Acceptance testing
and commissioning shall be performed in accordance with "AAPM Code of Practice
for Radiotherapy Accelerators" (1994) prepared by AAPM Radiation Therapy Task
Group 45, which is incorporated by reference and included in Appendix A, and
the manufacturer's contractual specifications;
(3) Acceptance testing and commissioning
shall be conducted before the first medical use following installation or
reinstallation of the therapeutic radiation machine;
(4) Full calibration shall include
measurement of all parameters required by Table II of "Comprehensive QA for
Radiation Oncology" (1994) prepared by AAPM Radiation Therapy Committee Task
Group 40, which Report and Table are incorporated by reference and included in
Appendix A;
(5) Full calibration
shall be performed in accordance with "AAPM Code of Practice for Radiotherapy
Accelerators" (1994) prepared by AAPM Radiation Therapy Task Group 45 which is
incorporated by reference and included in Appendix A;
(6) It shall not be necessary to complete all
elements of a full calibration at the same time, all applicable parameters (for
all energies) shall be completed at intervals not exceeding 12 calendar months,
unless a more frequent interval is required in Table II of "Comprehensive QA
for Radiation Oncology" (1994) prepared by AAPM Radiation Therapy Committee
Task Group 40, which is incorporated by reference and included in Appendix
A;
(7) The radiation therapy
physicist shall perform all elements of a full calibration necessary to
determine that all parameters are within acceptable limits as follows:
a. Whenever quality assurance check
measurements indicate that the radiation output differs by more than 5 percent
from the value obtained at the last full calibration and the difference cannot
be reconciled;
b. Therapeutic
radiation machines with multi-energy and/or multi-mode capabilities shall only
require measurements for those modes and/or energies that are not within their
acceptable range;
c. Following any
component replacement, major repair, or modification of components that could
significantly affect the characteristics of the radiation beam;
d. If the repair, replacement, or
modification does not affect all modes and/or energies, measurements shall be
performed on the effected mode/energy that is in most frequent clinical use at
the facility; and
e. The remaining
energies/modes may be validated with quality assurance check procedures against
the criteria in He-P 4047.07(u) (7)
;
(8) The registrant shall use the dosimetry
system described in He-P 4047.05(h) and (i)
to measure the radiation output for one set
of exposure conditions;
(9) The
remaining radiation measurements required in
He-P
4047.07(u) (2) -(6) may be made
using a dosimetry system that indicates relative dose rates;
(10) The registrant shall maintain a record
of each calibration in an auditable form for the life of the therapeutic
radiation machine; and
(11) The
record required in He-P 4047.07(u) (10)
shall include:
a. The date of the calibration;
b. The manufacturer's name;
c. Model number of the therapeutic
machine;
d. Serial number of the
therapeutic machine;
e. The model
numbers and serial numbers of the instruments used to calibrate the therapeutic
radiation machine; and
f. The
signature of the radiation therapy physicist responsible for performing the
calibration.
(v) Periodic quality assurance checks shall
meet the following requirements:
(1) Periodic
quality assurance checks shall be performed on all therapeutic radiation
machines subject to He-P 4047.07 at intervals not to
exceed those specified in "Comprehensive QA for Radiation Oncology" (1994),
prepared by AAPM Radiation Therapy Committee Task Group 40, which is
incorporated by reference and included in Appendix A;
(2) Quality assurance checks shall include
determination of central axis radiation output and a representative sampling of
periodic quality assurance checks contained in "Comprehensive QA for Radiation
Oncology" (1994) prepared by AAPM Radiation Therapy Committee Task Group 40
which is incorporated by reference and included in Appendix A;
(3) Representative sampling as required in
He-P
4047.07(v) (2) shall include all
referenced periodic quality assurance checks in an interval not to exceed 12
consecutive calendar months;
(4)
The registrant shall use a dosimetry system which has been intercompared within
the previous 12 months with the dosimetry system described in
He-P
4047.05(h) and (i) to make the
periodic quality assurance checks;
(5) The registrant shall perform periodic
quality assurance checks in accordance with procedures established by the
radiation therapy physicist;
(6)
The registrant shall review the results of each periodic radiation output check
according to the following procedures:
a. The
authorized user and radiation therapy physicist shall be immediately notified
if any parameter is not within its acceptable tolerance;
b. The therapeutic radiation machine shall
not be made available for subsequent medical use until the radiation therapy
physicist has determined that all parameters are within their acceptable
tolerances;
c. If all quality
assurance check parameters appear to be within their acceptable range, the
quality assurance check shall be reviewed and signed by either the authorized
user or radiation therapy physicist within 3 treatment days; and
d. The radiation therapy physicist shall
review and sign the results of each radiation output quality assurance check at
intervals not exceed 30 days;
(7) Therapeutic radiation machines subject to
He-P
4047.07 shall have safety quality assurance checks
listed in "Comprehensive QA for Radiation Oncology" (1994) prepared by AAPM
Radiation Therapy Committee Task Group 40, which is incorporated by reference
and included in Appendix A, performed at intervals not to exceed 7
days;
(8) To satisfy the
requirement of He-P 4047.07(v) (7)
, safety quality assurance checks shall
ensure proper operation of:
a. Electrical
interlocks at each external beam radiation therapy room entrance;
b. The "BEAM-ON", interrupt, and termination
switches;
c. Beam condition
indicator lights on the access doors, control console, and in the radiation
therapy room;
d. Viewing
systems;
e. Electrically operated
treatment room door(s) from inside and outside the treatment room;
and
f. At least one emergency power
cutoff switch, as follows:
1. If more than
one emergency power cutoff switch is installed and not all switches are tested
at once, each switch shall be tested on a rotating basis; and
2. Safety quality assurance checks of the
emergency power cutoff switches may be conducted at the end of the treatment
day in order to minimize possible stability problems with the therapeutic
radiation machine.
(9) The registrant shall promptly repair any
system identified in He-P 4047.07(v) (8)
that is not operating properly;
(10) The registrant shall maintain a record
of each quality assurance check required for 3 years; and
(11) The record required in
He-P
4047.07(v) (10) shall include:
a. The date of the quality assurance
check;
b. The manufacturer's
name;
c. The machine model
number;
d. The machine serial
number;
e. The manufacturer's name,
model number, and serial number of the instrument(s) used to measure the
radiation output of the therapeutic radiation machine; and
f. The signature of the individual who
performed the periodic quality assurance check.
(w) For intensity modulated radiation therapy
(IMRT), quality assurance checks shall:
(1)
Include commissioning and testing of the treatment planning and delivery
systems, routine quality assurance of the delivery system, and patient specific
validation of treatment plan;
(2)
Be performed in accordance with "Guidance document on delivery, treatment
planning, and clinical implementation of IMRT" (2003) Report of the IMRT
subcommittee of the AAPM radiation therapy committee: AAPM Report No. 82, which
is incorporated by reference and included in Appendix A;
(3) Be performed in accordance with the
manufacturer's contractual specifications.
#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06,
EXPIRES: 2-2-07; ss by #8808, eff
1-24-07